Tan Suwatanaviroj1, Weimin He1, Edith Pituskin2, Ian Paterson1, Jonathan Choy1, Harald Becher1.
Abstract
BACKGROUND: There are limited data on what is the minimum change that can be detected in cancer patients undergoing treatment with cardiotoxic drugs and are referred for monitoring left ventricular (LV) function.Entities:
Keywords: contrast echocardiography; ejection fraction; inter-observer variability; minimum detectable difference
Year: 2018 PMID: 29643125 PMCID: PMC5948195 DOI: 10.1530/ERP-18-0003
Source DB: PubMed Journal: Echo Res Pract ISSN: 2055-0464
Figure 1Tracings of LV borders by two readers and the volumes calculated from the tracings, the EF was reported as 48% by reader 1 and 49% by reader 2. (A) Four-chamber view, (B) two-chamber view. EDV, end-diastolic volume; ESV, end-systolic volume.
Baseline characteristics of study participants.
| Patient variable | Total ( |
|---|---|
| Age (years) | 59 ± 12 |
| Female | 120 (80%) |
| Body surface area (m2) | 1.8 ± 0.2 |
| Cancer diagnosis | |
| Breast | 115 (77%) |
| Hematologic | 27 (18%) |
| Other | 8 (5%) |
| Echocardiographic parameters | |
| LVEDV (mL) | 109.6 (93.6–131.8) |
| LVESV (mL) | 46.7 (37.2–62.9) |
| Index LVEDV (mL/m2) | 62.8 (55.0–72.6) |
| Index LVESV (mL/m2) | 26.7 (21.3–33.9) |
| LVEF (%) | 58.0 (50.6–62.8) |
Data are expressed as mean ± s.d., as median (interquartile range) or as number (%).
EDV, end-diastolic volume; EF, ejection fraction; ESV, end-systolic volume; LV, left ventricular.
Inter-observer variability and minimum detectable difference for EF measurements in each EF group.
| EF (%) | ICC | RME | SEM | MDD |
|---|---|---|---|---|
| <53 | 0.99 | 2.3 (0.3–3.8) | 1.1 | 3.0 |
| 53–60 | 0.81 | 1.5 (0.0–3.4) | 1.3 | 3.6 |
| >60 | 0.96 | 1.0 (0.0–2.4) | 1.1 | 3.0 |
Relative mean error (RME) is calculated as a percent and expressed as median (interquartile range). The standard error of measurement (SEM) and the minimum detectable difference (MDD) have the same units as the measurement of interest.
EF, ejection fraction; ICC, intra-class correlation coefficient.
Standard error of measurement and minimum detectable difference for EDV and ESV measurements in each EF group.
| EF (%) | EDV SEM (mL) | ESV SEM (mL) | EDV MDD (mL) | ESV MDD (mL) |
|---|---|---|---|---|
| <53 | 2.5 | 1.4 | 7.0 | 4.0 |
| 53–60 | 2.6 | 1.1 | 7.3 | 3.1 |
| >60 | 2.0 | 1.2 | 5.6 | 3.3 |
The SEM and MDD have the same units as the measurement of interest.
EDV, end-diastolic volume; EF, ejection fraction; ESV, end-systolic volume; MDD, minimum detectable difference; SEM, standard error of measurement.
Patients with real changes in EF (≥4%) in those patients who had follow-up studies.
| EF (%) | Number of patients with follow-up studies (%) | EF changes during follow-up (percentage points) | Number of patients with ‘real change’ (EF changes ≥4%) |
|---|---|---|---|
| <53 | 28 (56) | 10 ± 6 | 18 (36) |
| 53–60 | 31 (62) | 8 ± 5 | 21 (42) |
| >60 | 28 (56) | 8 ± 5 | 20 (40) |
Data are expressed as mean ± s.d. or as number (%).
EF, ejection fraction.