Mamadou Alpha Diallo1, Khadim Diongue2, Mame Cheikh Seck2, Mouhamadou Ndiaye2, Ibrahima Diallo3, Younouss Diedhiou2, Tolla Ndiaye2, Yaye Die Ndiaye2, Aida Sadikh Badiane2, Daouda Ndiaye2.
Abstract
BACKGROUND: In developing countries, malaria diagnosis relies on microscopy and rapid diagnostic tests. In Senegal, national malaria control program (NMCP) regularly conducts supervisory visits in health services where malaria microscopy is performed. In this study, expert microscopists assessed the performance of laboratory technicians in malaria microscopy.Entities:
Keywords: Diagnosis; EQA; External quality assessment; Malaria; Microscopy; Senegal
Mesh:
Year: 2018 PMID: 29544479 PMCID: PMC5853095 DOI: 10.1186/s12941-018-0261-1
Source DB: PubMed Journal: Ann Clin Microbiol Antimicrob ISSN: 1476-0711 Impact factor: 3.944
Fig. 1Malaria epidemiological strata with sites where supervision was conducted; 7 hospital laboratories: Dakar (3), Thies (1) Saint Louis (2), Kedougou (1); 11 health centre laboratories: Dakar (2), Thies (2), Saint Louis (4), Kedougou (3)
Fig. 2Trophozoites of P. falciparum with specific Maurer dots inside the infected red blood cells, Giemsa stained thin smear (×1000)
Reported errors from participants on panel testing slides
| Participants | Slide 1 | Slide 2 | Slide 3 | Slide 4–8 | Total errors |
|---|---|---|---|---|---|
| Hospital 1 | Negative |
| 2 | ||
| Hospital 5 |
| 1 | |||
| HC 1 | Negative |
| 2 | ||
| HC 5 |
| 1 | |||
| HC 8 | Negative |
|
| 3 | |
| HC 10 | Negative | 1 | |||
| Total errors | 4 | 3 | 0 | 3 | 10 |
Slide 1: P. falciparum, 192 p/μl
Slide 2: P. falciparum, 28,000 parasites with Maurer dots (see Fig. 1)
Slide 3: P. falciparum, with both trophozoites and gametocytes
Slides 4,5,6,7 and 8: No parasites
Errors reported during cross-checking of read slides
| Participants laboratories | Errors types | Total errors | ||||
|---|---|---|---|---|---|---|
| HFP | HFN | LFP | LFNa | Misidentificationa | ||
| Hospital 1 | 0 | 0 | 1 | 1 | 3 | 5 |
| Hospital 2 | 0 | 0 | 0 | 1 | 0 | 1 |
| Hospital 3 | 0 | 0 | 0 | 0 | 0 | |
| Hospital 4 | 0 | 0 | 0 | 1 | 1 | 2 |
| Hospital 5 | 0 | 0 | 0 | 0 | 0 | |
| Hospital 6 | 0 | 0 | 0 | 0 | 0 | |
| HC 1 | 0 | 0 | 0 | 0 | 0 | |
| HC 2 | 2 | 0 | 0 | 0 | 0 | 2 |
| HC 3 | 0 | 0 | 0 | 0 | 0 | |
| HC 4 | 0 | 0 | 0 | 0 | 0 | |
| HC 5 | 0 | 0 | 1 | 0 | 0 | 1 |
| HC 6 | 0 | 0 | 0 | 0 | 1 | 1 |
| HC 7 | 0 | 0 | 0 | 0 | 0 | |
| HC 8 | 0 | 0 | 0 | 0 | 0 | |
| HC 9 |
|
|
|
|
| |
| HC 10 |
|
|
|
|
| |
| HC 11 | 0 | 0 | 0 | 2 | 0 | 2 |
HC9 and HC10 were not involved as they could not store the required number of slides
aSee details on Table 3
Results of missed slides and misidentified slides when controlled by expert microscopists
| Results of participants | Results of controllers | |
|---|---|---|
| Misidentification | ||
| Hospital 1 |
|
|
| Hospital 4 |
|
|
| LFN | ||
| Hospital 1 | Negative | Spirochetes |
| Hospital 2 | Negative | Spirochetes |
| Hospital 4 | Negative | |
| HC 4 | Negative |
|
| HC 11 | Negative |
|
| HC 11 | Negative | |
Fig. 3P. falciparum trophozoite presenting an equatorial-like band in Giemsa stained thin smear (×1000)
Sensitivities, specificities, PPV and NPV of participant microscopists against expert microscopists
| Controllers | Total | |||
|---|---|---|---|---|
| Positive | Negative | |||
| Hospital technicians | Positive | 89 | 1 | 90 |
| Negative | 3 | 87 | 90 | |
| Total | 92 | 88 | 180 | |
| Sensitivity (%) | 96.7 | |||
| Specificity (%) | 98.9 | |||
| PPV (%) | 98.9 | |||
| NPV (%) | 96.7 | |||
| Health centers technicians | Positive | 132 | 3 | 135 |
| Negative | 2 | 133 | 135 | |
| Total | 134 | 136 | 270 | |
| Sensitivity (%) | 98.5 | |||
| Specificity (%) | 97.8 | |||
| PPV (%) | 97.8 | |||
| NPV (%) |
| |||