Are the same tapes really the same? Ultrasound study of laser-cut and mechanically cut TVT-O post-operative behavior.

Zdenek Rusavy1,2, Jaromir Masata3, Kamil Svabik3, Petr Hubka3, Karel Zvara4, Alois Martan3.   

Abstract

INTRODUCTION AND HYPOTHESIS: TVT-O production has been modified to laser cutting from mechanical cutting. We compared the behavior of laser and mechanically cut tension-free vaginal tape-obturator (TVT-O) using ultrasound at various time points after surgery.
METHODS: This is a retrospective analysis of clinical and ultrasound data from two previously reported randomized controlled trials with TVT-O. Behavior of mechanically cut TVT-O implanted in January 2007 to November 2009 and laser-cut TVT-O implanted in May 2010 to May 2012 was assessed by ultrasound at day 1, the 2nd week, the 3rd month, and the 1st and 2nd years post-operatively. Bladder neck and tape margins positions were described by coordinates in the orthogonal system calculated from polar coordinates. Tape mobility was measured as a change in the upper and lower tape margin position from rest to maximal Valsalva. Comparison of 2-year subjective and objective surgery outcomes was also performed.
RESULTS: In total, 68 mechanically cut and 50 laser-cut TVT-Os were implanted. Follow-up data were available from 49 and 45 women respectively. No differences in any baseline characteristics or bladder neck mobility were observed. Significantly lower tape mobility was observed on day 1 and week 2 after mechanically cut TVT-O, although subsequent mobility was comparable to laser-cut TVT-O. The subjective and objective surgery outcomes were comparable.
CONCLUSIONS: Although without clinical significance, early postoperative behavior of the mechanically cut and laser-cut TVT-O tapes differs. The less stiff, mechanically-cut TVT-O loosens within 2 weeks of implantation, whereas the stiffer, laser-cut TVT-O keeps its tension.

Entities:  

Keywords:  TVT-O; Tape elongation properties; Tape mobility; Tape stiffness; Ultrasound

Mesh:

Year:  2017        PMID: 29177544     DOI: 10.1007/s00192-017-3516-z

Source DB:  PubMed          Journal:  Int Urogynecol J        ISSN: 0937-3462            Impact factor:   2.894


  13 in total

Review 1.  Are biomechanical properties predictive of the success of prostheses used in stress urinary incontinence and pelvic organ prolapse? A systematic review.

Authors:  Altaf Mangera; Anthony J Bullock; Christopher R Chapple; Sheila Macneil
Journal:  Neurourol Urodyn       Date:  2011-10-28       Impact factor: 2.696

2.  Influence of TVT properties on outcomes of midurethral sling procedures: high-stiffness versus low-stiffness tape.

Authors:  Jens Christian Prien-Larsen; Thomas Prien-Larsen; Lars Cieslak; Ram B Dessau
Journal:  Int Urogynecol J       Date:  2016-01-15       Impact factor: 2.894

3.  Randomized trial of a comparison of the efficacy of TVT-O and single-incision tape TVT SECUR systems in the treatment of stress urinary incontinent women--2-year follow-up.

Authors:  Jaromir Masata; Kamil Svabik; Karel Zvara; Petra Drahoradova; Rachid El Haddad; Petr Hubka; Alois Martan
Journal:  Int Urogynecol J       Date:  2012-10       Impact factor: 2.894

4.  Tensile properties of five commonly used mid-urethral slings relative to the TVT.

Authors:  Pamela A Moalli; Noah Papas; Shawn Menefee; Mike Albo; Leslie Meyn; Steven D Abramowitch
Journal:  Int Urogynecol J Pelvic Floor Dysfunct       Date:  2008-01-09

Review 5.  Simultaneous inference in general parametric models.

Authors:  Torsten Hothorn; Frank Bretz; Peter Westfall
Journal:  Biom J       Date:  2008-06       Impact factor: 2.207

Review 6.  The biology behind fascial defects and the use of implants in pelvic organ prolapse repair.

Authors:  Jan Deprest; Fang Zheng; Maja Konstantinovic; Federico Spelzini; Filip Claerhout; Anneke Steensma; Yves Ozog; Dirk De Ridder
Journal:  Int Urogynecol J Pelvic Floor Dysfunct       Date:  2006-06

Review 7.  Single-incision sling operations for urinary incontinence in women.

Authors:  Arjun Nambiar; June D Cody; Stephen T Jeffery; Patricia Aluko
Journal:  Cochrane Database Syst Rev       Date:  2017-07-26

8.  Comparison of the efficacy of tension-free vaginal tape obturator (TVT-O) and single-incision tension-free vaginal tape (Ajust™) in the treatment of female stress urinary incontinence: a 1-year follow-up randomized trial.

Authors:  Jaromir Masata; Kamil Svabik; Karel Zvara; Petr Hubka; Ales Toman; Alois Martan
Journal:  Int Urogynecol J       Date:  2016-04-01       Impact factor: 2.894

9.  The 'iris effect': how two-dimensional and three-dimensional ultrasound can help us understand anti-incontinence procedures.

Authors:  H P Dietz; P D Wilson
Journal:  Ultrasound Obstet Gynecol       Date:  2004-03       Impact factor: 7.299

10.  Novel surgical technique for the treatment of female stress urinary incontinence: transobturator vaginal tape inside-out.

Authors:  Jean de Leval
Journal:  Eur Urol       Date:  2003-12       Impact factor: 20.096