Measurements and simulation of RF heating of implanted stereo-electroencephalography electrodes during MR scans.

Bhumi Bhusal1, Pallab Bhattacharyya2,3, Tanvir Baig1, Stephen Jones2,4, Michael Martens1.   

Abstract

PURPOSE: To assess RF-induced heating during MRI of patients with implanted stereo-electroencephalography electrodes.
METHODS: Simulations and experimental measurements using phantom and a head-only transmit/receive coil on a 3T MR system were performed to evaluate temperature increases at the tip of an 8-contact stereo-electroencephalography electrode and an insulated wire partially immersed into the phantom. The lengths of wire producing maximum (resonant condition) and minimum (anti-resonant condition) heating were evaluated for different entry modes and penetration depths.
RESULTS: For both wire and stereo-electroencephalography electrode, resonant lengths were close to odd integral multiples of RF quarter wavelength in air and antiresonant length close to even integral multiples of RF quarter wavelength, both being unaffected by the entry mode. In the resonant condition, temperature increased by as much as a factor of 10 higher than that at antiresonant condition. Larger penetration depths did not change resonant length, but did lead to increased RF heating.
CONCLUSION: For the partially immersed implants like stereo-electroencephalography electrode, the resonant lengths were found to be independent of the penetration depths and entry modes, although the temperature increases may vary. Avoiding such lengths of cables can reduce the risk of tissue heating during in vivo MRI.
© 2018 International Society for Magnetic Resonance in Medicine.

Entities:  

Keywords:  FDTD simulations; MR safety; RF heating; SEEG; functional MRI; implant

Mesh:

Substances:

Year:  2018        PMID: 29468721     DOI: 10.1002/mrm.27144

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  6 in total

1.  The effect of simulation strategies on prediction of power deposition in the tissue around electronic implants during magnetic resonance imaging.

Authors:  Bach T Nguyen; Julie Pilitsis; Laleh Golestanirad
Journal:  Phys Med Biol       Date:  2020-09-16       Impact factor: 3.609

2.  Evaluating the Safety of Simultaneous Intracranial Electroencephalography and Functional Magnetic Resonance Imaging Acquisition Using a 3 Tesla Magnetic Resonance Imaging Scanner.

Authors:  Yuya Fujita; Hui Ming Khoo; Miki Hirayama; Masaaki Kawahara; Yoshihiro Koyama; Hiroyuki Tarewaki; Atsuko Arisawa; Takufumi Yanagisawa; Naoki Tani; Satoru Oshino; Louis Lemieux; Haruhiko Kishima
Journal:  Front Neurosci       Date:  2022-06-23       Impact factor: 5.152

3.  Risk assessment of copper-containing contraceptives: the impact for women with implanted intrauterine devices during clinical MRI and CT examinations.

Authors:  Wiebke Neumann; Tanja Uhrig; Matthias Malzacher; Verena Kossmann; Lothar R Schad; Frank G Zoellner
Journal:  Eur Radiol       Date:  2018-11-19       Impact factor: 5.315

4.  Modeling radiofrequency responses of realistic multi-electrode leads containing helical and straight wires.

Authors:  Mikhail Kozlov; Marc Horner; Wolfgang Kainz
Journal:  MAGMA       Date:  2019-11-19       Impact factor: 2.310

5.  Safety of MRI in patients with retained cardiac leads.

Authors:  Bach T Nguyen; Bhumi Bhusal; Amir Ali Rahsepar; Kate Fawcett; Stella Lin; Daniel S Marks; Rod Passman; Donny Nieto; Richard Niemzcura; Laleh Golestanirad
Journal:  Magn Reson Med       Date:  2021-12-27       Impact factor: 3.737

6.  A perturbation approach for ultrafast calculation of RF field enhancements near medical implants in MRI.

Authors:  Peter R S Stijnman; Bart R Steensma; Cornelis A T van den Berg; Alexander J E Raaijmakers
Journal:  Sci Rep       Date:  2022-03-10       Impact factor: 4.379

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.