Bilal Yilmaz1, Ulvihan Ciltas1. 1. Department of Analytical Chemistry, Faculty of Pharmacy, Ataturk University, 25240 Erzurum, Turkey.
Abstract
Rapid, sensitive and specific methods were developed for the determination of diclofenac in pharmaceutical preparations by linear sweep voltammetry (LSV) and gas chromatography (GC) with mass spectrometry (MS) detection. The linearity was established over the concentration range of 5-35 μg/mL for LSV and 0.25-5 μg/mL for GC-MS method. The intra- and inter-day relative standard deviation (RSD) was less than 4.39% and 4.62% for LSV and GC-MS, respectively. Limits of quantification (LOQ) were determined as 4.8 and 0.15 μg/mL for LSV and GC-MS, respectively. No interference was found from tablet excipients at the selected assay conditions. The methods were applied for the quality control of commercial diclofenac dosage forms to quantify the drug and to check the formulation content uniformity.
Rapid, sensitive and specific methods were developed for the determination of diclofenac in pharmaceutical preparations by linear sweep voltammetry (LSV) and gas chromatography (GC) with mass spectrometry (MS) detection. The linearity was established over the concentration range of 5-35 μg/mL for LSV and 0.25-5 μg/mL for GC-MS method. The intra- and inter-day relative standard deviation (RSD) was less than 4.39% and 4.62% for LSV and GC-MS, respectively. Limits of quantification (LOQ) were determined as 4.8 and 0.15 μg/mL for LSV and GC-MS, respectively. No interference was found from tablet excipients at the selected assay conditions. The methods were applied for the quality control of commercial diclofenac dosage forms to quantify the drug and to check the formulation content uniformity.
Diclofenac is a nonsteroidal anti-inflammatory drug (NSAID) that is widely prescribed for the treatment of rheumatoid arthritis, osteoarthritis, musculoskeletal injuries and post-surgery analgesia in human and veterinary medicine [1]. Patients are frequently given special formulations of diclofenac or a co-treatment agent as a therapeutic strategy to attenuate the gastrointestinal tract complications that limit the use of diclofenac and other NSAIDs [2], [3]. Many patients prescribed diclofenac for arthritis also take additional drugs for other chronic health problems such as hypertension [4].To date, several methods for the determination of diclofenac have been reported. These include potentiometry [5], [6], [7], capillary zone electrophoresis [8], high-performance liquid chromatography (HPLC) [9], [10], [11], high-performance liquid chromatography–mass spectrometry (HPLC–MS) [12], spectrophotometry [13], [14], spectrofluorometry [15], [16], thin layer chromatography [17], gas chromatography [18], polarographic analysis [19], and spectroscopic methods [20], [21], [22], [23], [24]. An extensive literature survey revealed that there were several HPLC methods for the determination of diclofenac in blood plasma, whereas there was little other work disclosed only for the quantitative determination of diclofenac in pharmaceutical formulation samples. The reported HPLC methods were influenced by interference of endogenous substances and potential loss of drugs in the re-extraction procedure and involving lengthy, tedious and time-consuming plasma sample preparation and extraction processes and requiring a sophisticated and expensive instrumentation.On extensive survey of literature, no LSV method was reported till date for determination of diclofenac in pure and pharmaceutical dosage forms. The development of a new method capable of determining drug amount in pharmaceutical dosage forms is important. Electroanalytical techniques have been used for the determination of a wide range of drug compounds with the advantages that there is, in most instances, no need for derivatization and that these techniques are less sensitive to matrix effects than other analytical techniques. Additionally, application of electrochemistry involves the determination of electrode mechanism. Redox properties of drugs can give insights into their metabolic fate or their in vivo redox processes or pharmacological activity [25], [26], [27], [28]. Despite the analytical importance of the electrochemical behavior and oxidation mechanism of diclofenac, no report has been published on the voltammetric study of the electrochemical oxidation of diclofenac in nonaqueous media. It is well known that the experimental and instrumental parameters directly affect the electrochemical process and voltammetric response of drugs. Consequently, it would be interesting to investigate the oxidation process of diclofenac in aprotic media.Therefore, this paper describes a new LSV method for determination of diclofenac and a gas chromatography with MS detection. The LSV method was aimed at developing an easy and rapid assay method for diclofenac without any time-consuming sample preparation steps for routine analysis. GC method was attempted to demonstrate the utility of MS detection for determination of diclofenac with simple sample preparation and reasonable analysis time with high precision. In both the proposed methods, there is no need to extract the drug from the formulation excipient, thereby decreasing the error in quantization. Formulation samples can be directly used after dissolving and filtration. The developed methods were used to determine the total drug content in commercially available tablets of diclofenac.
