| Literature DB >> 29403841 |
Gehad G Mohamed1, F A Nour El-Dien1, Eman Y Z Frag1, Marwa El-Badry Mohamed1.
Abstract
The construction and performance characteristics of new sensitive and selective in situ modified screen printed (ISPE) and carbon paste (ICPE) electrodes for determination of naphazoline hydrochloride (NPZ-HCl) have been developed. The electrodes under investigation show potentiometric response for NPZ-HCl in the concentration range from 7.0×10-7 to 1.0×10-2 M at 25 °C and the electrode response is independent of pH in the range of 3.1-7.9. These sensors have slope values of 59.7±0.6 and 59.2±0.2 mV decade-1 with detection limit values of 5.6×10-7 and 5.9×10-7 M NPZ-HCl using ISPE and ICPE, respectively. These electrodes show fast response time of 4-7 s and 5-8 s and exhibits lifetimes of 28 and 30 days for ISPE and ICPE, respectively. Selectivity for NPZ-HCl with respect to a number of interfering materials was also investigated. It was found that there is no interference from the investigated inorganic cations, anions, sugars and other pharmaceutical excipients. The proposed sensors were applied for the determination of NPZ-HCl in pharmaceutical formulation using the direct potentiometric method. It showed a mean average recovery of 100.2% and 102.6% for ISPE and ICPE, respectively. The obtained results using the proposed sensors were in good agreement with those obtained using the official method. The proposed sensors show significantly high selectivity, response time, accuracy, precision, limit of detection (LOD) and limit of quantification (LOQ) compared with other proposed methods.Entities:
Keywords: Eye drops; ISPE and ICPE; Naphazoline hydrochloride; Potentiometry
Year: 2013 PMID: 29403841 PMCID: PMC5761002 DOI: 10.1016/j.jpha.2013.03.008
Source DB: PubMed Journal: J Pharm Anal ISSN: 2214-0883
Fig. 1The structure of naphazoline hydrochloride.
Effect of the nature and content of ion pairing agents on the performance of ISPE and ICPE sensors.
| Electrodes | Composition (mg) | |||||
|---|---|---|---|---|---|---|
| ICPE | ISPE | |||||
| Ion pair (mg) | Graphite (mg) | Plasticizer (TCP) (mg) | Slope (mV decade−1) | Ion pair (mg) | Slope (mV decade−1) | |
| NaTPB | 3 | 250 | 114.3 | 41.9±0.5 | 6 | 59.7±0.6 |
| 6 | 250 | 114.3 | 59.2±0.2 | 12 | 51.9±0.9 | |
| 9 | 250 | 114.3 | 56.1±0.9 | 16 | 46.5±0.9 | |
| 12 | 250 | 114.3 | 55.1±1.0 | 22 | 44.1±1.6 | |
| 16 | 250 | 114.3 | 44.9±0.7 | 30 | 40.3±1.0 | |
| 20 | 250 | 114.3 | 38.7±0.9 | |||
| PMA | 3 | 250 | 114.3 | 42.7±0.2 | 6 | 35.5±0.5 |
| 6 | 250 | 114.3 | 44.3±0.5 | 12 | 39.7±0.7 | |
| 9 | 250 | 114.3 | 39.1±0.8 | 16 | 16.9±0.3 | |
| 12 | 250 | 114.3 | 37.4±0.3 | 22 | 16.4±0.1 | |
| 16 | 250 | 114.3 | 36.0±0.2 | 30 | 15.8±0.5 | |
| 20 | 250 | 114.3 | 28.8±1.0 | |||
| STA | 3 | 250 | 114.3 | 47.9±0.4 | 6 | 14.9±0.2 |
| 6 | 250 | 114.3 | 38.9±0.3 | 12 | 20.1±0.3 | |
| 9 | 250 | 114.3 | 36.7±1.0 | 16 | 36.1±0.8 | |
| 12 | 250 | 114.3 | 33.1±0.5 | 22 | 31.1±0.6 | |
| 16 | 250 | 114.3 | 23.8±0.2 | 30 | 30.6±0.8 | |
| 20 | 250 | 114.3 | 21.6±1.0 | |||
| PTA | 3 | 250 | 114.3 | 36.4±0.2 | 6 | 28.7±0.9 |
| 6 | 250 | 114.3 | 40.6±0.4 | 12 | 38.4±0.4 | |
| 9 | 250 | 114.3 | 39.0±0.5 | 16 | 47.5±0.5 | |
| 12 | 250 | 114.3 | 26.4±0.6 | 22 | 35.5±0.9 | |
| 16 | 250 | 114.3 | 25.2±1.0 | 30 | 34.6±0.5 | |
| 20 | 250 | 114.3 | 25.0±0.6 | |||
| RN | 3 | 250 | 114.3 | 43.2±0.4 | 6 | 24.5±0.5 |
| 6 | 250 | 114.3 | 47.6±0.6 | 12 | 29.0±0.6 | |
| 9 | 250 | 114.3 | 46.8±0.6 | 16 | 38.0±0.7 | |
| 12 | 250 | 114.3 | 44.0±0.5 | 22 | 37.6±0.7 | |
| 16 | 250 | 114.3 | 42.2±0.9 | 30 | 36.1±1.3 | |
| 20 | 250 | 114.3 | 41.6±1.3 | |||
Fig. 2Effect of NaTPB on the performance of (A) ISPE and (B) ICPE sensors.
