K Filipovic1, B Bellmann2, J Lüker2, D Steven2, A Sultan2.
Abstract
We report a case of a 85-year old woman with a preexisting Transcatheter Pacing System (TPS) (Micra™ VR, Fa. Medtronic, Inc., Minneapolis, MN, USA) undergoing several external electrical cardioversions (CV) for symptomatic persistent atrial fibrillation (persAF). Due to bradycardia in the setting of atrial fibrillation a right apical TPS implantation was performed earlier. Four weeks prior to presentation at our facility an unsuccessful CV with a maximum biphasic energy level of 360J was performed, after which amiodarone was initiated. At the time of presentation three shocks with 100 J, 200 J and 360 J were delivered without sustained restoration of a stable sinus rhythm. Patches were in an anterior-posterior position. No complications and no significant changes in device parameters in comparison to the pre-acquired values were observed. To our knowledge, this is the first case report of an external CV in a patient with a TPS. External CV in patients with a preexisting TPS seems to be safe and feasible.Entities:
Keywords: Atrial fibrillation; Cardioversion; Leadless pacemaker
Year: 2017 PMID: 29269168 PMCID: PMC5840851 DOI: 10.1016/j.ipej.2017.12.002
Source DB: PubMed Journal: Indian Pacing Electrophysiol J ISSN: 0972-6292
Fig. 1Fig. 1 shows the position of the intracardiac pacemaker device in the right ventricle on a standard chest X-ray one day after implantation of the device.
Fig. 212-lead-ECG recordings prior to (a) and immediately after (b) external CV as well as intracardiac electrograms during testing of pacing (c) and sensing (d) functions immediately after CV are shown.
Table 1 presents the measurements of the TPS before and after the current CV. The initial parameters after device implantation, as well as parameters after the first CV four weeks prior were added for comparison. The pacemaker was programmed to VVIR mode with a lower limiting frequency of 55 beats per minute. The ventricular pacing percentage was 77.9%.
| Post-Implantation | Index CV 4 weeks earlier | Before CV | After CV | 2 weeks follow-up | |
|---|---|---|---|---|---|
| Battery voltage (V) | 3.15 | 3.17 | 3.13 | 3.13 | 3.11 |
| Sensing (mV) | 5.6 | 6.1 | 6.4 | 6.4 | 7.3 |
| Impedance (Ohm) | 510 | 560 | 570 | 570 | 590 |
| Stimulation threshold | 0.88V/0.24 | 0.50V/0.24 ms | 0.50V/0.24 ms | 0.50V/0.24 ms | 0.50V/0.20 ms |