Literature DB >> 29657380

Preprocedural Check of Central Venous Catheter Set.

Komal Anil Gandhi1, Tanvir Samra1.   

Abstract

Entities:  

Year:  2018        PMID: 29657380      PMCID: PMC5879865          DOI: 10.4103/ijccm.IJCCM_360_17

Source DB:  PubMed          Journal:  Indian J Crit Care Med        ISSN: 0972-5229


× No keyword cloud information.
Sir, Structural manufacturing defect in central venous catheter (CVC) set poses challenges during insertion. We report an unusual defect in the introducer needle of CVC set (Certofix® trio) and propose a checklist to detect the same. An ultrasound (US)-guided right internal jugular vein (RIJV) CVC insertion was planned in an elderly patient (American Society of Anesthesiologists III) scheduled for total hip replacement surgery. Before insertion, CVC and guidewire were visually inspected and all three lumens of catheter were flushed with saline. During this preprocedural checkout, resistance was encountered while advancing guidewire through side-port adapter of valve needle. Hence, guidewire insertion was tried through straight port of introducer needle, but it also failed. On careful observation, malalignment of lumen of needle and hub was seen. This was preventing the guidewire to negotiate needle hub [Figure 1a and b]. This malalignment also prevented forward flow of saline through syringe attached to the straight port of the introducer needle.
Figure 1

(a) (top) The guidewire in an aberrant track (arrow) when inserted through the side port of introducer needle. (b) (below) Failure to negotiate the guidewire (arrow) on insertion through the straight port of introducer needle

(a) (top) The guidewire in an aberrant track (arrow) when inserted through the side port of introducer needle. (b) (below) Failure to negotiate the guidewire (arrow) on insertion through the straight port of introducer needle A 16G intravenous (IV) cannula was then used as a rescue introducer needle and RIJV was punctured in single attempt under US guidance. The hollow metal stylet of cannula was removed and the guidewire was easily passed through the plastic catheter of 16G IV cannula. No procedure-related complications were observed. Manufacturing defects in CVC set are sporadic events but they can cause serious mishaps. Faults in design of the guidewire or in manufacturing process have caused breakage of a guidewire and its lodgment in IJV, necessitating fluoroscopic guidance for removal.[1] Hegde et al. encountered abnormal communication between two channels of CVC intraoperatively.[2] We could identify the said defect before insertion, thus highlighting the importance of preprocedural check. If unnoticed, multiple punctures in IJV would have been done with defective introducer needle. Such multiple attempts increase the incidence of mechanical complications such as hematomas, inadvertent arterial puncture, and pneumothorax.[3] Use of 16G IV cannula as an introducer needle was a temporary solution, but it is technically more challenging. There is no stiff part on the cannula to hold with sufficient stability during guidewire advancement, so the entire length of the IV plastic catheter needs to be inserted in the lumen of IJV. Moreover, absence of adapter at cannula hub increases difficulty for insertion of guidewire using dispenser. Lee et al. reported higher success rates and lower complications with thin-wall introducer needle technique when compared with a cannula-over-needle technique during IJV catheterization.[4] Hence, it is always advisable that physicians should perform preuse check of all medical equipment to avoid potential complications. A simplified preprocedural checklist for CVC set used in our hospital is described in Table 1.
Table 1

Preprocedural checklist for CVC

Preprocedural checklist for CVC

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.
  4 in total

1.  Broken guide wire--a fault of design?

Authors:  Enrico Monaca; Stefan Trojan; John Lynch; Manfred Doehn; Frank Wappler
Journal:  Can J Anaesth       Date:  2005-10       Impact factor: 5.063

2.  Communication between two channels of central venous catheter: a rare manufacturing defect.

Authors:  Harihar V Hegde; Roopa Sachidananda; Raghavendra P Rao; Shrinivas S Deshapande
Journal:  Anesth Analg       Date:  2009-12       Impact factor: 5.108

3.  Complications and failures of subclavian-vein catheterization.

Authors:  P F Mansfield; D C Hohn; B D Fornage; M A Gregurich; D M Ota
Journal:  N Engl J Med       Date:  1994-12-29       Impact factor: 91.245

4.  Comparison of Needle Insertion and Guidewire Placement Techniques During Internal Jugular Vein Catheterization: The Thin-Wall Introducer Needle Technique Versus the Cannula-Over-Needle Technique.

Authors:  Yong Hun Lee; Tae Kyong Kim; Yoo Sun Jung; Youn Joung Cho; Susie Yoon; Jeong-Hwa Seo; Yunseok Jeon; Jae Hyon Bahk; Deok Man Hong
Journal:  Crit Care Med       Date:  2015-10       Impact factor: 7.598

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.