Materials and methods
Chemicals
Diclofenac was obtained from Sigma (St. Louis, MO, USA). Acetonitrile and lithium perchlorate (LiClO4) were purchased from Fluka (Buchs, Switzerland). Diclomec, Dicloflam and Voltaren tablets were obtained from pharmacies (Erzurum, Turkey).
Voltammetric and chromatographic system
Electrochemical experiments were performed on a Gamry Potentiostat Interface 1000 controlled with software PHE 200 and PV 220. All measurements were carried out in a single-compartment electrochemical cell with a standard three-electrode arrangement. A platinum disk with an area of 0.72 cm2 and a platinum wire were used as the working and the counter electrodes, respectively. The working electrode was successively polished with 1.0, 0.3 and 0.05 µm alumina slurries (Buehler) on microcloth pads (Buehler). After each polishing, the electrode was washed with water and sonicated for 10 min in acetonitrile. Then, it was immersed into a hot piranha solution (3:1, H2SO4, 30% H2O2) for 10 min, and rinsed copiously with water. All potentials were reported versus Ag/AgCl/KCl (3.0 M) reference electrode (BAS Model MF-2078) at room temperature. The electrolyte solutions were degassed with purified nitrogen for 10 min before each experiment and bubbled with nitrogen during the experiment.Chromatographic analysis was carried out on an Agilent 6890N gas chromatography system equipped with 5973 series mass selective detector, 7673 series autosampler and chemstation (Agilent Technologies, Palo Alto, CA). HP-5 MS column (30 m×0.25 mm, 0.25 µm) was used for separation. Splitless injection was used and the carrier gas was helium at a flow rate of 1 μg/mL. The injector and detector temperatures were 250 °C. The MS detector parameters were transfer line temperature 280 °C, solvent delay 3 min and electron energy 70 eV. MS was run in electron impact mode with selected ion monitoring (SIM) for quantitative analysis.
Preparation of the standard and quality control (QC) solutions
For the LSV method, the stock standard solution of diclofenac was prepared in 0.1 M LiClO4/acetonitrile to a concentration of 100 μg/mL. For the GC–MS method, the stock solution of diclofenac was prepared in methanol solution to a concentration of 100 μg/mL. Standard solutions were prepared as 5–35 μg/mL (5, 10, 15, 20, 25, 30 and 35 μg/mL) for LSV method and 0.25–5 μg/mL (0.25, 1, 2, 3, 4 and 5 μg/mL) for the GC–MS method. The QC samples were prepared by adding aliquots of standard working solution of diclofenac to obtain the final concentrations of 7.5, 17.5 and 32.5 μg/mL for the LSV method and 0.75, 2.5 and 4.5 μg/mL for the GC–MS method.
Procedure for pharmaceutical preparations
A total of 10 tablets of diclofenac (Diclomec, Dicloflam and Voltaren) were accurately weighed and powdered. For the LSV method, an amount of this powder corresponding to one tablet diclofenac content was weighed and accurately transferred into a 100 mL calibrated flask and 50 mL of 0.1 M LiClO4/acetonitrile was added and then the flask was sonicated to 10 min at room temperature. The flask was filled to volume with 0.1 M LiClO4/acetonitrile. The resulting solutions in both the cases were filtered through Whatman filter paper no. 42 and suitably diluted to get a final concentration within the limits of linearity for the respective proposed method. For the GC–MS method, an appropriate volume of filtrate was diluted further with methanol so that the concentration of diclofenac in the final solution was within the working range, and then analyzed by GC–MS.
Data analysis
All statistical calculations were performed with the Statistical Product and Service Solutions (SPSS) for Windows, version 10.0. Correlations were considered statistically significant if calculated P values were 0.05 or less.
Results and discussion
Method development and optimization
The electrochemical behavior of diclofenac was investigated at the Pt disc electrode in anhydrous acetonitrile solution containing 0.1 M LiClO4 as the supporting electrolyte by using cyclic voltammetry (CV).Fig. 1 shows a typical cyclic voltammogram of 20 μg/mL diclofenac recorded under these conditions for describing the scan rate of 0.2 V/s. In the anodic sweep, two oxidation peaks were seen at about potentials of 0.87 and 1.27 V, respectively.
