| Literature DB >> 29339025 |
Vanessa Luisa Destro Fidêncio1, Regina Tangerino de Souza Jacob2, Liége Franzini Tanamati3, Érika Cristina Bucuvic3, Adriane Lima Mortari Moret2.
Abstract
INTRODUCTION: The frequency modulation system is a device that helps to improve speech perception in noise and is considered the most beneficial approach to improve speech recognition in noise in cochlear implant users. According to guidelines, there is a need to perform a check before fitting the frequency modulation system. Although there are recommendations regarding the behavioral tests that should be performed at the fitting of the frequency modulation system to cochlear implant users, there are no published recommendations regarding the electroacoustic test that should be performed.Entities:
Keywords: Assistive technology; Cochlear implant; Implante coclear; Pediatria; Pediatric; Percepção da fala; Speech perception; Tecnologia assistiva
Mesh:
Year: 2017 PMID: 29339025 PMCID: PMC9452272 DOI: 10.1016/j.bjorl.2017.11.001
Source DB: PubMed Journal: Braz J Otorhinolaryngol ISSN: 1808-8686
FM gain recommended for each speech processor model automatically calculated by the Phonak “FM SuccessWare” software, version 4.6.3.
| Speech processor model | Brand | FM gain recommended |
|---|---|---|
| Harmony | Advanced Bionics | +10 dB |
| Freedom | Cochlear | 0 dB |
| Nucleus 5 | Cochlear | +2 dB |
| Opus 2 | MED-EL | +10 dB |
Step 1: step-by-step electroacoustic evaluation test – IC only.
| a | Connect the IC speech processor to its respective listening check device |
| b | Place one side of the earphone (listening check device) on the HA-1 coupler through the use of putty and place them in the Verifit test box |
| c | In the Verifit choose: “test” – “test box measures” – “speechmap” |
| d | Place the IC speech processor in the test box with the microphone next to the reference microphone (Verifit) |
| e | Close the equipment and select the stimulus intensity 65 dB SPL and stimulus type “Speech-ISTS” |
| f | Writhe down the values obtained in the frequencies of 750, 1000 and 2000 Hz and calculate a mean of these values |
Step 2: step-by-step electroacoustic evaluation test – IC + FM system.
| a | Adjust the gain of FM system (company's software) |
| b | Plug the FM system receiver into listening check device of the speech processor |
| c | To put the speech processor in the sound attenuation box |
| d | Put the microphone of the FM system transmitter in the test box (Verifit) near the reference microphone |
| e | In the Verifit choose: “test” – “test box measures” – “speechmap” |
| f | Close the equipment and select the stimulus intensity 65 dB SPL and stimulus type “Speech-ISTS” |
| g | Writhe down the values obtained in the frequencies of 750, 1000 and 2000 Hz and calculate a mean of these values |
Results of the electroacoustic evaluation.
| Participant | Speech processor model | Mean of CI microphone (dB) | Mean of CI + FM microphone (dB) | Difference (dB) |
|---|---|---|---|---|
| 1 | Harmony | 74.6 | 75 | 0.4 |
| 2 | 79 | 78 | 1 | |
| 3 | 78.3 | 80.3 | 2 | |
| 4 | 75 | 79.3 | ||
| 5 | 87 | 85.3 | 1.7 | |
| 6 | 81.3 | 80.6 | 0.7 | |
| 7 | Opus 2 | 72.6 | 75.6 | 3 |
| 8 | 67.6 | 69 | 1.4 | |
| 9 | 71 | 74.6 | ||
| 10 | 72 | 75 | 3 | |
| 11 | 73.3 | 71.3 | 2 | |
| 12 | 74 | 73.6 | 0.4 | |
| 13 | 78.6 | 77.3 | 1.3 | |
| 14 | 75.3 | 74.4 | 0.9 | |
| 15 | 68.6 | 64 | ||
| 16 | 73 | 76.6 | ||
| 17 | 64.6 | 67.3 | 2.7 | |
| 18 | 78.6 | 82 | ||
| 19 | 72.6 | 75.3 | 2.7 | |
| 20 | 66 | 67.3 | 1.3 | |
| 21 | 77.3 | 78 | 0.7 | |
| 22 | 74 | 75 | 1 | |
| 23 | 70.6 | 72.6 | 2 | |
| 24 | 79.6 | 82 | 2.4 | |
| 25 | 54 | 53.6 | 0.4 | |
| 26 | 78 | 79.6 | 1.6 | |
| 27 | 78.6 | 80.6 | 2 | |
| 28 | 71.6 | 74.6 | 3 | |
| 29 | 70.6 | 71.6 | 1 | |
| 30 | 73 | 76 | 3 | |
| 31 | Freedom | 75.3 | 76.6 | 1.3 |
| 32 | 73.6 | 76 | 2.4 | |
| 33 | 68.6 | 75 | ||
| 34 | Nucleus 5 | 71.3 | 68.6 | 2.7 |
| 35 | 65.6 | 67.3 | 1.7 | |
| 36 | 73.3 | 70.3 | 3 | |
| 37 | 78.6 | 76.3 | 2.3 | |
| 38 | 69.6 | 72 | 2.4 | |
| 39 | 70 | 69.6 | 0.4 | |
| 40 | 84.3 | 82.6 | 1.7 | |
CI, cochlear implant; FM, frequency modulation system.
Description of the results before and after new FM gain adjustment.
| P | Initial gain (dB) | Result of the electroacoustic evaluation | Speech perception in noise test | Gain after adjustment (dB) | Result of the electroacoustic evaluation | Speech perception in noise test |
|---|---|---|---|---|---|---|
| 4 | +10 | 4.3 | 0 | +8 | 1.7 | −1.5 |
| 9 | +10 | 3.6 | – | – | – | – |
| 15 | +10 | 4.6 | – | +12 | 1.3 | – |
| 16 | +10 | 3.6 | +6 | +8 | 2 | −1.5 |
| 18 | +10 | 3.4 | 0 | +8 | 0.3 | −1.5 |
| 33 | 0 | 6.4 | – | – | – | – |
P, participant.
With initial gain.
After adjustment of gain.