| Literature DB >> 29354502 |
Seth K Bechis1, Joel E Abbott2, Roger L Sur1.
Abstract
BACKGROUND: To compare head to head two end-engaging nitinol stone retrieval devices available to urologists, in terms of durability, versatility and efficacy.Entities:
Keywords: Kidney stone; durability; nitinol; stone basket; tipless
Year: 2017 PMID: 29354502 PMCID: PMC5760394 DOI: 10.21037/tau.2017.11.30
Source DB: PubMed Journal: Transl Androl Urol ISSN: 2223-4683
Figure 1The two commercially available end-engaging Nitinol stone retrieval devices, NGage™ (Cook Urological Inc., Bloomington, IN, USA) and Dakota™ (Boston Scientific, Natick, MA, USA), hold stones like a basket but engage and release like a grasper. (A) Closeup of the NGage in open position; (B) opening of the Dakota at full deflection of the ureteroscope; (C,D) Dakota with OpenSure™ handle in normal open basket position (C) and with handle engaged in full open position (D).
Figure 2Durability, versatility and efficacy testing. (A) The basket was passed through a simulated endoscopic working channel and cycled in grasping and releasing synthetic stones; (B) red arrows depict areas of visible basket breakdown in baskets inspected after repetitive use; (C) stone model showing 1 and 9 mm synthetic model stones.
Efficacy testing: ability of 8 mm baskets to capture and release various stones
| 8 mm devices (N=15) | 7 mm stone | 8 mm stone | 9 mm stone | |||||
|---|---|---|---|---|---|---|---|---|
| Dakota | NGage | Dakota | NGage | Dakota | NGage | |||
| Capture & elevate stone | 100% (15/15) | 100% (15/15) | 100% (15/15) | 100% (15/15) | 20% (3/15) | Would not capture | ||
| Released stone by gravity | 100% (15/15) | 13% (2/15) | 7% (1/15) | 0% (0/15) | 0% (0/3) | NA | ||
| Retained stone; released by shaking device | NR | 100% (13/13) | 86% (12/14) | 13% (2/15) | 33% (1/3) | NA | ||
| Retained stone; released using OpenSure™ | NR | NA | NR | NA | NR | NA | ||
| Retained stone; manually removed | NR | NR | NR | 87% (13/15) | 67% (2/3) | NA | ||
NA, not applicable; NR, not required.
Efficacy testing: ability of 11 mm baskets to capture and release various stones
| 11 mm devices (N=15) | 9 mm stone | 10 mm stone | 11 mm stone | |||||
|---|---|---|---|---|---|---|---|---|
| Dakota | NGage | Dakota | NGage | Dakota | NGage | |||
| Capture & elevate stone | 100% (15/15) | 100% (15/15) | 100% (15/15) | 100% (15/15) | 100% (15/15) | Would not capture | ||
| Released stone by gravity | 100% (15/15) | 67% (10/15) | 100% (15/15) | 7% (1/15) | 0% (0/15) | NA | ||
| Retained stone; released by shaking device | NR | 100% (5/5) | NR | 100% (14/14) | 100% (15/15) | NA | ||
| Retained stone; released using OpenSure™ | NR | NA | NR | NA | NR | NA | ||
| Retained stone; manually removed | NR | NR | NR | NR | NR | NA | ||
NA, not applicable; NR, not required.