Literature DB >> 29194732

Doppler ultrasound triggering for cardiac MRI at 7T.

F Kording1, C Ruprecht1, B Schoennagel1, K Fehrs1, J Yamamura1, G Adam1, J Goebel2,3, K Nassenstein3, S Maderwald2, H H Quick2,4, O Kraff2.   

Abstract

PURPOSE: A prerequisite for cardiac MR (CMR) imaging is adequate synchronization of image acquisition with the cardiac cycle. Electrocardiogram triggering may be hampered by electromagnetic interferences at high field strength. The purpose of this work is to evaluate the feasibility of Doppler ultrasound triggering for CMR image synchronization at 7T ultra-high-field MRI.
METHODS: A custom-built Doppler ultrasound (DUS) trigger device was developed. Magnetic resonance compatibility was evaluated using E- and H-field probes and flip angle maps prior to the study. Cardiac MR was performed at 7T in 13 healthy subjects using DUS and pulse oximetry for triggering. For validation of the trigger signal, the electrocardiogram, pulse, and DUS signals were compared outside of the MR room. Breath-hold cine fast low-angle-shot sequences were acquired in short-axis and four-chamber view. Image quality was assessed by two senior radiologists and by measurement of endocardial blurring.
RESULTS: The maximal change in E- and H-field distributions with and without transducer was 5%. No interferences were observed between DUS and MRI in the B1 maps and during CMR imaging. Validation of the DUS trigger signal resulted in a high correlation to the electrocardiographic signal of r = 0.99. Analysis of image and trigger quality revealed no significant differences.
CONCLUSION: Doppler ultrasound was applied as a new trigger method in CMR at 7T. The transmission line and transducer were locally approved as 7T MR conditional, and were successfully tested for image synchronization at 7T. In the future, this method needs to be evaluated in a larger patient population. Magn Reson Med 80:239-247, 2018.
© 2017 International Society for Magnetic Resonance in Medicine. © 2017 International Society for Magnetic Resonance in Medicine.

Entities:  

Keywords:  7 Tesla; Doppler ultrasound; cardiac triggering; cardiovascular magnetic resonance imaging

Mesh:

Year:  2017        PMID: 29194732     DOI: 10.1002/mrm.27032

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  2 in total

Review 1.  Model checking for trigger loss detection during Doppler ultrasound-guided fetal cardiovascular MRI.

Authors:  Sven-Thomas Antoni; Sascha Lehmann; Maximilian Neidhardt; Kai Fehrs; Christian Ruprecht; Fabian Kording; Gerhard Adam; Sibylle Schupp; Alexander Schlaefer
Journal:  Int J Comput Assist Radiol Surg       Date:  2018-08-04       Impact factor: 2.924

2.  Doppler ultrasound cardiac gating of intracranial flow at 7T.

Authors:  Karin Markenroth Bloch; Fabian Kording; Johannes Töger
Journal:  BMC Med Imaging       Date:  2020-12-09       Impact factor: 1.930

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.