Vincent Masse1,2, Michael J Hartley3, Michael B Edmond1,4, Daniel J Diekema1,5.
Abstract
Background: Ultraviolet germicidal irradiation (UVGI) systems are gaining popularity, however objective comparisons of their characteristics are lacking. While environmental cultures and reduction of hospital-associated infections rates are excellent study endpoints, they are impractical for centers with limited resources who want to compare or optimize UVGI systems use.Entities:
Keywords: Radiometry; Terminal room disinfection; Test cards; UVGI; Ultraviolet
Mesh:
Substances:
Year: 2018 PMID: 29484174 PMCID: PMC5824448 DOI: 10.1186/s13756-018-0317-1
Source DB: PubMed Journal: Antimicrob Resist Infect Control ISSN: 2047-2994 Impact factor: 4.887
UVGI devices comparison
| Large units for patient room disinfection | Small units for bathrooms and other confined spaces | |||
|---|---|---|---|---|
| Name (manufacturer) | Tru-D (Tru-D SmartUVC, Memphis, TN) | Enlight (Clorox Healthcare - Optimum UV, Oakland, CA) | EDU 435 (Lumalier, Memphis, TN) | Turbo-UV (MRSA-UV, Palm Springs, FL) |
| Workflow and control | • One room, one device placement | • One room, multiple device placement | • Delay start function | • Remote control |
| Available literature | Most studied UVGI system | A few available studies | No studies available | No studies available |
| Price point (approximate, USD) | $90,000 | $45,000 | $4000 | $600 |
Live tests results with the large units
| Area | Power reading (mW/cm2) Tru-D Center 1 h runtime | Energy estimation (mJ) Tru-D Center 1 h runtime | Sum of Power readings (mW/cm2) Enlight 3 spots X 5 min | Energy estimation (mJ) Enlight 3 spots X 5 min | ||
|---|---|---|---|---|---|---|
| BR faucet | 0 | 0 | No change | 0.021 | 6.30 | No change |
| Flush handle | 0 | 0 | MRSA range | 0.007 | 2.10 | No change |
| HH station | 0.257 | 925 | 0.829 | 249 | ||
| Keyboard | 0.0628 | 226 | 0.9654 | 289 | ||
| Heart monitor | 0.319 | 1148 | 0.1875 | 56 | ||
| Sphygmo dial | 0.350 | 1260 | 0.336 | 101 | ||
| Crib | 0.243 | 875 | 0.3125 | 94 | ||
| Under table | 0 | 0 | MRSA range | 0.429 | 129 | |
| IV pump | 0.684 | 2462 | 0.614 | 184 | ||
| Floor, corner | 0.063 | 227 | MRSA range | 0.061 | 18 | No change |
| % of areas in range | No change | 30% | 10% | No change | 20% | 30% |
| MRSA or | 70% | 90% | MRSA or | 80% | 70% | |
| MRSA alone |
| 30% | MRSA alone | 10% | 0% | |
|
| 70% | 60% |
| 70% | 70% |
BR bathroom, HH hand hygiene, Sphygmo sphygmomanometer, C. diff C. difficile. The MRSA lethal energy dose is 12 mJ. The C. difficile lethal energy dose is 38.5 mJ
Live tests results with the small units
| Area | Turbo-UV on the floor | Turbo-UV raised 20 in. | EDU 435 raised 20 in. | |
|---|---|---|---|---|
| Bathroom faucet | No change | |||
| Toilet flush handle | ||||
| Toilet seat | No change | |||
| Shower drain | ||||
| Shower back wall | ||||
| Door handle | ||||
| Toilet paper roll | ||||
| % of areas in range | No change | 29% | 0% | 0% |
| MRSA or | 71% | 100% | 100% | |
| MRSA alone | 0% | 0% | 0% | |
|
| 71% | 100% | 100% | |
Fig. 1Power and Energy Output of the Large UVGI Units. a The power output in time is expressed in mW/cm2 for each distance. b The curves show the cumulative energy (in mJ/cm2) at each distance for each device. A power of 1 watt (W) applied for 1 second (s) produces 1 joule (J) of energy. The green line represents the C. difficile lethal energy dose (38.5 mJ). The dashed curves are the logarithmic regressions. 5 feet = 1.5 m; 8 feet = 2.4 m; 10 feet = 3 m; 15 feet = 4.6 m
Fig. 2Power and Energy Output of the Small UVGI Units. a The power output in time is expressed in mW/cm2 for each distance. b The curves show the cumulative energy (in mJ/cm2) at each distance for each device. A power of 1 watt (W) applied for 1 second (s) produces 1 joule (J) of energy. The green line represents the C. difficile lethal energy dose (38.5 mJ). 5 feet = 1.5 m; 8 feet = 2.4 m