Literature DB >> 29604095

Performance evaluation of a chemiluminescence microparticle immunoassay for CK-MB.

Zhi-Yuan Lin1,2, Yi-Zhen Fang1, Hong-Wei Jin1, Hua-Yue Lin1, Zhang Dai1, Qing Luo1, Hong-Wei Li1, Yan-Ling Lin1, Shui-Zhen Huang1, Lei Gao3, Fei-Hai Xu3, Zhong-Ying Zhang1.   

Abstract

BACKGROUND: To verify and evaluate the performance characteristics of a creatine kinase phosphokinase isoenzymes MB (CK-MB) assay kit, which produced by Xiamen Innodx Biotech Co. Ltd.
METHODS: Evaluation was carried out according to "Guidelines for principle of analysis performance evaluation of in vitro diagnostic reagent." The performance parameters included detection limit, linearity range, reportable range, recovery test, precision verification, interference test, cross-reactivity, matrix effect, and method comparison.
RESULTS: The detection limit was 0.1 ng/mL. The assay had clinical linearity over range of 0.1 ng/mL-500 ng/mL. Reportable range was from 0.1 ng/mL to 1000 ng/mL. The average percent of recovery was 99.66%. The coefficient of variation (CV) for within-run and between-run of low CK-MB sample was 5.55% and 6.16%, respectively. As for high-level sample, it was 7.88% and 7.80%. In medical decision level, the relative deviation (Bias) of all interference tests was lower than 15%. When the sample had mild-hemolysis; hemoglobin ≤15 g/L; triglyceride ≤17 mmol/L; bilirubin ≤427.5 μmol/L; rheumatoid factor ≤206U/mL, there was no significant interference to be found. Moreover, assay kit had no cross-reaction with CK-MM and CK-BB. At last, total diagnostic accuracy of kit was 93.24%, when compared with refer kit.
CONCLUSION: Overall the results of the verification study indicated the performance of kit is met the requirements of the clinical test.
© 2018 Wiley Periodicals, Inc.

Entities:  

Keywords:  CK-MB; chemiluminescence microparticle immunoassay; evaluation; kit; performance

Year:  2018        PMID: 29604095      PMCID: PMC6816894          DOI: 10.1002/jcla.22426

Source DB:  PubMed          Journal:  J Clin Lab Anal        ISSN: 0887-8013            Impact factor:   2.352


  10 in total

1.  [Serum creatine kinase in the diagnosis of myocardial infarct].

Authors:  J C DREYFUS; G SCHAPIRA; J RESNAIS; L SCEBAT
Journal:  Rev Fr Etud Clin Biol       Date:  1960-04

2.  A spectrophotometric method for the determination of creatine phosphokinase and myokinase.

Authors:  I T OLIVER
Journal:  Biochem J       Date:  1955-09       Impact factor: 3.857

3.  Clinical outcomes after detection of elevated cardiac enzymes in patients undergoing percutaneous intervention. IMPACT-II Investigators. Integrilin (eptifibatide) to Minimize Platelet Aggregation and Coronary Thrombosis-II.

Authors:  B E Tardiff; R M Califf; J E Tcheng; A M Lincoff; K N Sigmon; R A Harrington; K W Mahaffey; E M Ohman; P S Teirstein; J C Blankenship; M M Kitt; E J Topol
Journal:  J Am Coll Cardiol       Date:  1999-01       Impact factor: 24.094

4.  Mass concentration and activity concentration of creatine kinase isoenzyme MB compared in serum after acute myocardial infarction.

Authors:  J R Delanghe; A M De Mol; M L De Buyzere; I K De Scheerder; R J Wieme
Journal:  Clin Chem       Date:  1990-01       Impact factor: 8.327

5.  Performance evaluation of a chemiluminescence microparticle immunoassay for CK-MB.

Authors:  Zhi-Yuan Lin; Yi-Zhen Fang; Hong-Wei Jin; Hua-Yue Lin; Zhang Dai; Qing Luo; Hong-Wei Li; Yan-Ling Lin; Shui-Zhen Huang; Lei Gao; Fei-Hai Xu; Zhong-Ying Zhang
Journal:  J Clin Lab Anal       Date:  2018-03-31       Impact factor: 2.352

6.  Determination of CKMB activity and protein concentration and their application in the diagnosis of AMI.

Authors:  Y Song; J Zhong; L Zhao; T Shaoji
Journal:  J Tongji Med Univ       Date:  2000

Review 7.  Enzyme and muscle diseases.

Authors:  M Panteghini
Journal:  Curr Opin Rheumatol       Date:  1995-11       Impact factor: 5.006

Review 8.  An historical approach to the diagnostic biomarkers of acute coronary syndrome.

Authors:  Elisa Danese; Martina Montagnana
Journal:  Ann Transl Med       Date:  2016-05

Review 9.  Role of creatine kinase isoenzymes on muscular and cardiorespiratory endurance: genetic and molecular evidence.

Authors:  M Echegaray; M A Rivera
Journal:  Sports Med       Date:  2001       Impact factor: 11.928

10.  The diagnostic value of both troponin T and creatinine kinase isoenzyme (CK-MB) in detecting combined renal and myocardial injuries in asphyxiated infants.

Authors:  Wilson E Sadoh; Charles O Eregie; Damian U Nwaneri; Ayebo E Sadoh
Journal:  PLoS One       Date:  2014-03-13       Impact factor: 3.240

  10 in total
  3 in total

1.  Performance evaluation of a chemiluminescence microparticle immunoassay for CK-MB.

Authors:  Zhi-Yuan Lin; Yi-Zhen Fang; Hong-Wei Jin; Hua-Yue Lin; Zhang Dai; Qing Luo; Hong-Wei Li; Yan-Ling Lin; Shui-Zhen Huang; Lei Gao; Fei-Hai Xu; Zhong-Ying Zhang
Journal:  J Clin Lab Anal       Date:  2018-03-31       Impact factor: 2.352

2.  On the Challenges for the Diagnosis of SARS-CoV-2 Based on a Review of Current Methodologies.

Authors:  Isabela A Mattioli; Ayaz Hassan; Osvaldo N Oliveira; Frank N Crespilho
Journal:  ACS Sens       Date:  2020-12-03       Impact factor: 7.711

3.  Establishment of a new protein C detection system based on chromogenic substrate assay and its clinical diagnostic value for deep vein thrombosis.

Authors:  Wenfei Lu; Anshan Ge; Jinxia Zhao; Hui Rong; Chao Zhu; Limin Lun
Journal:  J Clin Lab Anal       Date:  2021-11-13       Impact factor: 2.352

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.