| Literature DB >> 29622791 |
Abstract
The aim of this report is to describe a new device that can validate, by automated auscultation, individual blood pressure (BP) readings taken by automated sphygmomanometers.The Accutension Stetho utilizes a smartphone application in conjunction with a specially designed stethoscope that interfaces directly into the smartphone via the earphone jack. The Korotkoff sounds are recorded by the application and are analyzed by the operator on the screen of the smartphone simultaneously with the images from the sphygmomanometer screen during BP estimation. Current auscultatory validation standards require at least 85 subjects and strict statistical criteria for passage. A device that passes can make no guarantee of accuracy on individual patients. The Accutension Stetho is an inexpensive smartphone/stethoscope kit combination that estimates precise BP values by auscultation to confirm the accuracy of an automated sphygmomanometer's readings on individual patients. This should be of great value for both professional and, in certain circumstances, self-measurement BP. Patients will avoid both unnecessary treatment and errors of underestimation of BP, in which the patient requires therapy. The Stetho's software has been validated in an independent ANSI/AAMI/ISO standard study. The Stetho has been shown to perform without difficulty in multiple deflation-based devices by many manufacturers.Entities:
Mesh:
Year: 2018 PMID: 29622791 PMCID: PMC6026489 DOI: 10.1038/s41371-018-0053-2
Source DB: PubMed Journal: J Hum Hypertens ISSN: 0950-9240 Impact factor: 3.012
Fig. 1Setup and use of Accutension Stetho in a BPM self-validation
Fig. 2Structure of Accutension Stetho: a stethoscope consisting of a stethoscope head, a tube and a microphone with an earphone plug and an app running on the smartphone
Fig. 3Korotkoff sounds strength, cuff pressure oscillation and cuff pressure vs. time during cuff deflation from a BPM. The dots mark the peak of Korotkoff sounds and the peak of cuff pressure oscillation, respectively, and they coincide with each other. An automated BPM refreshes the cuff pressure on its screen when it detects the cuff pressure oscillation peak
Devices with successful testing with the Accutension Stetho
| Brand | Model | Measure during deflation | Linear deflation |
|---|---|---|---|
| Omron | BP786 | Yes | Yes |
| BP760N | Yes | Yes | |
| HEM-7051 | Yes | Yes | |
| HEM-7111 | Yes | Yes | |
| HEM-J12 | Yes | Yes | |
| M3 | Yes | Yes | |
| MABIS HEALTHCARE | 04-596-008 | Yes | Yes |
| HOMEDICS | BPA-060 | Yes | Yes |
| Beurer | BM26 | Yes | Yes |
| LifeSource | UA-767 | Yes | Yes |
| ReliOn | BP200 | Yes | Yes |
| Microlife | 3MC1-PC | Yes | Yes |
| AND | UA705 | Yes | Yes |
| Panasonic | EW3108WQ | Noa | N/A |
aThe Panasonic device estimates BP using inflation-based technology