Literature DB >> 29480814

Estimation of patient-specific imaging dose for real-time tumour monitoring in lung patients during respiratory-gated radiotherapy.

Takehiro Shiinoki1, Ryota Onizuka, Daisuke Kawahara, Tatsuhiko Suzuki, Yuki Yuasa, Koya Fujimoto, Takuya Uehara, Hideki Hanazawa, Keiko Shibuya.   

Abstract

PURPOSE: To quantify the patient-specific imaging dose for real-time tumour monitoring in the lung during respiratory-gated stereotactic body radiotherapy (SBRT) in clinical cases using SyncTraX. METHODS AND MATERIALS: Ten patients who underwent respiratory-gated SBRT with SyncTraX were enrolled in this study. The imaging procedure for real-time tumour monitoring using SyncTraX was simulated using Monte Carlo. We evaluated the dosimetric effect of a real-time tumour monitoring in a critical organ at risk (OAR) and the planning target volume (PTV) over the course of treatment. The relationship between skin dose and gating efficiency was also investigated.
RESULTS: For all patients, the mean D50 to the PTV, ipsilateral lung, liver, heart, spinal cord and skin was 118.3 (21.5-175.9), 31.9 (9.5-75.4), 15.4 (1.1-31.6), 10.1 (1.3-18.1), 25.0 (1.6-101.8), and 3.6 (0.9-7.1) mGy, respectively. The mean D2 was 352.0 (26.5-935.8), 146.4 (27.3-226.7), 90.7 (3.6-255.0), 42.2 (4.8-82.7), 88.0 (15.4-248.5), and 273.5 (98.3-611.6) mGy, respectively. The D2 of the skin dose was found to increase as the gating efficiency decreased.
CONCLUSIONS: The additional dose to the PTV was at most 1.9% of the prescribed dose over the course of treatment for real-time tumour monitoring. For OARs, we could confirm the high dose region, which may not be susceptible to radiation toxicity. However, to reduce the skin dose from SyncTraX, it is necessary to increase the gating efficiency.

Entities:  

Mesh:

Year:  2018        PMID: 29480814     DOI: 10.1088/1361-6560/aab242

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  3 in total

1.  Determination of reproducibility of end-exhaled breath-holding in stereotactic body radiation therapy.

Authors:  Motoharu Sasaki; Hitoshi Ikushima; Kanako Sakuragawa; Michihiro Yokoishi; Akira Tsuzuki; Wataru Sugimoto
Journal:  J Radiat Res       Date:  2020-11-16       Impact factor: 2.724

2.  Estimation of effective imaging dose and excess absolute risk of secondary cancer incidence for four-dimensional cone-beam computed tomography acquisition.

Authors:  Yuki Yuasa; Takehiro Shiinoki; Ryota Onizuka; Koya Fujimoto
Journal:  J Appl Clin Med Phys       Date:  2019-10-08       Impact factor: 2.102

3.  A novel dynamic robotic moving phantom system for patient-specific quality assurance in real-time tumor-tracking radiotherapy.

Authors:  Takehiro Shiinoki; Fumitake Fujii; Koya Fujimoto; Yuki Yuasa; Tatsuhiro Sera
Journal:  J Appl Clin Med Phys       Date:  2020-04-13       Impact factor: 2.102

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.