Influence of inward pressure of the transducer on lateral abdominal muscle thickness during ultrasound imaging.

Hiroshi Ishida1, Sususmu Watanabe.   

Abstract

STUDY
DESIGN: Controlled laboratory study, technical note.
OBJECTIVES: The purpose of this study was to quantify changes in the thickness of the transversus abdominis, internal oblique, and external oblique muscles induced by different inward pressures of the transducer during ultrasound imaging (USI).
BACKGROUND: USI of the lateral abdominal muscles is increasingly used in managing musculoskeletal dysfunction. However, to the best of our knowledge, no study has evaluated the influence of different inward pressures of the transducer on the lateral abdominal muscle thickness during USI.
METHODS: Thirty healthy male volunteers participated in this study. The thickness of the transversus abdominis, internal oblique, and external oblique muscles was measured with USI by the same rater in 4 conditions of inward pressures of 0.5, 1.0, 2.0, and 4.0 N. Intraclass correlation coefficients (ICC1,1), with 95% confidence intervals, were calculated, and a repeated-measures analysis of variance was used to assess the influence of inward pressure on the thickness of the lateral abdominal muscles.
RESULTS: The thickness of the transversus abdominis, internal oblique, and external oblique muscles was significantly different among the 4 conditions (P<.038). The mean difference between the 0.5-N and 4.0-N conditions was greater than the minimal detectable change of the 0.5-N condition in the lateral abdominal muscles.
CONCLUSION: The difference in magnitude produced by the forces under different conditions was meaningful. When using a technique that involves a handheld transducer, the examiner should attempt to maintain consistent inward pressure of the transducer during USI to quantify the thickness of lateral abdominal muscles. J Orthop Sports Phys Ther 2012;42(9):815-818, Epub 19 April 2012. doi:10.2519/jospt.2012.4064.

Entities:  

Mesh:

Year:  2012        PMID: 22517245     DOI: 10.2519/jospt.2012.4064

Source DB:  PubMed          Journal:  J Orthop Sports Phys Ther        ISSN: 0190-6011            Impact factor:   4.751


  8 in total

1.  Sonographic morphometry of abductor pollicis brevis: can direct contact yield images comparable with those obtained by the water bath technique?

Authors:  Keitaro Fujino; Katsunori Ohno; Kenta Fujiwara; Atsushi Yokota; Masashi Neo
Journal:  J Med Ultrason (2001)       Date:  2019-04-15       Impact factor: 1.314

2.  Intra-rater reliability of transversus abdominis measurement by a novice examiner: Comparison of "freehand" to "probe force device" method of real-time ultrasound imaging.

Authors:  Vanessa L Kennedy; Carol A Flavell; Kenji Doma
Journal:  Ultrasound       Date:  2019-02-25

3.  Correlation between Peak Expiratory Flow and Abdominal Muscle Thickness.

Authors:  Hiroshi Ishida; Kenichi Kobara; Hiroshi Osaka; Tadanobu Suehiro; Tomotaka Ito; Chiharu Kurozumi; Susumu Watanabe
Journal:  J Phys Ther Sci       Date:  2014-11-13

4.  Reliability of ultrasonography measurement for the longus colli according to inward probe pressure.

Authors:  Byoung-Lock Jeong; Sung-Min Ha; In-Cheol Jeon; Ki-Hoon Hong
Journal:  J Phys Ther Sci       Date:  2015-11-30

Review 5.  Quantitative ultrasound: measurement considerations for the assessment of muscular dystrophy and sarcopenia.

Authors:  Michael O Harris-Love; Reza Monfaredi; Catheeja Ismail; Marc R Blackman; Kevin Cleary
Journal:  Front Aging Neurosci       Date:  2014-07-14       Impact factor: 5.750

6.  Interrater reliability of quantitative ultrasound using force feedback among examiners with varied levels of experience.

Authors:  Michael O Harris-Love; Catheeja Ismail; Reza Monfaredi; Haniel J Hernandez; Donte Pennington; Paula Woletz; Valerie McIntosh; Bernadette Adams; Marc R Blackman
Journal:  PeerJ       Date:  2016-06-21       Impact factor: 2.984

7.  Influence of the ultrasound transducer tilt on muscle thickness and echo intensity of the rectus femoris muscle of healthy subjects.

Authors:  Hiroshi Ishida; Tadanobu Suehiro; Keita Suzuki; Tsuyoshi Yoneda; Susumu Watanabe
Journal:  J Phys Ther Sci       Date:  2017-12-13

8.  Reliability of body composition assessment using A-mode ultrasound in a heterogeneous sample.

Authors:  Monica Miclos-Balica; Paul Muntean; Falk Schick; Horia G Haragus; Bogdan Glisici; Vasile Pupazan; Adrian Neagu; Monica Neagu
Journal:  Eur J Clin Nutr       Date:  2020-09-11       Impact factor: 4.016

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.