Attenuation values of renal parenchyma in virtual noncontrast images acquired from multiphase renal dual-energy CT: Comparison with standard noncontrast CT.

Yuan-Mao Lin1, Yi-You Chiou1, Mei-Han Wu1, Shan-Su Huang2, Shu-Huei Shen3.   

Abstract

OBJECTIVES: To compare the renal parenchyma attenuation of virtual noncontrast (VNC) images derived from multiphase renal dual-energy computed tomography (DECT) with standard noncontrast (SNC) images, and to determine the optimum phase for VNC images.
MATERIALS AND METHODS: Twenty-nine men and 16 women (mean age, 61 ± 13 years; range, 37-89 years) underwent dynamic renal DECT (100/Sn140 kVp) were included in this institutional review board-approved retrospective study. There were four phases of the scan, which included noncontrast, corticomedullary (CMP), nephrographic (NP), and excretory phases (EP). The VNC images was generated from CMP, NP and EP. CT numbers of SNC images and VNC images of each phases were measured in the renal cortex and medulla. Mean standard deviation of subcutaneous fat was measured as image noise on SNC and VNC images. Radiation dose was recorded and potential radiation dose reduction was estimated. Results were tested for statistical significance using the unpaired t-test and agreement using Bland-Altman plot analysis.
RESULTS: The difference in mean attenuation between SNC and each phase of VNC images were ≤4 HU. The mean attenuation of renal cortex and medulla was 33.2 ± 4.4 HU, and 34.2 ± 4.8 HU in SNC, 33.6 ± 7.6 HU and 31.1 ± 8.3 HU in VNC of CMP, 34.8 ± 8.6 HU and 35.6 ± 8.5 HU in VNC of NP, 31.5 ± 7.6 HU and 32.4 ± 7.5 HU in VNC of EP. In VNC of CMP, the attenuation of the cortex was higher than the medulla (p < 0.05), and the attenuation of medulla was significant lower than that of SNC (p < 0.01). In VNC of NP, the attenuation of renal cortex was higher than SNC (p < 0.05). In VNC of EP, the attenuation of cortex and medulla were lower than SNC (p < 0.05), and inadequate iodine subtraction in collecting system was noted. Image noise was significantly greater in SNC (p < 0.001). Mean radiation dose reduction achievable by removing the SNC was 12.3% ± 0.9%.
CONCLUSIONS: VNC images from multiphase renal DECT were similar to SNC images. Using the nephrographic phase can gives more comparable VNC images to SNC images in renal parenchyma than other phases.
Copyright © 2018 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Attenuation value; Dual-energy computed tomography; Multiphase renal CT; Radiation dose; Standard noncontrast; Virtual noncontrast

Mesh:

Year:  2018        PMID: 29571782     DOI: 10.1016/j.ejrad.2018.02.001

Source DB:  PubMed          Journal:  Eur J Radiol        ISSN: 0720-048X            Impact factor:   3.528


  6 in total

1.  Compatibility of true and virtual unenhanced attenuation in rapid kV-switching dual energy CT.

Authors:  İlkay Çamlıdağ
Journal:  Diagn Interv Radiol       Date:  2020-03       Impact factor: 2.630

2.  Dual-energy CT arthrography: a feasibility study.

Authors:  Rashpal Sandhu; Mercan Aslan; Nancy Obuchowski; Andrew Primak; Wadih Karim; Naveen Subhas
Journal:  Skeletal Radiol       Date:  2020-09-18       Impact factor: 2.199

3.  The Delayed Nephrogram: Point-of-care Quantitative Measurement, Validation as an Indicator of Obstruction, and Novel Use as a Predictor of Renal Functional Impairment.

Authors:  Marshall C Strother; Eric Y Cho; Matt Loecher; David Strauss; Akhil Chandra; Elizabeth Handorf; Jian Yu; David Y T Chen; Robert Uzzo; Laura Levin; Jordan Anaokar; Alexander Kutikov
Journal:  Eur Urol Focus       Date:  2022-02-15

4.  Dual-Energy CT Images: Pearls and Pitfalls.

Authors:  Anushri Parakh; Simon Lennartz; Chansik An; Prabhakar Rajiah; Benjamin M Yeh; Frank J Simeone; Dushyant V Sahani; Avinash R Kambadakone
Journal:  Radiographics       Date:  2021 Jan-Feb       Impact factor: 5.333

5.  Quantitative accuracy of virtual non-contrast images derived from spectral detector computed tomography: an abdominal phantom study.

Authors:  Jasmin A Holz; Hatem Alkadhi; Kai R Laukamp; Simon Lennartz; Carola Heneweer; Michael Püsken; Thorsten Persigehl; David Maintz; Nils Große Hokamp
Journal:  Sci Rep       Date:  2020-12-09       Impact factor: 4.379

6.  Diagnostic performance of single-phase dual-energy CT to differentiate vascular and nonvascular incidental renal lesions on portal venous phase: comparison with CT.

Authors:  Domenico Mastrodicasa; Martin J Willemink; Nikhil Madhuripan; Ranjit Singh Chima; Amanzo A Ho; Yuqin Ding; Daniele Marin; Bhavik N Patel
Journal:  Eur Radiol       Date:  2021-06-11       Impact factor: 7.034

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.