| Literature DB >> 29549265 |
Mel O'Leary1,2, Daria Boscolo3, Nicole Breslin4,5, Jeremy M C Brown4,6, Igor P Dolbnya7, Chris Emerson4,5, Catarina Figueira4,5, Oliver J L Fox7, David Robert Grimes4,5,8, Vladimir Ivosev4,9, Annette K Kleppe7, Aaron McCulloch4,5, Ian Pape7, Chris Polin4, Nathan Wardlow4, Fred J Currell4,5.
Abstract
Absolute measurements of the radiolytic yield of Fe3+ in a ferrous sulphate dosimeter formulation (6 mM Fe2+), with a 20 keV x-ray monoenergetic beam, are reported. Dose-rate suppression of the radiolytic yield was observed at dose rates lower than and different in nature to those previously reported with x-rays. We present evidence that this effect is most likely to be due to recombination of free radicals radiolytically produced from water. The method used to make these measurements is also new and it provides radiolytic yields which are directly traceable to the SI standards system. The data presented provides new and exacting tests of radiation chemistry codes.Entities:
Year: 2018 PMID: 29549265 PMCID: PMC5856745 DOI: 10.1038/s41598-018-21813-z
Source DB: PubMed Journal: Sci Rep ISSN: 2045-2322 Impact factor: 4.379
Figure 1Total least squares fit of the data. Part A showing the best linear fit plotted (c = 1.51 ± 0.06 μmol/J and m = −0.37 ± 0.09 (nmol/J)/(Gy/s), as indicated on the graph) against the data and error bars (plotted as ellipses). In part B, there is a contour plot of the likelihood that a fit fits the data. Each contour is about a standard deviation further away from the best fit, which is indicated by the point in the center of the contours[21]. The intercept of the best fit line agrees well with the model by Yamaguchi[22]. The blue line shows the predicted change in G-value due to spectral hardening effects. With an intercept at 1.498 μmol/J and a slope of −2.11 × 10−2 (nmol/J)/(Gy/s), i.e. two orders of magnitude smaller than the observed effect.
Figure 2Measured amount of Fe3+ as a function of radiation energy absorbed by the sample. The slope of the line of best fit through the data is the G-value (1.42 ± 0.08 μmol/J). Error bars (indicated by ellipses) are ± two standard errors and indicate statistical errors alone. These measurements were taken with a dose rate of 150 ± 20 Gy/s.
Figure 3Measured amount of Fe3+ as a function of radiation energy absorbed by the sample; showing the turnover in production from the depletion of oxygen. Error bars are ± two standard errors and indicate statistical errors alone. These measurements were taken on B16[29,31] with 16 keV X-rays (monoenergetic beam) at a dose rate of 575 ± 17 Gy/s.
Figure 4(a) Photograph of the chamber (b) Sketch of the chamber (c) An exploded view of the chamber showing its component parts[1]: front block[2], kapton windows[3], o-rings[4], the chamber block[5], Inlet PEEK tubing[6], Outlet PEEK tubing[7], Back block that is connected to supporting structure.
Figure 5Schematic diagram of the set up on the beam line, after the monochromator and beam collimation system.