Hiroshi Ishida1, Tadanobu Suehiro1, Keita Suzuki1, Tsuyoshi Yoneda1, Susumu Watanabe1.
Abstract
[Purpose] This study aimed to assess the influence of ultrasound (US) transducer tilt on muscle thickness and echo intensity of the rectus femoris muscle (RF) in healthy subjects.Entities:
Keywords: Echo intensity; Muscle thickness; Ultrasound
Year: 2017 PMID: 29643602 PMCID: PMC5890228 DOI: 10.1589/jpts.29.190
Source DB: PubMed Journal: J Phys Ther Sci ISSN: 0915-5287
Intrarater reliablitity
| ICC | 95% CI | SEM | MDC | |
|---|---|---|---|---|
| Transducer tilt (°) | 0.961 | (0.888–0.987) | 0.9 | 2.6 |
| Muscle thickness (mm) | 0.989 | (0.968–0.996) | 0.4 | 1.0 |
| Echo intensity | 0.974 | (0.924–0.991) | 0.6 | 1.7 |
ICC: intra-class correlation coefficients; CI: confidence interval; SEM: standard error of measurement; MDC: minimal detectable change.
Mean (± standard deviation) transducer tilt, muscle thickness and echo intensity under the reference and three experimental conditions
| Reference | + 3° | + 6° | + 9° | |
|---|---|---|---|---|
| Transducer tilt (°) | 2.8 ± 4.8 | 5.9 ± 4.8 a | 8.8 ± 4.8 a b | 11.9 ± 4.7 a b c |
| Difference | 3.1 ± 0.2 | 6.0 ± 0.2 | 9.0 ± 0.2 | |
| Muscle thickness (mm) | 18.9 ± 3.5 | 18.4 ± 3.3 | 18.3 ± 3.4 | 18.3 ± 3.4 |
| Difference | –0.4 ± 0.5 | –0.5 ± 0.5 | –0.5 ± 0.5 | |
| Echo intensity | 75.2 ± 3.8 | 73.6 ± 5.3 | 71.7 ± 6.0 a b | 69.3 ± 6.0 a b c |
| Difference | –1.6 ± 3.7 | –3.5 ± 4.5 | –5.9 ± 5.1 |
aSignificantly different from the reference condition (p<0.05; Bonferroni’s test). bSignificantly different from the +3°condition (p<0.05; Bonferroni’s test). cSignificantly different from the +6°condition (p<0.05; Bonferroni’s test).