Correlation of size-dependent conversion factor and body-mass-index using size-specific dose estimates formalism in CT examinations.

Babak Alikhani1, Tobias Getzin2, Till Frederik Kaireit2, Kristina Imeen Ringe2, Leila Jamali2, Frank Wacker2, Thomas Werncke2, Hans-Jürgen Raatschen2.   

Abstract

PURPOSE: The aim of this investigation is to establish the relationship between the size-dependent conversion factor (fsize) and the body-mass-index (BMI) and to test whether BMI can be substituted for the conventionally used patients' anterior-posterior (AP) and lateral (LAT) dimensions for calculation of fsize. By calculating fsize on the basis of BMI instead of the AP and LAT measurements, size-specific dose estimates (SSDE) could be determined prior to image acquisition.
METHODS: Our institute utilizes a dose monitoring software to record radiation exposure during CT examinations. The datasets gathered during each examination contain information regarding the scan protocol, the volumetric computed tomography dose index (CTDIvol), SSDE and BMI. fsize is traditionally calculated through measurement of AP and LAT dimensions. In this work, the dose monitoring system calculates AP and LAT diameters at the middle of scout views. For purpose of this investigation, we used data from 13544 patients who underwent CT examinations of the torso, head or knee (both) to compare fsize as calculated from the AP and LAT dimensions to fsize calculated as a function of BMI.
RESULTS: In the examinations of the torso, we observed an exponentially decreasing correlation between fsize and BMI. In the examinations of the head and knee (both), fsize reflected an almost independent behavior to BMI.
CONCLUSION: This study demonstrates that it is possible to estimate fsize by using the patients' BMI for the torso as well as the head and knee CT, thereby enabling calculation of the probable SSDE prior to image acquisition on the basis of the presumed CTDIvol provided by the scanner. By providing information on the expected patient dose prior to image acquisition, this method is advantageous over the traditional calculation of fsize via the AP and LAT dimensions.
Copyright © 2018 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Dose monitoring software; Radiation dose; Size-dependent conversion factor; Size-specific dose estimates

Mesh:

Year:  2018        PMID: 29496070     DOI: 10.1016/j.ejrad.2018.01.027

Source DB:  PubMed          Journal:  Eur J Radiol        ISSN: 0720-048X            Impact factor:   3.528


  3 in total

1.  CAN THE SIZE-SPECIFIC DOSE ESTIMATE BE DERIVED FROM THE BODY MASS INDEX? A FEASIBILITY STUDY.

Authors:  Beatrice Steiniger; Chris Klippel; Ulf Teichgräber; Jürgen R Reichenbach; Martin Fiebich
Journal:  Radiat Prot Dosimetry       Date:  2022-05-13       Impact factor: 0.954

2.  Size-Specific Dose Estimates of Radiation Based on Body Weight and Body Mass Index for Chest and Abdomen-Pelvic CTs.

Authors:  Jian Xu; Xiangquan Wang; Panfeng Yang; Kuangnan Luo; Xiaolong He
Journal:  Biomed Res Int       Date:  2020-07-10       Impact factor: 3.411

3.  Using body mass index to estimate individualised patient radiation dose in abdominal computed tomography.

Authors:  Siobhan O'Neill; Richard G Kavanagh; Brian W Carey; Niamh Moore; Michael Maher; Owen J O'Connor
Journal:  Eur Radiol Exp       Date:  2018-11-28
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.