Literature DB >> 29205219

Experimental analysis of insertion torques and forces of threaded and press-fit acetabular cups by means of ex vivo and in vivo measurements.

Danny Vogel1, Andreas Rathay1, Stephanie Teufel1, Martin Ellenrieder1, Carmen Zietz1, Manuela Sander2, Rainer Bader1.   

Abstract

PURPOSE: In THA a sufficient primary implant stability is the precondition for successful secondary stability. Industrial foams of different densities have been used for primary stability investigations. The aim of this study was to analyse and compare the insertion behaviour of threaded and press-fit cups in vivo and ex vivo using bone substitutes with various densities.
METHODS: Two threaded (Bicon Plus®, Trident® TC) and one press-fit cup (Trident PSL®) were inserted by orthopaedic surgeons (S1, S2) into 10, 20 and 31 pcf blocks, using modified surgical instruments allowing measurements of the insertion forces and torques. Furthermore, the insertion behaviour of two cups were analysed intraoperatively.
RESULTS: Torques for the threaded cups increased while bone substitute density increased. Maximum insertion torques were observed for S2 with 102 Nm for the Bicon Plus® in 20 pcf blocks and 77 Nm for the Trident® TC in 31 pcf blocks, which compares to the in vivo measurement (85 Nm). The average insertion forces for the press-fit cup varied from 5.2 to 6.8 kN (S1) and 7.2-11.5 kN (S2) ex vivo. Intraoperatively an average insertion force of 8.0 kN was determined.
CONCLUSIONS: Implantation behaviour was influenced by acetabular cup design, bone substitute and experience of the surgeon. No specific density of bone substitute could be favoured for ex vivo investigations on the implantation behaviour of acetabular cups. The use synthetic bone blocks of high density (31 pcf) led to problems regarding cup orientation and seating. Therefore, bone substitutes used should be critically scrutinized in terms of the comparability to the in vivo situation.

Entities:  

Mesh:

Year:  2017        PMID: 29205219

Source DB:  PubMed          Journal:  Acta Bioeng Biomech        ISSN: 1509-409X            Impact factor:   1.073


  2 in total

1.  Surgeon-stratified cohort analysis of 1976 cementless Zweymüller total hip arthroplasties from a single hospital with 23,255 component years of follow-up.

Authors:  Samo Roškar; Vane Antolič; Blaž Mavčič
Journal:  Arch Orthop Trauma Surg       Date:  2020-06-09       Impact factor: 3.067

2.  How screw connections influence the primary stability of acetabular cups under consideration of different bone models.

Authors:  Gafar Adam Ahmed; Linda Auge; Jessica Loke; Carlos Alfonso Fonseca Ulloa; Christian Fölsch; Markus Rickert; Bernd Alexander Ishaque; Alexander Jahnke
Journal:  J Orthop       Date:  2020-06-06