EXPERIMENTAL EVALUATION OF DOSIMETRIC CHARACTERIZATION OF GAFCHROMIC EBT3 AND EBT-XD FILMS FOR CLINICAL CARBON ION BEAMS.

Shunsuke Yonai1, Chinatsu Arai1,2, Kaoru Shimoyama1,2, Nathalie Fournier-Bidoz1.   

Abstract

Radiochromic film is a very useful tool for 2D dosimetric measurements in radiotherapy because it is self-developing and has very high-spatial resolution. However, considerable care has to be taken in ion beam radiotherapy owing to the quenching effect of high-linear energy transfer (LET) radiation. In this study, the dose responses of GAFchromic EBT3 and EBT-XD films were experimentally investigated using the clinical carbon ion beam at the Heavy Ion Medical Accelerator in Chiba. Results showed that the relations between absorbed dose and net optical density could be expressed well using an equation proposed by Reinhardt (2015). The quenching effect was evaluated by determining their relative efficiencies for photon irradiation as a function of LET. A correction equation derived in this study allowed the absorbed dose to be determined in the small irradiation field used for carbon ion radiotherapy eye treatments. This study contributes to establishing an absolute dosimetry procedure for heavy ion beams using radiochromic film.

Entities:  

Mesh:

Substances:

Year:  2018        PMID: 29415148     DOI: 10.1093/rpd/ncy006

Source DB:  PubMed          Journal:  Radiat Prot Dosimetry        ISSN: 0144-8420            Impact factor:   0.972


  4 in total

1.  Experimental dosimetry of EDR2 films in scanning carbon-ion irradiation.

Authors:  Weiwei Wang; Yu Deng; Zhijie Huang
Journal:  J Appl Clin Med Phys       Date:  2022-05-20       Impact factor: 2.243

2.  A Modular System for Treating Moving Anatomical Targets With Scanned Ion Beams at Multiple Facilities: Pre-Clinical Testing for Quality and Safety of Beam Delivery.

Authors:  Michelle Lis; Wayne Newhauser; Marco Donetti; Moritz Wolf; Timo Steinsberger; Athena Paz; Marco Durante; Christian Graeff
Journal:  Front Oncol       Date:  2021-03-19       Impact factor: 6.244

3.  Dosimetric Validation of a System to Treat Moving Tumors Using Scanned Ion Beams That Are Synchronized With Anatomical Motion.

Authors:  Michelle Lis; Wayne Newhauser; Marco Donetti; Moritz Wolf; Timo Steinsberger; Athena Paz; Christian Graeff
Journal:  Front Oncol       Date:  2021-09-08       Impact factor: 6.244

4.  Linear energy transfer-independent calibration of radiochromic film for carbon-ion beams.

Authors:  Mutsumi Tashiro; Motohiro Kawashima
Journal:  Phys Imaging Radiat Oncol       Date:  2022-08-08
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.