Damien Offner1,2,3, Florence Fioretti1,2,3, Anne-Marie Musset2,3.
Abstract
OBJECTIVES: To assess the efficacy of three continuous water disinfection systems for dental units under real conditions of dental care. DESIGN AND SETTINGS: A prospective study carried out from 45 days to 20 months on the water microbial quality of the dental units is benefited from three different systems: two chemical treatment systems (IGN EVO/Calbenium/IGN Cartridge and Sterispray) and one physical treatment system (BacTerminator). Studied items were six dental units of the Dental Medicine and Oral Surgery Center within the University Hospital of Strasbourg (HUS), France. RESULTS AND DISUCUSSION: The IGN EVO/Calbenium/IGN Cartridge and Sterispray systems showed an immediate and long-term efficacy on contaminated dental unit waterlines. However, the first system offers ergonomic advantages (automatic system, action on the water from the water supply network). The BacTerminator system took longer to be effective and was less effective than the other two.Entities:
Year: 2016 PMID: 29607068 PMCID: PMC5842837 DOI: 10.1038/bdjopen.2016.7
Source DB: PubMed Journal: BDJ Open ISSN: 2056-807X
Water treatment systems assigned to the units groups
| System | IGN EVO Calbenium | Sterispray | BacTerminator |
| Type of water | Water of hospital quality | Distilled water | Water of hospital quality |
| Dosing system | Automatic | Manual | No dosage |
Culture conditions of the output water samples collected from the unit and standards used[11]
| HPC at 22 °C | 1 ml | 12 | 5±3 | 72 h at 22 °C on agar PCA by inclusion or filtration | ISO 6222[ |
| HPC at 36 °C | 1 ml | 12 | 5±3 | 48 h at 36 °C on agar PCA by inclusion or filtration | ISO 6222[ |
| Coliform bacteria and | 100 ml | 6 | Ambient (⩽25 °C) | 24 h at 36 °C on agar TTC by membrane filtration | ISO 9308-1[ |
| 100 ml | 24 | 5±3 | 2nd inspection after 48 h | ||
| Enterococci | 100 ml | 6 | Ambient (⩽25 °C) | 48 h at 36 °C on Slanetz agar by membrane filtration | ISO 7899-2[ |
| 100 ml | 24 | 5±3 | |||
| 100 ml | Undisclosed | Undisclosed | 48 h at 36 °C on cetrimide agar by membrane filtration | ISO 16266[ |
Abbreviations: HPC, heterotrophic plate count; PCA, plate count agar=agar used to count revivifiable aerobic mircroorganisms; TTC, tergitol=medium used for the search and count of coliform bacteria.
Levels of microbiological water quality and interpretation of the output water sampling results from the unit[11]
| • HPC at 22 °C ⩽100 c.f.u./ml • HPC at 37 °C ⩽10 c.f.u./ml • Absence of pathogenic germs | • Results COMPLIANT with the expected values |
| • HPC at 22 °C >100 u.f.c./ml and <200 u.f.c./ml • HPC at 27 °C >10 u.f.c./ml and <30 u.f.c./ml • Absence of pathogenic germs | • ACCEPTABLE results The unit can be used. A new sampling procedure is scheduled at D+3 |
| • HPC at 22 °C | • NON COMPLIANT results SUSPEND THE USE of the unit and initiate Dialox treatment. A new sampling is scheduled at D+15 after Dialox treatment, until results are compliant |
Abbreviation: HPC, heterotrophic plate count.
Figure 1Microbiological monitoring of the output water from units 1 and 2 equipped with the IGN EVO Calbenium system.
Figure 2Microbiological monitoring of the output water from units 3 and 4 equipped with the Sterispray.
Figure 3Microbiological monitoring of the output water from unit 5 equipped with the BacTerminator system, showing the correction of technical defects after samples collection at D+ 15.
Figure 4Microbiological monitoring of the output water from unit 6 equipped with the BacTerminator system.