Fig. 1
(A) Cyclic voltammogram for the oxidation of 20 μg/mL diclofenac and (B) blank in acetonitrile containing 0.1 M LiClO4 at Pt disk electrode. Scan rate: 0.2 V/s.
(A) Cyclic voltammogram for the oxidation of 20 μg/mL diclofenac and (B) blank in acetonitrile containing 0.1 M LiClO4 at Pt disk electrode. Scan rate: 0.2 V/s.To gain a deeper insight into the voltammetric waves, the effect of scan rate on the anodic peak currents (İm) and peak potentials (Ep) was studied in the range of 0.01–1 V/s the potential scan rates in acetonitrile solution containing 20 μg/mL of diclofenac. Scan rate dependency experiments showed that the peak currents for peak varied linearly with the scan rate (ν), which points out the adsorption-controlled process. However, the plots of logarithm of peak currents versus logarithm of scan rates for 20 μg/mL concentration of diclofenac displayed straight lines with 0.497 slope, which are close to the theoretical value of 0.5 expected for an ideal diffusion-controlled electrode process [29]. log Im–log ν curve is more eligible for this aim; therefore, a diffusional process for peak should be considered. These results suggested that the redox species were diffusing freely from solution and not precipitating onto the electrode surface. The reason for this behavior might be due to the solubility of the intermediate species in acetonitrile or poor adherence of products on the electrode surface.The oxidation peak potential (E) for peaks shifted toward more positive values with increasing scan rate. The relationship between the peak potential and scan rate is described by the following equation:and from the variation of peak potential with scan rate αna can be determined, where α is the transfer coefficient and na is the number of electrons transferred in the rate determining step. According to this equation, the plots of the peak potentials versus ln ν for oxidation peak show linear relationship. The slope indicates the value of αna is 0.38 for peak. On the base of the above results, the na is 1 and then the value of α is calculated to be 0.38, which is reasonable for the most of irreversible electrode processes. Based on the above discussions the oxidation process of diclofenac is controlled by the diffusion step and one electron and one proton are involved in the reaction.During GC–MS method development, a capillary column coated with 5% phenyl and 95% dimethylpolysiloxane was used for separation. The injection port and detector temperature was set to 250 °C. Different temperature programs were investigated to give an optimum temperature program as follows: initial temperature was 150 °C, held for 1 min, increased to 220 °C at 20 °C/min, held for 1 min, and finally to 300 °C at 10 °C/min with a final maintenance of 1.0 min. The injector volume was 1 μL in splitless mode.
Method validation
To ensure optimization of the methods in light of the standardization rules, the developed methods were validated in terms of specificity, linearity, precision, accuracy, LOD, LOQ, recovery and the stability effect which was investigated by analyzing the pure diclofenac solution and drug samples [30].
Specificity
All the solutions were scanned from 0.5 to 1.5 V and checked for change in the peaks at respective potentials (Fig. 2).
Fig. 2
(A) Linear sweep voltammograms for different concentrations of diclofenac in acetonitrile solution containing 0.1 M LiCIO4 (5, 10, 15, 20, 25, 30 and 35 μg/mL) and (B) mean calibration graph (n=6).
(A) Linear sweep voltammograms for different concentrations of diclofenac in acetonitrile solution containing 0.1 M LiCIO4 (5, 10, 15, 20, 25, 30 and 35 μg/mL) and (B) mean calibration graph (n=6).In a separate study, the specificity of the method was investigated by observing interferences between diclofenac and the excipients. For GC–MS, electron impact mode with selected ion monitoring (SIM) was used for quantitative analysis (m/z 214 for diclofenac).The mass spectra of the diclofenac are shown in Fig. 3. The retention time of diclofenac in GC–MS method was approximately 5.9 min with good peak shape (Fig. 4).
Fig. 3
(A) MS spectra of diclofenac and (B) chemical structure of diclofenac.
Fig. 4
(A) GC–MS chromatograms of diclofenac (0.25, 0.5, 1, 2, 3, 4 and 5.0 μg/mL) [Selected ion monitoring (SIM) mode, m/z 214 for diclofenac] and (B) mean calibration graph (n=6).
(A) MS spectra of diclofenac and (B) chemical structure of diclofenac.(A) GC–MS chromatograms of diclofenac (0.25, 0.5, 1, 2, 3, 4 and 5.0 μg/mL) [Selected ion monitoring (SIM) mode, m/z 214 for diclofenac] and (B) mean calibration graph (n=6).
