OBJECTIVE: This study aimed to investigate the influence of injection rates of calibrating standard solution on monitoring pulse indicator continuous cardiac output (PICCO, made in Germany), and thereby to provide significant references for clinical practice.Entities:
Keywords: Calibrating standard solution; Injection rate; Pulse indicator continuous cardiac output
Mesh:
Substances:
Year: 2018 PMID: 29548292 PMCID: PMC5857084 DOI: 10.1186/s12938-018-0453-0
Source DB: PubMed Journal: Biomed Eng Online ISSN: 1475-925X Impact factor: 2.819
Comparison of PICCO monitor parameters at the drip speed of 2–4, 5–7, and 8–10 s
| 2–4 s | 5–7 s | 8–10 s | Normal value | F |
| |
|---|---|---|---|---|---|---|
| GEF | 30.35 ± 4.87 | 27.47 ± 5.47 | 23.28 ± 5.32 | 25–35 | 49.113 | < 0.001 |
| GEDI | 713.36 ± 73.67 | 585.75 ± 126.94 | 545.53 ± 140.37 | 680–800 | 58.660 | < 0.001 |
| CI | 4.61 ± 1.28 | 4.32 ± 1.87 | 4.02 ± 1.81 | 3.0–5.0 | 3.088 | = 0.047 |
Whole body ejection fraction (GEF), systemic end-diastolic volume (GEDI), cardiac index (CI), which were all consistent with color Doppler surveillance
| Color ultrasonic diagnosis | KAPPA |
| ||||
|---|---|---|---|---|---|---|
| Normal | Abnormal | |||||
| Injection rate 2–4 s | GEF | Normal | 78 | 0 | 0.771 | < 0.001 |
| Abnormal | 9 | 21 | ||||
| GEDI | Normal | 50 | 2 | 0.907 | < 0.001 | |
| Abnormal | 3 | 53 | ||||
| CI | Normal | 64 | 1 | 0.761 | < 0.001 | |
| Abnormal | 9 | 34 | ||||
| Injection rate 5–7 s | GEF | Normal | 60 | 0 | 0.464 | 0.048 |
| Abnormal | 27 | 21 | ||||
| GEDI | Normal | 17 | 6 | 0.214 | 0.007 | |
| Abnormal | 36 | 49 | ||||
| CI | Normal | 39 | 0 | 0.426 | < 0.001 | |
| Abnormal | 34 | 35 | ||||
| Injection rate 8–10 s | GEF | Normal | 19 | 7 | − 0.055 | 0.269 |
| Abnormal | 68 | 14 | ||||
| GEDI | Normal | 7 | 4 | 0.06 | 0.308 | |
| Abnormal | 46 | 51 | ||||
| CI | Normal | 12 | 0 | 0.113 | 0.011 | |
| Abnormal | 61 | 35 | ||||
Fig. 1Correlation analysis between transesophageal heart color ultrasound monitor and PICCO parameters. 1a Monitoring data of global ejection fraction (GEF) in 2–4 s; 1b Monitoring data of cardiac index (CI) in 2–4 s; 1c Monitoring data of global end diastolic volume index (GEDI) in 2–4 s; 2a Monitoring data of global ejection fraction (GEF) in 5–7 s; 2b Monitoring data of cardiac index (CI) in 5–7 s; 2c Monitoring data of global end diastolic volume index (GEDI) in 5–7 s; 3a Monitoring data of global ejection fraction (GEF) in 8–10 s; 3b Monitoring data of cardiac index (CI) in 8–10 s; 3c Monitoring data of global end diastolic volume index (GEDI) in 8–10 s