Literature DB >> 29345557

Co-registration of cone beam CT and preoperative MRI for improved accuracy of electrode localization following cochlear implantation.

A S Dragovic1, A K Stringer1, L Campbell2, C Shaul1, S J O'Leary1, R J Briggs2.   

Abstract

OBJECTIVES: To investigate the clinical usefulness and practicality of co-registration of Cone Beam CT (CBCT) with preoperative Magnetic Resonance Imaging (MRI) for intracochlear localization of electrodes after cochlear implantation.
METHODS: Images of 20 adult patients who underwent CBCT after implantation were co-registered with preoperative MRI scans. Time taken for co-registration was recorded. The images were analysed by clinicians of varying levels of expertise to determine electrode position and ease of interpretation.
RESULTS: After a short learning curve, the average co-registration time was 10.78 minutes (StdDev 2.37). All clinicians found the co-registered images easier to interpret than CBCT alone. The mean concordance of CBCT vs. co-registered image analysis between consultant otologists was 60% (17-100%) and 86% (60-100%), respectively. The sensitivity and specificity for CBCT to identify Scala Vestibuli insertion or translocation was 100 and 75%, respectively. The negative predictive value was 100%. DISCUSSION: CBCT should be performed following adult cochlear implantation for audit and quality control of surgical technique. If SV insertion or translocation is suspected, co-registration with preoperative MRI should be performed to enable easier analysis. There will be a learning curve for this process in terms of both the co-registration and the interpretation of images by clinicians.

Entities:  

Keywords:  Co-registration; Cochlear implantation; Cone beam CT; Electrode localization; Preoperative MRI; Scala tympani; Scala vestibuli; Translocation

Mesh:

Year:  2018        PMID: 29345557     DOI: 10.1080/14670100.2017.1419548

Source DB:  PubMed          Journal:  Cochlear Implants Int        ISSN: 1467-0100


  3 in total

1.  Cone-beam CT versus Multidetector CT in Postoperative Cochlear Implant Imaging: Evaluation of Image Quality and Radiation Dose.

Authors:  R A Helal; R Jacob; M A Elshinnawy; A I Othman; I M Al-Dhamari; D W Paulus; T T Abdelaziz
Journal:  AJNR Am J Neuroradiol       Date:  2021-01-07       Impact factor: 3.825

2.  Best Fit 3D Basilar Membrane Reconstruction to Routinely Assess the Scalar Position of the Electrode Array after Cochlear Implantation.

Authors:  Renato Torres; Jean-Yves Tinevez; Hannah Daoudi; Ghizlene Lahlou; Neil Grislain; Eugénie Breil; Olivier Sterkers; Isabelle Mosnier; Yann Nguyen; Evelyne Ferrary
Journal:  J Clin Med       Date:  2022-04-07       Impact factor: 4.241

3.  Evaluation of artifacts of cochlear implant electrodes in cone beam computed tomography.

Authors:  Nicholas Bevis; Thomas Effertz; Dirk Beutner; Christian Gueldner
Journal:  Eur Arch Otorhinolaryngol       Date:  2020-07-15       Impact factor: 2.503