Linearity
For the LSV and GC–MS measurements, the solutions were prepared by dilution of the stock solution of diclofenac to reach a concentration range of 5–35 μg/mL (5, 10, 15, 20, 25, 30 and 35 μg/mL) and 0.25–5 μg/mL (0.25, 0.5, 1, 2, 3, 4 and 5 μg/mL), respectively. Calibration curves were constructed for diclofenac standard by plotting the concentration of diclofenac versus voltammogram and peak area response. The calibration curve constructed was evaluated by its correlation coefficient. The correlation coefficients (r) of all the calibration curves were consistently greater than 0.99. The regression equations were calculated from the calibration graphs, along with the standard deviations of the slope and intercept on the ordinate. The results are shown in Table 1.
Table 1
Linearity of diclofenac.
Method
Range (µg/mL)
LR
Sa
Sb
R
LOD (µg/mL)
LOQ (µg/mL)
LSV
5–35
y=0.0886x−0.240
0.043
0.012
0.992
1.6
4.8
GC–MS
0.25–5
y=732.34x−40.907
11.096
2.081
0.999
0.05
0.15
Based on three calibration curves, LR: linear regression, Sa: standard deviation of intercept of regression line, Sb: standard deviation of slope of regression line, R: coefficient of correlation, x: diclofenac concentration, LOD: limit of detection, LOQ: limit of quantification.
Linearity of diclofenac.Based on three calibration curves, LR: linear regression, Sa: standard deviation of intercept of regression line, Sb: standard deviation of slope of regression line, R: coefficient of correlation, x: diclofenac concentration, LOD: limit of detection, LOQ: limit of quantification.
Precision and accuracy
The precision of the LSV and GC–MS methods was determined by repeatability (intra-day) and intermediate precision (inter-day). Repeatability was evaluated by analyzing QC samples six times per day at three different concentrations. The intermediate precision was evaluated by analyzing the same samples once daily for two days. The RSD of the predicted concentrations from the regression equation was taken as precision. The accuracy of this analytic method was assessed as the percentage relative error. For all the concentrations studied, intra- and inter-day relative standard deviation values were ≤4.62% and for all concentrations of diclofenac the relative errors were ≤6.29%. These results are given in Table 2.
Table 2
Precision and accuracy of diclofenac.
Method
Added (µg/mL)
Intra-day
Inter-day
Found±SDa (µg/mL)
Precision (% RSD)b
Accuracyc
Found±SDa (µg/mL)
Precision (% RSD)b
Accuracyc
LSV
7.5
7.88±0.23
2.91
5.06
7.29±0.19
2.61
−2.80
17.5
18.13±0.30
1.65
3.60
18.60±0.32
1.72
6.29
32.5
33.44±1.47
4.39
2.89
33.02±1.59
4.62
1.60
GC–MS
0.75
0.77±0.03
3.89
2.67
0.74±0.02
2.30
−1.33
2.5
2.48±0.09
3.62
−2.08
2.49±0.06
2.65
−0.40
4.5
4.47±0.07
1.56
−1.17
4.48±0.05
1.21
−0.44
SD: standard deviation of six replicate determinations.
RSD: relative standard deviation, average of six replicate determinations.
Accuracy: (found–added)/added ×100.
Precision and accuracy of diclofenac.SD: standard deviation of six replicate determinations.RSD: relative standard deviation, average of six replicate determinations.Accuracy: (found–added)/added ×100.
LOD and LOQ
For LSV measurements, LOD and LOQ of diclofenac were determined using calibration standards. The LOD and LOQ values were calculated as 3.3 σ/S and 10 σ/S, respectively, where S is the slope of the calibration curve and σ is the standard deviation of y-intercept of regression equation (n=6) [31].For GC–MS measurements, the LOD and LOQ of diclofenac were determined by injecting progressively low concentration of the standard solution under the chromatographic conditions. The lowest concentration assayed was regarded as LOQ, where the signal/noise ratio was at least 10:1. The LOD was defined as a signal/noise ratio of 3:1. The LOD and LOQ for LSV were 1.6 and 4.8 μg/mL, for GC–MS 0.05 and 0.15 μg/mL, respectively. Among the two methods, GC–MS was more sensitive than LSV (Table 1).
