Validation of a smartphone auscultatory blood pressure kit Accutension XYZ-110 in adults according to the ANSI/AAMI/ISO 81060-2: 2013 standard.

Guang Chu1, Zhi Zhang, Mengdan Xu, Daini Huang, Qiuyan Dai.   

Abstract

OBJECTIVE: The aim of this study was to validate the accuracy of the Accutension XYZ-110 blood pressure (BP) kit according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization (ANSI/AAMI/ISO) 81060-2:2013 standard. PARTICIPANTS AND METHODS: Systolic blood pressure (SBP) and diastolic blood pressure (DBP) were measured simultaneously on the same arm in 85 Chinese adults (female : male=48 : 37) with a mean age of 43.2 years using the mercury sphygmomanometer (two observers) and the Accutension XYZ-110 device (one supervisor). The ANSI/AAMI/ISO 81060-2:2013 standard for the validation of BP-measuring devices in adults was followed precisely. A total of 255 comparison pairs were obtained for analysis.
RESULTS: The mean device-observer difference in the 255 separate BP data pairs was 2.45±2.24 mmHg for SBP and 0.69±2.09 mmHg for DBP. The data were in accordance with the criterion 1 of the ANSI/AAMI/ISO 81060-2:2013 standard requirements (≤5±8 mmHg). In addition, the mean device-observer difference of the 85 participants was 2.45±1.47 mmHg for SBP and 0.69±1.36 mmHg for DBP. The device accuracy also fulfilled the criterion 2 with the SD of less than or equal to 6.47 for SBP and less than or equal to 6.90 mmHg for DBP.
CONCLUSION: The Accutension XYZ-110 BP kit fulfilled the requirements of the ANSI/AAMI/ISO 81060-2:2013 standard, and hence could be recommended for both clinical and self/home BP measurement in adults.

Entities:  

Mesh:

Year:  2017        PMID: 28505014     DOI: 10.1097/MBP.0000000000000268

Source DB:  PubMed          Journal:  Blood Press Monit        ISSN: 1359-5237            Impact factor:   1.444


  6 in total

1.  History and Justification of a National Blood Pressure Measurement Validated Device Listing.

Authors:  Jordana B Cohen; Raj S Padwal; Michael Gutkin; Beverly B Green; Michael J Bloch; F Wilford Germino; Domenic A Sica; Anthony J Viera; Benjamin M Bluml; William B White; Sandra J Taler; Steven Yarows; Daichi Shimbo; Raymond R Townsend
Journal:  Hypertension       Date:  2019-02       Impact factor: 10.190

2.  A convenient method to verify the accuracy of oscillometric blood pressure monitors by the auscultatory method: A smartphone-based app.

Authors:  Zhi Zhang; Weichun Xi; Bingjiang Wang; Guang Chu; Fang Wang
Journal:  J Clin Hypertens (Greenwich)       Date:  2018-12-20       Impact factor: 3.738

3.  How to ensure personalized accuracy in home blood pressure devices: Should we play it by ear?

Authors:  Jennifer Ringrose; Raj Padwal
Journal:  J Clin Hypertens (Greenwich)       Date:  2018-12-20       Impact factor: 3.738

4.  The Accutension Stetho, an automated auscultatory device to validate automated sphygmomanometer readings in individual patients.

Authors:  Bruce S Alpert
Journal:  J Hum Hypertens       Date:  2018-04-06       Impact factor: 3.012

Review 5.  Applications of artificial intelligence for hypertension management.

Authors:  Kelvin Tsoi; Karen Yiu; Helen Lee; Hao-Min Cheng; Tzung-Dau Wang; Jam-Chin Tay; Boon Wee Teo; Yuda Turana; Arieska Ann Soenarta; Guru Prasad Sogunuru; Saulat Siddique; Yook-Chin Chia; Jinho Shin; Chen-Huan Chen; Ji-Guang Wang; Kazuomi Kario
Journal:  J Clin Hypertens (Greenwich)       Date:  2021-02-03       Impact factor: 3.738

6.  Lung Auscultation of Hospitalized Patients with SARS-CoV-2 Pneumonia via a Wireless Stethoscope.

Authors:  Pengyu Zhang; Bingjian Wang; Yan Liu; Muge Fan; Yong Ji; Hao Xu; Mengdan Xu; Songwen Chen; Qing Li; Zhi Zhang
Journal:  Int J Med Sci       Date:  2021-01-28       Impact factor: 3.738

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.