Home blood pressure-measuring devices: a comparative study of accuracy.

C E Evans1, R B Haynes, C H Goldsmith, S A Hewson.   

Abstract

Self monitoring of blood pressure by hypertensive patients has been shown to be of value and is increasingly popular. However, the accuracy and ease of use of at least some of the equipment for home use is questionable. We tested 23 pairs of home blood pressure devices, using static methods, trained observers with volunteer subjects and hypertensive patients. Eleven (48%) of the 23 pairs of devices tested were found inconsistent with their duplicate and failed the standards for automated devices of the Association for the Advancement of Medical Instrumentation. Several of the devices, all sold for home use by lay people, contained no instructions whatsoever and even patients using those machines with instructions required further training in blood pressure measurement. While home blood pressure monitoring is useful in certain circumstances, care must be taken in the choice of device, and training is required for all devices.

Entities:  

Mesh:

Year:  1989        PMID: 2926132

Source DB:  PubMed          Journal:  J Hypertens        ISSN: 0263-6352            Impact factor:   4.844


  4 in total

Review 1.  Economic factors in the initiation of antihypertensive therapy.

Authors:  I Kawachi
Journal:  Pharmacoeconomics       Date:  1992-10       Impact factor: 4.981

2.  Cardiovascular reactivity to a naturally occurring stressor: development and psychometric evaluation of a psychophysiological assessment procedure.

Authors:  R L Hazlett; S Falkin; W Lawhorn; E Friedman; S N Haynes
Journal:  J Behav Med       Date:  1997-12

Review 3.  Modern approaches to blood pressure measurement.

Authors:  J A Staessen; E T O'Brien; L Thijs; R H Fagard
Journal:  Occup Environ Med       Date:  2000-08       Impact factor: 4.402

Review 4.  Report of the Canadian Hypertension Society Consensus Conference: 2. Diagnosis of hypertension in adults.

Authors:  R B Haynes; Y Lacourcière; S W Rabkin; F H Leenen; A G Logan; N Wright; C E Evans
Journal:  CMAJ       Date:  1993-08-15       Impact factor: 8.262

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.