Impact of a monolithic silicon detector operating in transmission mode on clinical photon beams.

Kananan Utitsarn1, Ziyad A Alrowaili2, Nauljan Stansook3, Marco Petasecca4, Martin Carolan5, Vladimir L Perevertaylo6, Michael Lerch7, Anatoly Rosenfeld8.   

Abstract

PURPOSE: To investigate the effect on surface dose, as a function of different field sizes and distances from the solid water phantom to transmission detector (Dsd), of using the monolithic silicon detector MP512T in transmission mode.
METHODS: The influence of operating the MP512T in transmission mode on the surface dose of a phantom for SSD 100cm was evaluated by using a Markus IC. The MP512T was fixed to an adjustable stand holder and was positioned at different Dsd, ranging from 0.3 to 24 cm. For each Dsd, measurements were carried out for irradiation field sizes of 5 × 5cm2, 8 × 8 cm2 and 10 × 10 cm2. Measurements were obtained under two different operational setups, (i) with the MP512T face-up and (ii) with the MP512T face-down. In addition, the transmission factors for the MP512T and the printed circuit board were only evaluated using a Farmer IC.
RESULTS: For all Dsd and all field sizes, the MP512T led to the surface dose increasing by less than 25% when in the beam. For Dsd >18 cm the surface dose increase is less than 5%, and negligible for field size 5 × 5 cm2. The difference in the surface dose perturbation for the MP512T operating face up or operating face down is negligible (<2%) for all field sizes. The transmission factor of the MP512T ranged from 1.020 to 0.9950 for all measured Dsd and field sizes.
CONCLUSION: The study demonstrated that positioning the MP512T in air between the Linac head and the phantom produced negligible perturbation of the surface dose for Dsd >18 cm, and was completely transparent for 6 MV photon beams.
Copyright © 2017 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2017        PMID: 29195553     DOI: 10.1016/j.ejmp.2017.10.017

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  2 in total

1.  Two-dimensional real-time quality assurance dosimetry system using μ-Al2O3:C,Mg radioluminescence films.

Authors:  Luana F Nascimento; Dirk Verellen; Jo Goossens; Lara Struelens; Filip Vanhavere; Paul Leblans; Mark Akselrod
Journal:  Phys Imaging Radiat Oncol       Date:  2020-10-05

2.  Two-dimensional solid-state array detectors: A technique for in vivo dose verification in a variable effective area.

Authors:  Kananan Utitsarn; Giordano Biasi; Nauljun Stansook; Ziyad A Alrowaili; Marco Petasecca; Martin Carolan; Vladimir L Perevertaylo; Wolfgang A Tomé; Tomas Kron; Michael L F Lerch; Anatoly B Rosenfeld
Journal:  J Appl Clin Med Phys       Date:  2019-10-14       Impact factor: 2.102

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.