Literature DB >> 29431700

Influence of tidal volume and positive end-expiratory pressure on ventilation distribution and oxygenation during one-lung ventilation.

Zhanqi Zhao1, Wei Wang, Zuojing Zhang, Meiying Xu, Inez Frerichs, Jingxiang Wu, Knut Moeller.   

Abstract

OBJECTIVE: The aim of the study was to explore the feasibility of titrating tidal volume (V T) and positive end-expiratory pressure (PEEP) during one-lung ventilation (OLV) based on ventilation distribution and oxygenation. APPROACH: Twenty-four consecutive patients requiring intubation with a double-lumen tube and subsequent OLV for thoracic surgical procedures were examined prospectively in lateral posture. Electrical impedance tomography (EIT), blood gases, respiratory mechanics were successfully measured in 21 patients at various combinations of V T (4 ml kg-1, 6 ml kg-1, 8 ml kg-1 body weight) and PEEP (0 cm H2O, 4 cm H2O, 8 cm H2O) during OLV. MAIN
RESULTS: Low V T and low PEEP resulted in low global respiratory system compliance (C rs). Arterial partial pressure of O2 (PaO2) decreased with falling V T. Regional C rs measured with EIT showed high values at high V T and high PEEP in all but two patients. Regional C rs in mid and most dependent regions indicated tidal recruitment/derecruitment in eight patients at 8 ml kg-1 of V T and 4 cm H2O of PEEP; in four patients at 8 ml kg-1 and 0 cm H2O; in one patient at 6 ml kg-1 and 8 cm H2O. The changes in regional C rs induced by decreasing PEEP from 8 to 4 cm H2O were much smaller than those from 4 to 0 cm H2O. Ventilation distribution was most inhomogeneous with V T of 8 ml kg-1. All measures differed significantly among various V T and PEEP steps (p  <  0.05). SIGNIFICANCE: By using EIT in combination with PaO2, it is feasible to titrate V T and PEEP at the bedside during OLV.

Entities:  

Mesh:

Substances:

Year:  2018        PMID: 29431700     DOI: 10.1088/1361-6579/aaaeb2

Source DB:  PubMed          Journal:  Physiol Meas        ISSN: 0967-3334            Impact factor:   2.833


  3 in total

Review 1.  Electrical impedance tomography in perioperative medicine: careful respiratory monitoring for tailored interventions.

Authors:  Elena Spinelli; Tommaso Mauri; Alberto Fogagnolo; Gaetano Scaramuzzo; Annalisa Rundo; Domenico Luca Grieco; Giacomo Grasselli; Carlo Alberto Volta; Savino Spadaro
Journal:  BMC Anesthesiol       Date:  2019-08-07       Impact factor: 2.217

2.  Cardio-respiratory physiology during one-lung ventilation: complex interactions in need of advanced monitoring.

Authors:  Ines Marongiu; Elena Spinelli; Tommaso Mauri
Journal:  Ann Transl Med       Date:  2020-04

3.  PEEP guided by electrical impedance tomography during one-lung ventilation in elderly patients undergoing thoracoscopic surgery.

Authors:  Kun Liu; Chengya Huang; Meiying Xu; Jingxiang Wu; Inez Frerichs; Knut Moeller; Zhanqi Zhao
Journal:  Ann Transl Med       Date:  2019-12
  3 in total