Response characteristics of ISPE and ICPE sensors.
| Parameter | ISPE | ICPE |
|---|---|---|
| Slope (mV decade−1) | 59.7±0.6 | 59.20±0.2 |
| Correlation coefficient, | 0.996 | 0.974 |
| Detection limit (M) | 5.6×10–7 | 5.9×10–7 |
| Quantification limit (M) | 1.9×10–6 | 1.9×10–6 |
| Working pH range | 3.1–7.9 | 3.1–7.9 |
| Life time (days) | 28 | 30 |
| Concentration range (M) | 7.0×10−7–1.0×10–2 | 7.0×10−7–1.0×10–2 |
| SD | 0.023 | 0.008 |
| RSD (%) | 0.5 | 0.2 |
| Response time (s) | 4–7 | 5–8 |
| Isothermal coefficient (V/°C) | 0.0022 | 0.0008 |
Fig. 3Effect of pH on (A) ISPE and (B) ICPE modified with NaTPB ion pairing agent.
Selectivity coefficients values for ISPE and ICPE sensors.
| Interferingion | ICPE | ISPE | ||
|---|---|---|---|---|
| MPM | SSM | MPM | SSM | |
| Glucose | 3.00×10–2 | – | 1.50×10–2 | – |
| Lactose | 1.65×10–2 | – | 1.25×10–2 | – |
| Maltose | 1.60×10–2 | – | 0.80×10–2 | – |
| Starch | 2.00×10–2 | – | 3.00×10–2 | – |
| Fructose | 5.00×10–3 | – | 9.71×10–3 | – |
| Glycine | 1.00×10–2 | – | 8.75×10–3 | – |
| NH4+ | – | 4.70×10–4 | – | 1.48×10–3 |
| K+ | – | 1.29×10–3 | – | 9.66×10–4 |
| Na+ | – | 2.00×10–4 | – | 5.01×10–4 |
| Cu2+ | – | 6.83×10–1 | – | 5.64×10–2 |
| Ni2+ | – | 2.81×10–2 | – | 3.69×10–2 |
| Zn2+ | – | 9.67×10–1 | – | 3.42×10–2 |
| Ca2+ | – | 7.98×10–6 | – | 1.36×10–5 |
| Cd2+ | – | 7.10×10–6 | – | 2.72×10–5 |
| Cr3+ | – | 2.43×10–1 | – | 8.57×10–3 |
Fig. 4The dynamic response time of (A) ISPE and (B) ICPE potentiometric sensors modified with NaTPB ion pairing agent.
Fig. 5Effect of soaking time on the ICPE.
