| Literature DB >> 29403735 |
Palani Venkatesh1, Mahesh Daggumati2.
Abstract
Simultaneous quantification of Lamivudine and Zidovudine in tablets by HPTLC method was developed and validated. The chromatograms were developed using a mobile phase of toluene:ethyl acetate:methanol (4:4:2, v/v/v) on pre-coated plate of silica gel GF aluminum TLC plate and quantified by densitometric absorbance mode at 276 nm. The Rf values were 0.41±0.03 and 0.60±0.04 for Lamivudine and Zidovudine, respectively. The linearity of the method was found to be within the concentration range of 50-250 ng/spot for Lamivudine and for Zidovudine, it was 100-500 ng/spot. The lower limits of detection and quantification were 2.23 ng/spot and 7.90 ng/spot for Lamivudine and 2.90 ng/spot and 8.85 ng/spot for Zidovudine. The method was also validated for precision, specificity and recovery. This developed method was used to analyze fixed-dose tablets (Duovir, Cipla Ltd) samples of Lamivudine and Zidovudine.Entities:
Keywords: Lamivudine; Methanol; Normal-phase HPTLC; Zidovudine
Year: 2011 PMID: 29403735 PMCID: PMC5760830 DOI: 10.1016/j.jpha.2011.11.002
Source DB: PubMed Journal: J Pharm Anal ISSN: 2214-0883
Figure 1(A) Lamivudine and (B) Zidovudine.
Linearity results.
| Components | Concentration range (ng/spot) | Equation for regression line | |
|---|---|---|---|
| Lamivudine | 50–250 | 0.9991 | |
| Zidovudine | 100–500 | 0.9995 | |
Accuracy and intermediate precision.
| Drugs | Concentration (ng/spot) | Intra-day | Inter-day | ||
|---|---|---|---|---|---|
| Accuracy (%) | Precision (% RSD) | Accuracy (%)a | Precision (% RSD) | ||
| Lamivudine | 50 | 98.57±1.27 | 0.48 | 98.84±1.34 | 0.38 |
| 100 | 99.83±1.24 | 0.52 | 99.74±1.52 | 0.62 | |
| 150 | 99.58±1.32 | 0.46 | 99.28±1.27 | 0.39 | |
| Zidovudine | 100 | 99.26±1.63 | 0.34 | 99.05±1.12 | 0.43 |
| 200 | 98.69±1.52 | 0.45 | 99.21±1.36 | 0.82 | |
| 300 | 99.34±1.34 | 0.68 | 99.28±1.27 | 0.52 | |
Mean of three replicate.
Figure 2Chromatogram of fixed-dose tablet (Duovir, Cipla) solution of Lamivudine and Zidovudine. Mobile phase—toluene:ethyl acetate:methanol (4:4:2, v/v/v), detection at 276 nm.