Recovery
To determine the accuracy of the LSV and GC–MS methods and to study the interference of formulation additives, the recovery was checked at three different concentration levels. Analytical recovery experiments were performed by adding known amount of pure drugs to pre-analyzed samples of commercial dosage forms. The recovery values were calculated by comparing concentration obtained from the spiked samples with actual added concentrations. These values are also listed in Table 3.
Table 3
Recovery of diclofenac in pharmaceutical preparations.
Commercial preparation
Method
n
Found (mg) (Mean±SD)
Recovery
RSDa (%)
Confidence interval
Diclomec (100 mg/tablet)
LSV
6
99.7±0.5
99.7
0.5
98.0–101.2
GC–MS
6
101.1±1.0
101.1
1.0
99.0–102.5
Dicloflam (50 mg/tablet)
LSV
6
49.4±1.3
98.8
2.6
48.3–51.2
GC–MS
6
50.5±1.5
101.0
3.1
49.8–51.3
Voltaren (75 mg/tablet)
LSV
6
74.8±2.1
99.7
2.8
74.7–75.9
GC–MS
6
74.7±2.3
99.6
3.1
74.3–76.1
SD: standard deviation of six replicate determinations, RSD: relative standard deviation.
Average of six replicate determinations.
Recovery of diclofenac in pharmaceutical preparations.SD: standard deviation of six replicate determinations, RSD: relative standard deviation.Average of six replicate determinations.
Ruggedness
Determination of diclofenac using both the LSV and GC–MS methods was carried out by a different analyst on the same instrument with the same standard (Table 4).
Table 4
The results of analyses of diclofenac by a different analyst.a
Method
Added (µg/mL)
Found (µg/mL) (Mean±SD)
Recovery
RSDa (%)
LSV
5
4.9±0.1
98
2.0
15
14.8±0.2
99
1.3
35
35.4±0.7
101
1.9
GC–MS
0.5
0.5±0.01
100
2.0
15
14.8±0.1
99
0.6
35
34.0±0.3
97
0.9
Mean measurements of six replicate determinations.
The results of analyses of diclofenac by a different analyst.aMean measurements of six replicate determinations.The results showed no statistical differences between different operators, suggesting that the developed method was rugged.
Stability
Stability studies indicated that the samples were stable when kept at room temperature, +4 °C and −20 °C for 24 h (short-term) and −20 °C for 72 h (long-term).The results are given in Table 5, where the percent ratios were within the acceptance range of 90–110%.
Table 5
Stability of diclofenac in solution at different temperatures (Recovery±RSD, %).
Method
Added (µg/mL)
Room temperature
4 °C
−20 °C
8 h
24 h
24 h
72 h
24 h
72 h
LSV
0.75
101.1±4.46
99.2±2.94
102.3±3.84
98.6±3.15
101.2±4.12
99.4±3.87
2.5
101.8±5.02
101.4±3.64
98.2±5.14
101.9±3.84
99.4±3.47
99.7±3.09
4.5
98.5±3.17
99.2±4.76
98.2±4.06
97.4±3.75
97.8±5.05
101.2±4.64
GC–MS
1.0
99.3±5.18
102.2±3.29
99.7±3.16
101.2±4.51
101.1±2.96
98.6±3.55
3.0
101.3±4.74
101.4±4.47
102.1±4.08
99.4±3.71
98.6±1.84
98.9±2.44
5.0
102.6±3.44
98.5±4.60
101.4±3.14
97.8±3.01
99.7±3.49
97.8±2.27
RSD: Relative standard deviation of six replicate determinations.
Stability of diclofenac in solution at different temperatures (Recovery±RSD, %).RSD: Relative standard deviation of six replicate determinations.
