Mojtaba Shamsipur1, Fahimeh Jalali, Sohrab Ershad.
Abstract
A novel diclofenac ion-selective electrode is prepared, characterized and used in pharmaceutical analysis. The diclofenac complex with hexadecylpyridinium bromide is obtained in situ by soaking the PVC-membranes in a 1x10(-2) M diclofenac solution. Among four different solvent mediators tested, dibutyl phthalate (DBP) exhibited a proper behavior including Nernstian slope of the calibration curve, fast response time and good reproducibility of the emf values. The electrode exhibits a Nernstian slope of -59+/-1 mV decade(-1) for diclofenac in the concentration range 1.0x10(-5) to 1.0x10(-2) M with a limit of detection of 4.0x10(-6) M. The electrode displays a good selectivity for diclofenac with respect to a number of common inorganic and organic species. It can be used in a pH range of 6.0-9.0. The membrane sensor was successfully applied to the determination of diclofenac in its tablets as well as for its recovery from blood serum and urine samples.Entities:
Mesh:
Substances:
Year: 2005 PMID: 15862671 DOI: 10.1016/j.jpba.2004.07.051
Source DB: PubMed Journal: J Pharm Biomed Anal ISSN: 0731-7085 Impact factor: 3.935