Radiographic comparison of finned, cementless central pegged glenoid component and conventional cemented pegged glenoid component in total shoulder arthroplasty: a prospective randomized study.

Christopher M Kilian1, Brent J Morris2, Kyle R Sochacki3, Mufaddal M Gombera2, Richard E Haigler4, Daniel P O'Connor5, T Bradley Edwards2.   

Abstract

BACKGROUND: Radiographic lucency of the glenoid component remains a problem after cement fixation in primary total shoulder arthroplasty. Glenoid component design likely contributes to rates of glenoid lucency. The purpose of this study was to prospectively compare radiographic lucency between a finned, cementless central pegged glenoid component (CL component) and a conventional cemented pegged glenoid component (P component) on immediate postoperative and minimum 2-year follow-up radiographs.
METHODS: Fifty-four patients undergoing total shoulder arthroplasty were prospectively randomized to receive an all-polyethylene CL component or a conventional all-polyethylene P component. Three raters graded glenoid lucency and bone interdigitation on immediate postoperative and latest follow-up radiographs. Patients who had undergone revision surgery or had died before evaluation were excluded. Minimum 2-year follow-up was required for inclusion of radiographic evaluation.
RESULTS: Fifty patients met inclusion criteria; 42 patients (84%; 20 CL and 22 P) were available for follow-up with the original glenoid implant in place. The mean follow-up duration was 35 months (24-64 months). There were no significant differences in glenoid radiolucency between CL (1/20 [5%]) and P (2/22 [9%]) components at last follow-up (P = .999). Five patients (25%) in the CL group had bone interdigitation. No instances of aseptic glenoid loosening occurred.
CONCLUSION: There were no significant differences in the rate of glenoid lucency between the 2 groups at immediate or an average 35-month follow-up. Both techniques appear to be viable options for initial glenoid component fixation, with CL components allowing possible osseointegration, imparting potential long-term stability.
Copyright © 2017 Journal of Shoulder and Elbow Surgery Board of Trustees. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Cementless central pegged; CortiLoc; aseptic loosening, finned peg; pegged glenoid; radiolucency; total shoulder arthroplasty

Mesh:

Substances:

Year:  2017        PMID: 29246679     DOI: 10.1016/j.jse.2017.09.014

Source DB:  PubMed          Journal:  J Shoulder Elbow Surg        ISSN: 1058-2746            Impact factor:   3.019


  5 in total

1.  One and two-year clinical outcomes for a polyethylene glenoid with a fluted peg: one thousand two hundred seventy individual patients from eleven centers.

Authors:  Frederick A Matsen; Joseph P Iannotti; R Sean Churchill; Lieven De Wilde; T Bradley Edwards; Matthew C Evans; Edward V Fehringer; Gordon I Groh; James D Kelly; Christopher M Kilian; Giovanni Merolla; Tom R Norris; Giuseppe Porcellini; Edwin E Spencer; Anne Vidil; Michael A Wirth; Stacy M Russ; Moni Neradilek; Jeremy S Somerson
Journal:  Int Orthop       Date:  2018-12-03       Impact factor: 3.075

2.  Anatomic versus reverse shoulder arthroplasty: a mid-term follow-up comparison.

Authors:  Bradley S Schoch; Joseph J King; Joseph Zuckerman; Thomas W Wright; Chris Roche; Pierre-Henri Flurin
Journal:  Shoulder Elbow       Date:  2020-05-16

3.  Clinical and radiographic outcomes of total shoulder arthroplasty with a partially cemented all-polyethylene pegged bone-ingrowth glenoid component: a systematic review.

Authors:  Michael-Alexander Malahias; Lazaros Kostretzis; Ioannis Gkiatas; Efstathios Chronopoulos; Emmanouil Brilakis; Emmanouil Antonogiannakis
Journal:  Shoulder Elbow       Date:  2020-07-27

4.  Bone Integration and Prevalence of Radiolucent Lines around the Pegs of Minimally Cemented Glenoid Components in Total Shoulder Arthroplasty.

Authors:  Márcio Schiefer; Gláucio Siqueira; Alan Figueira; Patrícia Martins Souza; Martim Teixeira Monteiro; Geraldo Motta Filho
Journal:  Rev Bras Ortop (Sao Paulo)       Date:  2020-09-25

5.  The effect of radial mismatch on radiographic glenoid loosening.

Authors:  Bradley S Schoch; Thomas W Wright; Joseph D Zuckerman; Pierre-Henri Flurin; Charlotte Bolch; Chris P Roche; Joseph J King
Journal:  JSES Open Access       Date:  2019-11-18
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.