Evaluation of motion artifact metrics for coronary CT angiography.

Hongfeng Ma1, Eric Gros2, Aniko Szabo3, Scott G Baginski4, Zachary R Laste4, Naveen M Kulkarni4, Darin Okerlund2, Taly G Schmidt1.   

Abstract

PURPOSE: This study quantified the performance of coronary artery motion artifact metrics relative to human observer ratings. Motion artifact metrics have been used as part of motion correction and best-phase selection algorithms for Coronary Computed Tomography Angiography (CCTA). However, the lack of ground truth makes it difficult to validate how well the metrics quantify the level of motion artifact. This study investigated five motion artifact metrics, including two novel metrics, using a dynamic phantom, clinical CCTA images, and an observer study that provided ground-truth motion artifact scores from a series of pairwise comparisons.
METHOD: Five motion artifact metrics were calculated for the coronary artery regions on both phantom and clinical CCTA images: positivity, entropy, normalized circularity, Fold Overlap Ratio (FOR), and Low-Intensity Region Score (LIRS). CT images were acquired of a dynamic cardiac phantom that simulated cardiac motion and contained six iodine-filled vessels of varying diameter and with regions of soft plaque and calcifications. Scans were repeated with different gantry start angles. Images were reconstructed at five phases of the motion cycle. Clinical images were acquired from 14 CCTA exams with patient heart rates ranging from 52 to 82 bpm. The vessel and shading artifacts were manually segmented by three readers and combined to create ground-truth artifact regions. Motion artifact levels were also assessed by readers using a pairwise comparison method to establish a ground-truth reader score. The Kendall's Tau coefficients were calculated to evaluate the statistical agreement in ranking between the motion artifacts metrics and reader scores. Linear regression between the reader scores and the metrics was also performed.
RESULTS: On phantom images, the Kendall's Tau coefficients of the five motion artifact metrics were 0.50 (normalized circularity), 0.35 (entropy), 0.82 (positivity), 0.77 (FOR), 0.77(LIRS), where higher Kendall's Tau signifies higher agreement. The FOR, LIRS, and transformed positivity (the fourth root of the positivity) were further evaluated in the study of clinical images. The Kendall's Tau coefficients of the selected metrics were 0.59 (FOR), 0.53 (LIRS), and 0.21 (Transformed positivity). In the study of clinical data, a Motion Artifact Score, defined as the product of FOR and LIRS metrics, further improved agreement with reader scores, with a Kendall's Tau coefficient of 0.65.
CONCLUSION: The metrics of FOR, LIRS, and the product of the two metrics provided the highest agreement in motion artifact ranking when compared to the readers, and the highest linear correlation to the reader scores. The validated motion artifact metrics may be useful for developing and evaluating methods to reduce motion in Coronary Computed Tomography Angiography (CCTA) images.
© 2017 American Association of Physicists in Medicine.

Entities:  

Keywords:  CCTA; cardiac imaging; motion artifacts

Mesh:

Year:  2018        PMID: 29222954      PMCID: PMC5807165          DOI: 10.1002/mp.12720

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  17 in total

1.  Pairwise comparison versus Likert scale for biomedical image assessment.

Authors:  Andrew S Phelps; David M Naeger; Jesse L Courtier; Jack W Lambert; Peter A Marcovici; Javier E Villanueva-Meyer; John D MacKenzie
Journal:  AJR Am J Roentgenol       Date:  2015-01       Impact factor: 3.959

2.  Dual-source CT cardiac imaging: initial experience.

Authors:  Thorsten R C Johnson; Konstantin Nikolaou; Bernd J Wintersperger; Alexander W Leber; Franz von Ziegler; Carsten Rist; Sonja Buhmann; Andreas Knez; Maximilian F Reiser; Christoph R Becker
Journal:  Eur Radiol       Date:  2006-05-13       Impact factor: 5.315

3.  Simultaneous misalignment correction for approximate circular cone-beam computed tomography.

Authors:  Y Kyriakou; R M Lapp; L Hillebrand; D Ertel; W A Kalender
Journal:  Phys Med Biol       Date:  2008-10-20       Impact factor: 3.609

4.  Cardiac motion correction based on partial angle reconstructed images in x-ray CT.

Authors:  Seungeon Kim; Yongjin Chang; Jong Beom Ra
Journal:  Med Phys       Date:  2015-05       Impact factor: 4.071

5.  Improving best-phase image quality in cardiac CT by motion correction with MAM optimization.

Authors:  Christopher Rohkohl; Herbert Bruder; Karl Stierstorfer; Thomas Flohr
Journal:  Med Phys       Date:  2013-03       Impact factor: 4.071

6.  Automatic phase determination for retrospectively gated cardiac CT.

Authors:  R Manzke; Th Köhler; T Nielsen; D Hawkes; M Grass
Journal:  Med Phys       Date:  2004-12       Impact factor: 4.071

7.  The impact of small motion on the visualization of coronary vessels and lesions in cardiac CT: A simulation study.

Authors:  Francisco Contijoch; J Webster Stayman; Elliot R McVeigh
Journal:  Med Phys       Date:  2017-05-26       Impact factor: 4.071

8.  Automated selection of the optimal cardiac phase for single-beat coronary CT angiography reconstruction.

Authors:  D Stassi; S Dutta; H Ma; A Soderman; D Pazzani; E Gros; D Okerlund; T G Schmidt
Journal:  Med Phys       Date:  2016-01       Impact factor: 4.071

9.  A fully four-dimensional, iterative motion estimation and compensation method for cardiac CT.

Authors:  Qiulin Tang; Jochen Cammin; Somesh Srivastava; Katsuyuki Taguchi
Journal:  Med Phys       Date:  2012-07       Impact factor: 4.071

10.  The diagnostic and prognostic value of coronary CT angiography in asymptomatic high-risk patients: a cohort study.

Authors:  Fabian Plank; Guy Friedrich; Wolfgang Dichtl; Andrea Klauser; Werner Jaschke; Wolfgang-Michael Franz; Gudrun Feuchtner
Journal:  Open Heart       Date:  2014-08-12
View more
  4 in total

1.  Motion-corrected coronary calcium scores by a convolutional neural network: a robotic simulating study.

Authors:  Yaping Zhang; Niels R van der Werf; Beibei Jiang; Robbert van Hamersvelt; Marcel J W Greuter; Xueqian Xie
Journal:  Eur Radiol       Date:  2019-10-18       Impact factor: 5.315

2.  Quantitative assessment of motion effects in dual-source dual-energy CT and dual-source photon-counting detector CT.

Authors:  Zaki Ahmed; Kishore Rajendran; Hao Gong; Cynthia McCollough; Shuai Leng
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2022-04-04

3.  Motion correction for routine X-ray lung CT imaging.

Authors:  Doil Kim; Jiyoung Choi; Duhgoon Lee; Hyesun Kim; Jiyoung Jung; Minkook Cho; Kyoung-Yong Lee
Journal:  Sci Rep       Date:  2021-02-12       Impact factor: 4.379

4.  Task-dependent estimability index to assess the quality of cardiac computed tomography angiography for quantifying coronary stenosis.

Authors:  Ehsan Samei; Taylor Richards; William P Segars; Melissa A Daubert; Alex Ivanov; Geoffrey D Rubin; Pamela S Douglas; Udo Hoffmann
Journal:  J Med Imaging (Bellingham)       Date:  2021-01-09
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.