Literature DB >> 29723533

Early Degeneration Caused by Cusp Tear of First-Generation Trifecta Bioprosthesis.

Simone Eichinger1, Akmal M A Badreldin1, Walter B Eichinger2.   

Abstract

The Trifecta aortic valve has excellent hemodynamic performance as result of an expansive valve design with a bovine pericardial sheet externally mounted on a titanium stent. We report 2 cases of early Trifecta valve degeneration, both caused by partial rupture of one of the leaflet cusps 3 and 4 years post-implant. Post discharge, both patients had routinely performed echocardiography check-ups, without signs of valve failure. Transesophageal echocardiography performed during emergency hospital readmission due to severe dyspnea revealed transvalvular aortic regurgitation without signs of endocarditis. Urgent redo-surgery was successful in both cases, replacing the Trifecta prosthesis with a different bovine bioprosthesis.
Copyright © 2018 The Society of Thoracic Surgeons. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2018        PMID: 29723533     DOI: 10.1016/j.athoracsur.2018.03.063

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  5 in total

1.  Impact of postoperative patient-prosthesis mismatch as a risk factor for early structural valve deterioration after aortic valve replacement with Trifecta bioprosthesis.

Authors:  Tatsuto Wakami; Shigeki Koizumi; Tadaaki Koyama
Journal:  J Cardiothorac Surg       Date:  2022-07-08       Impact factor: 1.522

2.  Early surgical aortic bioprosthesis failure caused by cusp tear: A word of caution.

Authors:  Thibault Schaeffer; Denis Berdajs; Oliver Reuthebuch
Journal:  Clin Case Rep       Date:  2020-11-28

3.  Trifecta versus Perimount Magna Ease aortic valves: Failure mechanisms.

Authors:  Ryo Suzuki; Toshiro Ito; Masato Suzuki; Shunsuke Ohori; Ryo Takayanagi; Shiro Miura
Journal:  Asian Cardiovasc Thorac Ann       Date:  2022-05-22

4.  Early failure of the Trifecta GT bioprostheses.

Authors:  Lise Tchouta; Daniel Liesman; Karen Kim; Shinichi Fukuhara
Journal:  JTCVS Tech       Date:  2020-08-12

5.  Commentary: Trifecta valve: Does a word of caution prevail?

Authors:  Ko Bando
Journal:  JTCVS Tech       Date:  2020-11-18
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.