Literature DB >> 29230939

Normative values and factors affecting water-perfused esophageal high-resolution impedance manometry for a Chinese population.

P-H Tseng1, R K M Wong2,3, J-F Wu4, C-C Chen1, C-H Tu1, Y-C Lee1, H-C Lee4, H-P Wang1, M-S Wu1.   

Abstract

BACKGROUND: Combined esophageal high-resolution impedance manometry (HRIM) measures multiple pressures and bolus transit simultaneously, facilitating detailed assessment of esophageal motility. Currently, normative values for water-perfused HRIM systems for Chinese populations are lacking.
METHODS: Healthy volunteers were enrolled for comprehensive anthropometric measures, blood biochemistry tests, and an HRIM study using 22 water-perfused pressure sensors and 12 impedance channels. Ten 5-mL liquid swallows of saline at 30-second intervals were conducted. The following parameters were calculated: distal contractile integral (DCI), distal latency (DL), lower esophageal sphincter (LES) basal pressure, 4-second integrated relaxation pressure (IRP-4s), and complete bolus transit percentage. Normal values were established based on the 5th and 95th percentiles. KEY
RESULTS: All 66 participants (34 male, 32 female, aged 21-64 years) completed the study and tolerated the HRIM procedure well. The upper normal limit (95th percentile) of IRP-4 second was 20 mmHg. The 5th-95th percentile range for DCI, DL, and complete bolus transit was 99-2186 mmHg●s●cm, 6.2-11.3 second, and 50%-100%, respectively. Age was negatively correlated with DL. Females had significantly higher upper limits for IRP-4s and median DCI than males. Multivariate analyses confirmed that IRP-4s was higher in females, and that higher body mass index and waist circumference were associated with reduced DL and better bolus transit, respectively. CONCLUSIONS AND INFERENCES: We established normative values for the water-perfused HRIM system for a Chinese population. Gender and anthropometric factors may affect various major HRIM parameters and should be taken into account when interpreting HRIM results in clinical practice.
© 2017 John Wiley & Sons Ltd.

Entities:  

Keywords:  anthropometrics; esophageal motility disorder; high-resolution impedance manometry (HRIM); normative values; water-perfused

Mesh:

Substances:

Year:  2017        PMID: 29230939     DOI: 10.1111/nmo.13265

Source DB:  PubMed          Journal:  Neurogastroenterol Motil        ISSN: 1350-1925            Impact factor:   3.598


  2 in total

1.  VALIDATION OF A NEW WATER-PERFUSED HIGH-RESOLUTION MANOMETRY SYSTEM.

Authors:  Rogério Mariotto; Fernando A M Herbella; Vera Lucia Ângelo Andrade; Francisco Schlottmann; Marco G Patti
Journal:  Arq Bras Cir Dig       Date:  2021-01-25

2.  Bolus transit of upper esophageal sphincter on high-resolution impedance manometry study correlate with the laryngopharyngeal reflux symptoms.

Authors:  Jia-Feng Wu; Wei-Chung Hsu; I-Jung Tsai; Tzu-Wei Tong; Yu-Cheng Lin; Chia-Hsiang Yang; Ping-Huei Tseng
Journal:  Sci Rep       Date:  2021-10-14       Impact factor: 4.379