Abstract

PURPOSE: To report a potentially serious complication resulting from a faulty guide wire during central venous catheterization. CLINICAL FEATURES: We report breakage of a guide wire with potential severe implications for a 47-yr-old critical care patient. The tip of the J-wire became lodged in the internal jugular vein, and required fluoroscopic guidance for removal, which occurred without complication.
CONCLUSION: Breakage of a guide wire during central venous catheter insertion has been very rarely reported. However, inherent faults in design or in the manufacturing process of the guide wire could lead to this rare complication. A simple bedside test is proposed to detect breakage in the core section of the guide wire.

Entities:  

Mesh:

Year:  2005        PMID: 16189330     DOI: 10.1007/BF03021773

Source DB:  PubMed          Journal:  Can J Anaesth        ISSN: 0832-610X            Impact factor:   5.063


  18 in total

1.  Complications of Central Venous Catheterisation: Breakage of guidewire-a disaster averted.

Authors:  Qutaiba A Tawfic; Pradipta Bhakta; Jyoti Burad; Pragyandipta Mishra; Rajini Kausalya
Journal:  Sultan Qaboos Univ Med J       Date:  2011-10-25

2.  [Knotted Seldinger wire].

Authors:  A Fiala; B Glodny; I H Lorenz
Journal:  Anaesthesist       Date:  2015-12-11       Impact factor: 1.041

3.  An intra-arterial knotted guide wire.

Authors:  Bosco Chandra Kumar Asayas; Prem Anand John; Sushrut Takroo
Journal:  Indian J Surg       Date:  2012-12-08       Impact factor: 0.656

4.  Successful removal of kinked J-guide wire under fluoroscopic guidance during central venous catheterization -A case report-.

Authors:  Hyun-Seok Han; Young-Tae Jeon; Hyo-Seok Na; Jin-Young Hwang; Eun-Joo Choi; Mi-Hyun Kim
Journal:  Korean J Anesthesiol       Date:  2011-05-31

5.  Fracture of J-tipped guidewire during central venous catheterization and its successful removal under fluoroscopic guidance -A case report-.

Authors:  Sun Kyung Park; In-Kyong Yi; Jae-Ho Lee; Dae-Hee Kim; Sook Young Lee
Journal:  Korean J Anesthesiol       Date:  2012-11-16

6.  Damaged guidewire by the introducer needle tip while inserting central venous catheter in subclavian vein by supraclavicular approach.

Authors:  Rakesh Garg; Vijay Kumar Ramaiah; R S Chouhan
Journal:  Saudi J Anaesth       Date:  2010-09

7.  Oxygen desaturation following methylene blue injection: Not always spurious.

Authors:  Uma Hariharan; Rajesh Sood; Arindam Choudhury; Rakesh Garg; Jaswinder Kaur
Journal:  Saudi J Anaesth       Date:  2011-01

8.  Intracranial entrapment of a haemodialysis catheter guidewire.

Authors:  Natasha Hemicke Langer; Lars Hein; Morten Heiberg Bestle
Journal:  BMJ Case Rep       Date:  2020-08-25

9.  Jugular Venous Catheterization: A Case of Knotting.

Authors:  E Erkılıç; M Aksoy; T Gümüş; S Sarı; E Kesimci
Journal:  Case Rep Med       Date:  2015-11-08

10.  Loss of guide wire: a lesson learnt review of literature.

Authors:  Rajiv Srivastav; Vishal Yadav; Dimpy Sharma; Vikas Yadav
Journal:  J Surg Tech Case Rep       Date:  2013-07
View more