Guang Chu1, Zhi Zhang, Mengdan Xu, Daini Huang, Qiuyan Dai. 1. Department of Cardiology, Shanghai General Hospital of Nanjing Medical University and Shanghai Jiaotong University, Shanghai, People's Republic of China.
Abstract
OBJECTIVE: The aim of this study was to validate the accuracy of the Accutension XYZ-110 blood pressure (BP) kit according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization (ANSI/AAMI/ISO) 81060-2:2013 standard. PARTICIPANTS AND METHODS: Systolic blood pressure (SBP) and diastolic blood pressure (DBP) were measured simultaneously on the same arm in 85 Chinese adults (female : male=48 : 37) with a mean age of 43.2 years using the mercury sphygmomanometer (two observers) and the Accutension XYZ-110 device (one supervisor). The ANSI/AAMI/ISO 81060-2:2013 standard for the validation of BP-measuring devices in adults was followed precisely. A total of 255 comparison pairs were obtained for analysis. RESULTS: The mean device-observer difference in the 255 separate BP data pairs was 2.45±2.24 mmHg for SBP and 0.69±2.09 mmHg for DBP. The data were in accordance with the criterion 1 of the ANSI/AAMI/ISO 81060-2:2013 standard requirements (≤5±8 mmHg). In addition, the mean device-observer difference of the 85 participants was 2.45±1.47 mmHg for SBP and 0.69±1.36 mmHg for DBP. The device accuracy also fulfilled the criterion 2 with the SD of less than or equal to 6.47 for SBP and less than or equal to 6.90 mmHg for DBP. CONCLUSION: The Accutension XYZ-110 BP kit fulfilled the requirements of the ANSI/AAMI/ISO 81060-2:2013 standard, and hence could be recommended for both clinical and self/home BP measurement in adults.
OBJECTIVE: The aim of this study was to validate the accuracy of the Accutension XYZ-110 blood pressure (BP) kit according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization (ANSI/AAMI/ISO) 81060-2:2013 standard. PARTICIPANTS AND METHODS: Systolic blood pressure (SBP) and diastolic blood pressure (DBP) were measured simultaneously on the same arm in 85 Chinese adults (female : male=48 : 37) with a mean age of 43.2 years using the mercury sphygmomanometer (two observers) and the Accutension XYZ-110 device (one supervisor). The ANSI/AAMI/ISO 81060-2:2013 standard for the validation of BP-measuring devices in adults was followed precisely. A total of 255 comparison pairs were obtained for analysis. RESULTS: The mean device-observer difference in the 255 separate BP data pairs was 2.45±2.24 mmHg for SBP and 0.69±2.09 mmHg for DBP. The data were in accordance with the criterion 1 of the ANSI/AAMI/ISO 81060-2:2013 standard requirements (≤5±8 mmHg). In addition, the mean device-observer difference of the 85 participants was 2.45±1.47 mmHg for SBP and 0.69±1.36 mmHg for DBP. The device accuracy also fulfilled the criterion 2 with the SD of less than or equal to 6.47 for SBP and less than or equal to 6.90 mmHg for DBP. CONCLUSION: The Accutension XYZ-110 BP kit fulfilled the requirements of the ANSI/AAMI/ISO 81060-2:2013 standard, and hence could be recommended for both clinical and self/home BP measurement in adults.
Authors: Jordana B Cohen; Raj S Padwal; Michael Gutkin; Beverly B Green; Michael J Bloch; F Wilford Germino; Domenic A Sica; Anthony J Viera; Benjamin M Bluml; William B White; Sandra J Taler; Steven Yarows; Daichi Shimbo; Raymond R Townsend Journal: Hypertension Date: 2019-02 Impact factor: 10.190