Computer simulated modeling of healthy and diseased right ventricular and pulmonary circulation.

Jody Chou1, Joseph B Rinehart2.   

Abstract

We have previously developed a simulated cardiovascular physiology model for in-silico testing and validation of novel closed-loop controllers. To date, a detailed model of the right heart and pulmonary circulation was not needed, as previous controllers were not intended for use in patients with cardiac or pulmonary pathology. With new development of controllers for vasopressors, and looking forward, for combined vasopressor-fluid controllers, modeling of right-sided and pulmonary pathology is now relevant to further in-silico validation, so we aimed to expand our existing simulation platform to include these elements. Our hypothesis was that the completed platform could be tuned and stabilized such that the distributions of a randomized sample of simulated patients' baseline characteristics would be similar to reported population values. Our secondary outcomes were to further test the system in representing acute right heart failure and pulmonary artery hypertension. After development and tuning of the right-sided circulation, the model was validated against clinical data from multiple previously published articles. The model was considered 'tuned' when 100% of generated randomized patients converged to stability (steady, physiologically-plausible compartmental volumes, flows, and pressures) and 'valid' when the means for the model data in each health condition were contained within the standard deviations for the published data for the condition. A fully described right heart and pulmonary circulation model including non-linear pressure/volume relationships and pressure dependent flows was created over a 6-month span. The model was successfully tuned such that 100% of simulated patients converged into a steady state within 30 s. Simulation results in the healthy state for central venous volume (3350 ± 132 ml) pulmonary blood volume (405 ± 39 ml), pulmonary artery pressures (systolic 20.8 ± 4.1 mmHg and diastolic 9.4 ± 1.8 mmHg), left atrial pressure (4.6 ± 0.8 mmHg), PVR (1.0 ± 0.2 wood units), and CI (3.8 ± 0.5 l/min/m2) all met criteria for acceptance of the model, though the standard deviations of LAP and CI were somewhat narrower than published comparators. The simulation results for right ventricular infarction also fell within the published ranges: pulmonary blood volume (727 ± 102 ml), pulmonary arterial pressures (30 ± 4 mmHg systolic, 12 ± 2 mmHg diastolic), left atrial pressure (13 ± 2 mmHg), PVR (1.6 ± 0.3 wood units), and CI (2.0 ± 0.4 l/min/m2) all fell within one standard deviation of the reported population values and vice-versa. In the pulmonary hypertension model, pulmonary blood volume of 615 ± 90 ml, pulmonary arterial pressures of 80 ± 14 mmHg systolic, 36 ± 7 mmHg diastolic, and the left atrial pressure of 11 ± 2 mmHg all met criteria for acceptance. For CI, the simulated value of 2.8 ± 0.4 l/min/m2 once again had a narrower spread than most of the published data, but fell inside of the SD of all published data, and the PVR value of 7.5 ± 1.6 wood units fell in the middle of the four published studies. The right-ventricular and pulmonary circulation simulation appears to be a reasonable approximation of the right-sided circulation for healthy physiology as well as the pathologic conditions tested.

Entities:  

Keywords:  Computer modeling; Pulmonary circulation; Pulmonary hypertension; Right ventricular infarction; Right-heart; Simulation

Mesh:

Year:  2018        PMID: 29330837     DOI: 10.1007/s10877-018-0099-2

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  22 in total

1.  DETERMINATION OF PULMONARY BLOOD VOLUME BY SINGLE INTRAVENOUS INJECTION OF ONE INDICATOR IN PATIENTS WITH NORMAL AND HIGH PULMONARY VASCULAR PRESSURES.

Authors:  F M DEFREITAS; E Z FARACO; N NEDEL; D F DEAZEVEDO; J ZADUCHLIVER
Journal:  Circulation       Date:  1964-09       Impact factor: 29.690

2.  Relationship of pulmonary arterial capacitance and mortality in idiopathic pulmonary arterial hypertension.

Authors:  Srijoy Mahapatra; Rick A Nishimura; Paul Sorajja; Stephen Cha; Michael D McGoon
Journal:  J Am Coll Cardiol       Date:  2006-01-26       Impact factor: 24.094

3.  Closed-loop fluid resuscitation: robustness against weight and cardiac contractility variations.

Authors:  Joseph Rinehart; Christine Lee; Maxime Cannesson; Guy Dumont
Journal:  Anesth Analg       Date:  2013-11       Impact factor: 5.108

4.  Feasibility of automated titration of vasopressor infusions using a novel closed-loop controller.

Authors:  Joseph Rinehart; Michael Ma; Michael-David Calderon; Maxime Cannesson
Journal:  J Clin Monit Comput       Date:  2017-01-25       Impact factor: 2.502

5.  Right ventricular infarction. Clinical and hemodynamic features.

Authors:  J N Cohn; N H Guiha; M I Broder; C J Limas
Journal:  Am J Cardiol       Date:  1974-02       Impact factor: 2.778

6.  Pulmonary artery reflection for differentially diagnosing primary pulmonary hypertension and chronic pulmonary thromboembolism.

Authors:  Y Nakayama; N Nakanishi; T Hayashi; N Nagaya; F Sakamaki; N Satoh; H Ohya; S Kyotani
Journal:  J Am Coll Cardiol       Date:  2001-07       Impact factor: 24.094

7.  Quantification of right ventricular afterload in patients with and without pulmonary hypertension.

Authors:  Jan-Willem Lankhaar; Nico Westerhof; Theo J C Faes; Koen M J Marques; J Tim Marcus; Piet E Postmus; Anton Vonk-Noordegraaf
Journal:  Am J Physiol Heart Circ Physiol       Date:  2006-05-12       Impact factor: 4.733

8.  Radionuclide determined pulmonary blood volume in ischaemic heart disease.

Authors:  W J Hannan; J Vojacek; H M Connell; N G Dewhurst; A L Muir
Journal:  Thorax       Date:  1981-12       Impact factor: 9.139

9.  Evaluation of a novel closed-loop fluid-administration system based on dynamic predictors of fluid responsiveness: an in silico simulation study.

Authors:  Joseph Rinehart; Brenton Alexander; Yannick Le Manach; Christoph Hofer; Benoit Tavernier; Zeev N Kain; Maxime Cannesson
Journal:  Crit Care       Date:  2011-11-23       Impact factor: 9.097

10.  Pharmacokinetic-pharmacodynamic modelling of the cardiovascular effects of drugs - method development and application to magnesium in sheep.

Authors:  Richard N Upton; Guy L Ludbrook
Journal:  BMC Pharmacol       Date:  2005-03-10
View more
  2 in total

1.  Closed-loop vasopressor control: in-silico study of robustness against pharmacodynamic variability.

Authors:  Joseph Rinehart; Alexandre Joosten; Michael Ma; Michael-David Calderon; Maxime Cannesson
Journal:  J Clin Monit Comput       Date:  2018-12-11       Impact factor: 2.502

2.  Semi-mechanistic modelling platform to assess cardiac contractility and haemodynamics in preclinical cardiovascular safety profiling of new molecular entities.

Authors:  Raja Venkatasubramanian; Teresa A Collins; Lawrence J Lesko; Jerome T Mettetal; Mirjam N Trame
Journal:  Br J Pharmacol       Date:  2020-06-18       Impact factor: 8.739

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.