| Literature DB >> 29722356 |
Chengyu Shi1, Xiaoli Tang1, Maria Chan1.
Abstract
OBJECTIVE: The newly released Respiratory Gating for Scanners (RGSC; Varian Medical Systems, Palo Alto, CA, USA) system has limited existing quality assurance (QA) protocols and pertinent publications. Herein, we report our experiences of the RGSC system acceptance and QA.Entities:
Keywords: RGSC; breathing curves; computed tomography; dynamic phantom; quality assurance
Year: 2017 PMID: 29722356 PMCID: PMC5927385 DOI: 10.1002/pro6.34
Source DB: PubMed Journal: Precis Radiat Oncol ISSN: 2398-7324
Acceptance tasks of a new Respiratory Gating for Scanners system[7]
| Task | Purpose | Expected result |
|---|---|---|
| 1 | Software version | Software version ≥1.7 |
| 2 | Scanner vendor and camera type | Scanner manufacturer should be verified and camera type is correct |
| 3 | Integration with ARIA system | ARIA is higher than version 13 |
| 4 | Calibration of camera | Pass |
| 5 | Integration with scanner | Hardware connection and software control |
| 6 | Camera motion position accuracy in 3D | ≤2 mm |
| 7 | Export breathing curve | File can be exported to destination |
| 8 | Export patient into database | File can be exported into the database |
| 9 | Export breathing pattern as DICOM | DICOM file can be exported |
| 10 | Documents | User related documents are available from vendor |
DICOM, Digital Imaging and Communications in Medicine.
FIGURE 1Scheme of Respiratory Gating for Scanners (RGSC) and computed tomography (CT) scanner connections
FIGURE 2Diagram of spatial location accuracy tests. The white plate provides background reflection and reference locations to place the reflector block. The newly designed reflector block with four markers was aligned with the wall lasers and at a location with a number on the white plate
FIGURE 3QUASAR phantom setup for dynamic localization accuracy and screenshot of Respiratory Gating for Scanners software (here 8 s per breathing cycle)
Static physics coordinate accuracy results
| Direction | Shift amount | ||||||
|---|---|---|---|---|---|---|---|
| 0.2 cm | 0.5 cm | 1 cm | 3 cm | 5 cm | 7 cm | 10 cm | |
| X-Left of the observer | 0.2 | 0.5 | 1.0 | 3.0 | 5.0 | 7.0 | 10.0 |
| X-Right of the observer | 0.2 | 0.5 | 1.0 | 3.0 | 5.0 | 7.0 | 10.0 |
| Y-In to the gantry | 0.2 | 0.5 | 1.0 | 3.0 | 4.8 | 6.8 | 9.8 |
| Y-Out of the gantry | 0.2 | 0.5 | 1.0 | 2.9 | 4.8 | 6.8 | 9.8 |
| Z-Up shift | 0.2 | 0.5 | 1.0 | 3.0 | 5.0 | 7.0 | 9.9 |
| Z-Down shift | 0.2 | 0.5 | 1.0 | 3.0 | 5.0 | 7.0 | 10.0 |
FIGURE 4Comparison between programmed period and recorded period
FIGURE 5Comparison of the calculated period for patient-specific breathing curves recorded by Real-time Position Management (RPM) and Respiratory Gating for Scanners (RGSC) systems
FIGURE 6Trace plot for the first 30 min since the camera was plugged in. The y-axis shows the distance away from the calibration point. The x-axis shown time elapsed since the camera was plugged in
Recommended quality assurance tasks for daily, monthly, and annual quality assurance
| Period | Tasks | Tolerance |
|---|---|---|
| Daily | Calibration of camera | Pass/fail |
| Monthly | Calibration of camera | Pass/fail |
| Spatial localization accuracy | ≤2 mm7 | |
| Dynamic localization accuracy – regular | ≤0.1 s3 for the period of range of 3–10 s | |
| Annual | Calibration of camera | Pass/fail |
| Integration of peripheral equipment | End-to-end pass/fail | |
| Spatial localization accuracy | ≤2 mm | |
| Dynamic localization accuracy-regular | ≤ 0.1 s3 for the period of range of 3–10 s | |
| Dynamic localization accuracy-irregular | ≤ 0.2 s |
Based on the authors’ experience.