Determination of NPZ-HCl in pharmaceutical formulation (Neozoline) using ISPE and ICPE.
| Method | Taken (mg) | ICPE | ISPE | British Pharmacopoeia | ||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Found (mg) | Recovery (%)±SD | RSD (%) | Found (mg) | Recovery (%)±SD | RSD (%) | Found (mg) | Recovery (%)±SD | RSD (%) | ||
| Standard addition method | 0.50 | 0.483 | 96.60±0.23 | 0.238 | 0.489 | 97.80±0.38 | 0.391 | 0.494 | 98.75±0.35 | 0.358 |
| 1.23 | 1.199 | 97.50±0.71 | 0.725 | 1.200 | 97.56±0.28 | 0.289 | 1.22 | 99.1±0.28 | 0.286 | |
| 3.00 | 2.921 | 97.36±0.50 | 0.508 | 2.928 | 97.60±0.57 | 0.580 | 2.96 | 98.67±0.50 | 0.502 | |
| 6.00 | 5.849 | 97.48±0.74 | 0.762 | 5.826 | 97.10±0.28 | 0.291 | 5.91 | 98.55±0.64 | 0.646 | |
| Potentiometric titration method | 0.50 | 0.488 | 97.63±0.89 | 0.913 | 0.488 | 97.55±0.33 | 0.338 | |||
| 1.23 | 1.197 | 97.32±0.26 | 0.270 | 1.184 | 96.26±0.35 | 0.367 | ||||
| 3.00 | 2.921 | 97.36±0.92 | 0.944 | 2.913 | 97.10±0.28 | 0.291 | ||||
| 6.00 | 5.974 | 99.75±0.35 | 0.644 | 5.805 | 96.75±0.35 | 0.365 | ||||
| Calibration method | 0.50 | 0.513 | 102.60±0.53 | 0.528 | 0.501 | 100.20±0.28 | 0.275 | |||
| 1.23 | 1.235 | 100.40±0.53 | 0.528 | 1.247 | 101.40±0.18 | 0.177 | ||||
| 3.00 | 2.922 | 97.40±0.57 | 0.581 | 2.996 | 99.87±0.92 | 0.921 | ||||
| 6.00 | 5.805 | 96.75±0.35 | 0.365 | 5.952 | 99.20±0.42 | 0.428 | ||||
| Standard addition method: | Standard addition method: | |||||||||
| Potentiometric titration: | Potentiometric titration: | |||||||||
| Calibration method: | Calibration method: | |||||||||
Tabulated F value at 95% confidence level is 6.39 (n=5).
Tabulated t value at 95% confidence level is 2.776 (n=5).
Evaluation of accuracy and precision (intra- and inter-day) of the ISPE and ICPE sensors.
| Drug | Electrode type | Taken (mg) | Intra-day | Inter-day | ||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Found (mg) | Recovery (%) | SD | RSD (%) | Found (mg) | Recovery (%) | SD | RSD (%) | |||
| NPZ-HCl | ICPE | 12.337 | 12.394 | 100.46 | 0.004 | 0.149 | 11.89 | 96.38 | 0.035 | 1.163 |
| 1.2337 | 1.212 | 98.24 | 0.008 | 0.209 | 1.28 | 103.8 | 0.036 | 0.915 | ||
| 0.1234 | 0.123 | 99.68 | 0.008 | 0.167 | 0.125 | 101.3 | 0.009 | 0.179 | ||
| ISPE | 12.337 | 12.74 | 103.3 | 0.009 | 0.3 | 12.17 | 98.65 | 0.026 | 0.867 | |
| 1.2337 | 1.245 | 100.9 | 0.057 | 1.422 | 1.245 | 100.92 | 0.005 | 0.137 | ||
| 0.1234 | 0.126 | 102.1 | 0.023 | 0.47 | 0.125 | 101.3 | 0.009 | 0.179 | ||
| Neozoline | ICPE | 5.5 | 5.5 | 100.0 | 0.008 | 0.167 | 5.482 | 99.67 | 0.021 | 0.416 |
| 2.25 | 2.247 | 99.87 | 0.005 | 0.183 | 2.256 | 100.3 | 0.036 | 1.185 | ||
| 0.5 | 0.482 | 96.40 | 0.011 | 0.566 | 0.476 | 95.20 | 0.033 | 1.618 | ||
| ISPE | 5.5 | 5.498 | 99.96 | 0.008 | 0.167 | 5.493 | 99.87 | 0.011 | 0.228 | |
| 2.25 | 2.239 | 99.51 | 0.009 | 0.3 | 2.253 | 100.1 | 0.027 | 0.899 | ||
| 0.5 | 0.479 | 95.80 | 0.036 | 1.775 | 0.476 | 95.30 | 0.035 | 1.731 | ||