Assessment of test methods for evaluating effectiveness of cleaning flexible endoscopes.

Rebecca E Washburn1, Jennifer J Pietsch2.   

Abstract

BACKGROUND: Strict adherence to each step of reprocessing is imperative to removing potentially infectious agents. Multiple methods for verifying proper reprocessing exist; however, each presents challenges and limitations, and best practice within the industry has not been established. Our goal was to evaluate endoscope cleaning verification tests with particular interest in the evaluation of the manual cleaning step. The results of the cleaning verification tests were compared with microbial culturing to see if a positive cleaning verification test would be predictive of microbial growth.
METHODS: This study was conducted at 2 high-volume endoscopy units within a multisite health care system. Each of the 90 endoscopes were tested for adenosine triphosphate, protein, microbial growth via agar plate, and rapid gram-negative culture via assay. The endoscopes were tested in 3 locations: the instrument channel, control knob, and elevator mechanism.
RESULTS: This analysis showed substantial level of agreement between protein detection postmanual cleaning and protein detection post-high-level disinfection at the control head for scopes sampled sequentially.
CONCLUSIONS: This study suggests that if protein is detected postmanual cleaning, there is a significant likelihood that protein will also be detected post-high-level disinfection. It also infers that a cleaning verification test is not predictive of microbial growth.
Copyright © 2018 Association for Professionals in Infection Control and Epidemiology, Inc. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  ATP; Cleaning verification; culture; endoscope; manual cleaning verification; protein

Mesh:

Substances:

Year:  2018        PMID: 29329921     DOI: 10.1016/j.ajic.2017.11.014

Source DB:  PubMed          Journal:  Am J Infect Control        ISSN: 0196-6553            Impact factor:   2.918


  3 in total

1.  Turbulent fluid flow is a novel closed-system sample extraction method for flexible endoscope channels of various inner diameters.

Authors:  Seo Yean Sohn; Michelle J Alfa; Richard Lai; Yacoob Tabani; Mohamed E Labib
Journal:  J Microbiol Methods       Date:  2019-11-20       Impact factor: 2.363

2.  No relation between adenosine triphosphate after manual cleaning and presence of microorganisms on endoscopes after automated high-level disinfection.

Authors:  Judith A Kwakman; Arjan W Rauwers; Jolanda G Buijs; Woutrinus de Groot; Margreet C Vos; Marco J Bruno
Journal:  Endosc Int Open       Date:  2022-09-14

3.  Evaluation of an overnight non-culture test for detection of viable Gram-negative bacteria in endoscope channels.

Authors:  Harminder Singh; Donald R Duerksen; Gale Schultz; Carol Reidy; Pat DeGagne; Nancy Olson; Zoann Nugent; Michelle J Alfa
Journal:  Endosc Int Open       Date:  2019-01-30
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.