Assessment of intrathoracic blood volume as an indicator of cardiac preload: single transpulmonary thermodilution technique versus assessment of pressure preload parameters derived from a pulmonary artery catheter.

C Wiesenack1, C Prasser, C Keyl, G Rödīg.   

Abstract

OBJECTIVE: To analyze the clinical value of a new device (PiCCO) for cardiac output measurement and volume preload parameter assessment, based on transpulmonary thermodilution technique, as an alternative to the pulmonary artery thermodilution technique and assessment of pressure preload parameters derived from the pulmonary artery catheter.
DESIGN: Prospective, controlled, clinical study.
SETTING: University hospital. PARTICIPANTS: Eighteen patients with ejection fraction >50% undergoing coronary artery bypass graft surgery.
INTERVENTIONS: A baseline measurement was performed after induction of anesthesia under clinical steady-state conditions (T1). Hypovolemia, defined as central venous pressure (CVP) <10 mmHg and pulmonary capillary wedge pressure (PCWP) <12 mmHg, was treated by infusion of 6% hydroxyethyl starch 200/0.5 (7 mL/kg). After 10 minutes, a second measurement (T2) was performed.
MEASUREMENTS AND MAIN RESULTS: The mean difference (bias) between transpulmonary thermodilution cardiac output and pulmonary artery thermodilution cardiac output did not differ at the 2 sample points. Changes in pressure preload parameters of the pulmonary artery catheter (CVP, PCWP) did not correlate with changes in cardiac output or stroke volume, whereas changes in volume preload parameter intrathoracic blood volume (ITBV) of the PiCCO correlated significantly with changes in cardiac output and stroke volume (r = 0.55, p < 0.05; r = 0.62, p < 0.01).
CONCLUSION: These results suggest that increased cardiac preload is more reliably reflected by ITBV than by CVP or PCWP. The assessment of ITBV by the transpulmonary single indicator dilution technique is an interesting alternative to the pressure preload parameters. Copyright 2001 by W.B. Saunders Company

Entities:  

Mesh:

Year:  2001        PMID: 11687999     DOI: 10.1053/jcan.2001.26536

Source DB:  PubMed          Journal:  J Cardiothorac Vasc Anesth        ISSN: 1053-0770            Impact factor:   2.628


  31 in total

1.  Calculated global end-diastolic volume does not correspond to the largest heart blood volume: a bias for cardiac function index?

Authors:  François Gilles Brivet; Frédéric Jacobs; Patrice Colin
Journal:  Intensive Care Med       Date:  2004-09-11       Impact factor: 17.440

Review 2.  Reproducibility of transpulmonary thermodilution cardiac output measurements in clinical practice: a systematic review.

Authors:  Raphaël Giraud; Nils Siegenthaler; Paolo Merlani; Karim Bendjelid
Journal:  J Clin Monit Comput       Date:  2016-01-11       Impact factor: 2.502

Review 3.  Methods in pharmacology: measurement of cardiac output.

Authors:  Bart F Geerts; Leon P Aarts; Jos R Jansen
Journal:  Br J Clin Pharmacol       Date:  2011-03       Impact factor: 4.335

4.  Comparison of the accuracy of transpulmonary thermodilution measurement using indicators of different temperatures.

Authors:  Shulan Chen; Pingdong Lin; Zhenshuang Du; Fangchen Lan; Shanshan Wu; Tiegang Zhong; Xiaohua Liang; Hongyu Liu; Cuiping Zeng; Chenghua Zhang
Journal:  Int J Clin Exp Med       Date:  2014-12-15

Review 5.  Pulse waveform hemodynamic monitoring devices: recent advances and the place in goal-directed therapy in cardiac surgical patients.

Authors:  Adham Hendy; Şerban Bubenek
Journal:  Rom J Anaesth Intensive Care       Date:  2016-04

6.  Effects of high-frequency oscillatory ventilation on systemic and cerebral hemodynamics and tissue oxygenation: an experimental study in pigs.

Authors:  Jan Florian Heuer; Philip Sauter; Jürgen Barwing; Peter Herrmann; Thomas A Crozier; Annalen Bleckmann; Tim Beißbarth; Onnen Moerer; Michael Quintel
Journal:  Neurocrit Care       Date:  2012-10       Impact factor: 3.210

Review 7.  Clinical review: practical recommendations on the management of perioperative heart failure in cardiac surgery.

Authors:  Alexandre Mebazaa; Antonis A Pitsis; Alain Rudiger; Wolfgang Toller; Dan Longrois; Sven-Erik Ricksten; Ilona Bobek; Stefan De Hert; Georg Wieselthaler; Uwe Schirmer; Ludwig K von Segesser; Michael Sander; Don Poldermans; Marco Ranucci; Peter C J Karpati; Patrick Wouters; Manfred Seeberger; Edith R Schmid; Walter Weder; Ferenc Follath
Journal:  Crit Care       Date:  2010-04-28       Impact factor: 9.097

8.  Estimation of left ventricular systolic function by single transpulmonary thermodilution.

Authors:  Alain Combes; Jean-Baptiste Berneau; Charles-Edouard Luyt; Jean-Louis Trouillet
Journal:  Intensive Care Med       Date:  2004-04-23       Impact factor: 17.440

Review 9.  Monitoring cardiac function in intensive care.

Authors:  S M Tibby; I A Murdoch
Journal:  Arch Dis Child       Date:  2003-01       Impact factor: 3.791

Review 10.  Clinical review: Goal-directed therapy in high risk surgical patients.

Authors:  Nicholas Lees; Mark Hamilton; Andrew Rhodes
Journal:  Crit Care       Date:  2009-10-26       Impact factor: 9.097