Comparison of Central Corneal Thickness with Ultrasound Pachymetry, Noncontact Specular Microscopy and Spectral Domain Optical Coherence Tomography.

Sevil Karaman Erdur1, Goktug Demirci1, Funda Dikkaya1, Mehmet Selim Kocabora1, Mustafa Ozsutcu1.   

Abstract

BACKGROUND: To determine the agreement of central corneal thickness (CCT) measurements taken with ultrasonic pachymetry (USP), spectral domain optical coherence tomography (SD-OCT) and noncontact specular microscopy (NSM).
METHODS: A prospective, observational, cross-sectional study was performed in the outpatient ophthalmology clinic. CCT was measured in a total of 147 eyes of 147 consecutive healthy patients with USP, NSM, and SD-OCT. Same examiner performed all examinations. Bland-Altman plots were used to evaluate the agreement between instruments.
RESULTS: The average CCT values obtained by USP, NSM, and SD-OCT were 555 ± 37 µm, 554 ± 34 µm, and 546 ± 34 µm, respectively. There was a strong correlation between instruments: USP with SD-OCT (r = 0.937, p < 0.01), USP with NSM (r = 0.943, p < 0.01) SD-OCT with NSM (r = 0.975, p < 0.01) for CCT. The mean differences (lower/upper limit of agreement) for CCT measurements were -10 ± 12.9 µm (15.28/-35.28) between SD-OCT and USP, -8.1 ± 7.7 µm (7/-23.2) between SD-OCT and NSM, and 1.8 ± 12.3 µm (25.9/-22.3) between USP and NSM.
CONCLUSIONS: USP and NSM were found to have comparable CCT measurements and these two methods can be used correspondingly. However, CCT measurements by SD-OCT were lower when compared to other methods.

Entities:  

Keywords:  Corneal thickness; glaucoma; optical coherence tomography; pachymetry; specular microscopy

Mesh:

Year:  2018        PMID: 29509052     DOI: 10.1080/08820538.2018.1448091

Source DB:  PubMed          Journal:  Semin Ophthalmol        ISSN: 0882-0538            Impact factor:   1.975


  3 in total

1.  Central Corneal Thickness Measurements with Different Imaging Devices: Ultrasound Pachymetry, Noncontact Specular Microscopy, and Tono-Pachymetry.

Authors:  Ali Ceylan; Irem Onal; Burak Mergen; Yusuf Yildirim
Journal:  Beyoglu Eye J       Date:  2022-02-18

2.  Comparison of Six Methods of Central Corneal Thickness Measurement in Healthy Eyes.

Authors:  Turgay Ucak; Erel Icel; Nurdan Gamze Tasli; Yucel Karakurt; Hayati Yilmaz; Adem Ugurlu; Mehmet Demir
Journal:  Beyoglu Eye J       Date:  2021-02-12

3.  Effects of Type 2 Diabetes Mellitus and Smoking on Changes in Corneal Endothelial Morphology and Cell Density.

Authors:  Marija Antičić-Eichwalder; Susanne Lex; Stephanie Sarny; Jakob Schweighofer; Ivana Marić; Yosuf El-Shabrawi
Journal:  Cornea       Date:  2021-11-22       Impact factor: 3.152

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.