Comparison of the methods
Birajdar et al. [10] developed a reverse phase HPLC method for the simultaneous determination of rabeprazole and diclofenac in tablet. The method was based on HPLC separation of both drugs in reverse phase mode using Phenomenox C18 column with Waters HPLC system by using the mobile phase composition of acetonitrile and 50 mM ammonium acetate buffer (pH 3.6) (60:40, v/v) at flow rate of 1 mL/min. Detection wavelength used was at 254 nm. Linearity was obtained in the concentration range of 1.0–3.2 μg/mL for rabeprazole and 6.0–16.0 μg/mL for diclofenac.Sastry et al. [13] described a spectrophotometric method for the determination of diclofenac sodium in bulk samples and pharmaceutical preparations with p-N,N dimethylphenylenediamine as a solvent and having maximum absorbance at 670 nm. The reaction was sensitive enough to permit the determination of 2.0–24 μg/mL.Agrawal et al. [14] described two methods for the determination of diclofenac. In the first method, diclofenac reduced iron(III) to iron(II) having a maximum absorbance at 520 nm. The reaction obeys Beer׳s law for concentrations of 10–80 μg/mL. In the second method, diclofenac was treated with methylene blue in the presence of phosphate buffer (pH 6.8) and the complex was extracted with chloroform. The complex had a maximum absorbance at 640 nm and linearity was in the range of 5–40 μg/mL.Marcela et al. [15] developed a spectrofluorometric method for the determination of diclofenac. The method was based on its reaction with cerium (IV) in an acidic solution and measurement of the fluorescence of the Ce(III) ions produced. The absorbance was measured at 356 and 250 nm with double distilled water as solvent.Thongchai et al. [17] developed a high-performance thin layer chromatographic method for the determination of diclofenac sodium in pharmaceutical formulations. The drug was extracted from the sample and then various aliquots of this solution were spotted automatically by means of Camag Linomat IV on a silica gel 60 F254 aluminum plate, using a mixture of toluene: ethyl acetate: glacial acetic acid (60:40:1, v/v/v) as mobile phase. The spot areas were quantified by densitometry at 282 nm. Linear calibration curve was obtained over the range of 5–80 μg/mL.For LSV and GC–MS measurements, calibration curves were linear over the concentration range of 5–35 and 0.25–5 μg/mL for diclofenac, which is as good as or superior to that reported in other papers [10], [13], [14], [15], [17].Also, LSV and GC–MS methods were applied for the determination of diclofenac in three commercial tablets (Fig. 5, Fig. 6).
Fig. 5
Linear sweep voltammogram of Voltaren tablet solution containing 15 and 30 μg/mL of diclofenac.
Fig. 6
GC–MS chromatogram of Voltaren tablet solution containing 1 and 4 μg/mL of diclofenac.
Linear sweep voltammogram of Voltaren tablet solution containing 15 and 30 μg/mL of diclofenac.GC–MS chromatogram of Voltaren tablet solution containing 1 and 4 μg/mL of diclofenac.The results show the high reliability and reproducibility of two methods. The results were statistically compared using the F-test. At 95% confidence level, the calculated F-values do not exceed the theoretical values (Table 6). Therefore, there is no significant difference between LSV and GC–MS methods.
Table 6
Comparison of the proposed and reported methods for determination of diclofenac.
Method
Recovery (Mean±SD, %)
Confidence limits
P value
F-test
Official method (USP XXIV) (HPLC)
99.4±0.39
99.3±2.18
0.342
Fc=1.72
LSV
99.7±2.63
99.1±3.86
Ft=3.00
GC–MS
100.2±7.63
100.1±3.97
SD: standard deviation of six replicate determinations, Fc: calculated F-value, Ft: tabulated F-value, Ho hypothesis: no statistically significant difference exists between three methods, Ft>Fc: Ho hypothesis is accepted (P>0.05).
Comparison of the proposed and reported methods for determination of diclofenac.SD: standard deviation of six replicate determinations, Fc: calculated F-value, Ft: tabulated F-value, Ho hypothesis: no statistically significant difference exists between three methods, Ft>Fc: Ho hypothesis is accepted (P>0.05).USP XXIV has recommended HPLC method for analysis of diclofenac in pure and dosage forms (tablet). The method recommends use of a mobile phase of methanol and phosphate buffer (pH 2.5) at a flow rate of 1 μg/mL.Also, the proposed LSV and GC–MS methods were compared with the HPLC method of USP XXIV [32]. There was no significant difference between the three methods with respect to mean values and standard deviations at the 95% confidence level (Table 6). Therefore, this is suggested that the two methods are equally applicable.
Conclusion
In the present work, two new methods have been developed and validated for routine determination of diclofenac in pharmaceutical preparations. Linearity range, precision, accuracy, LOD and LOQ are suitable for the quantification of diclofenac in pharmaceutical preparations. The sample recoveries in three formulations were in good agreement with their respective label claims. No extraction procedure is involved. According to the statistical comparison of the results, there is no significant difference between LSV and GC–MS methods. The proposed methods can be used for the routine quality control analysis of diclofenac in pharmaceutical preparations in a total time of 6 min.
Authors: Saad S M Hassan; Wagiha H Mahmoud; Mohamed A F Elmosallamy; Mahra H Almarzooqi Journal: J Pharm Biomed Anal Date: 2005-09-01 Impact factor: 3.935