| Literature DB >> 29172387 |
Kohei Kawaguchi1, Kazuhiko Michishita1, Takeshi Manabe1, Yoshiyuki Akasaka1, Junya Higuchi1.
Abstract
PURPOSE: The KneeAlign2 (KA2, OrthoAlign Inc.) accelerometer-based portable navigation system and patient-specific instrumentation (PSI; Signature, ZimmerBiomet) are widely used for ideal femoral component alignment in total knee arthroplasty (TKA). However, there has been no comparative study of the KA2 system, PSI, and conventional intramedullary instrumentation (CON). The purpose of this study was to compare the accuracy in achieving proper femoral component alignment and clinical features by using the KA2 navigation system, PSI, and CON.Entities:
Keywords: Arthroplasty; Femoral component alignment; Knee; Navigation; Patient-specific instrumentation
Year: 2017 PMID: 29172387 PMCID: PMC5718798 DOI: 10.5792/ksrr.17.032
Source DB: PubMed Journal: Knee Surg Relat Res ISSN: 2234-0726
Fig. 1(A) Anterioposterior radiograph of the entire lower limb after total knee arthroplasty. (B) Lateral radiograph of the knee. CFA: coronal femoral angle, HKA: hip-knee-ankle angle, CTA: coronal tibial angle, SFA: sagittal femoral angle, STA: sagittal tibial angle.
Preoperative Patient Characteristics
| Characteristic | KA2 (n=32) | PSI (n=31) | CON (n=32) | p-value |
|---|---|---|---|---|
| Age (yr) | 76.0±5.2 | 74.2±8.0 | 72.3±5.1 | 0.059 |
| Sex (female:male) | 26:6 | 27:4 | 25:7 | 0.196 |
| BMI (kg/m2) | 25.3±2.5 | 25.9±3.2 | 27.2±3.5 | 0.057 |
| Preop HKA (°) | 167.2±4.8 | 167.7±4.5 | 168.1±6.1 | 0.404 |
| Preop ROM (°) | 111.1±16.1 | 114.8±17.4 | 107.5±22.8 | 0.158 |
| Preop KSS | 46.2±8.4 | 48.8±3.8 | 47.3±5.9 | 0.266 |
| Preop Knee function score | 42.9±21.3 | 47.2±9.4 | 44.3±12.9 | 0.528 |
Values are presented as mean±standard deviation.
KA2: KneeAlign2, PSI: patient-specific instrumentation, CON: conventional intramedullary instrument, BMI: body mass index, HKA: hip-knee-ankle angle, ROM: range of motion, KSS: Knee Society score.
Individual Component Position and Limb Alignment
| Characteristic | KA2 | PSI | CON | p-value |
|---|---|---|---|---|
| Postoperative HKA (°) | 178.9±1.6 | 178.5±2.2 | 177.6±2.2 | <0.01 |
| Coronal femoral angle (°) | 89.1±1.1 | 88.1±1.9 | 87.6±2.1 | <0.01 |
| Coronal tibial angle (°) | 89.3±1.6 | 86.7±2.6 | 89.6±1.8 | <0.01 |
Values are presented as mean±standard deviation.
In the coronal plane, a positive value indicates varus alignment and a negative value indicates valgus alignment on the femoral side. In the sagittal plane, a positive value indicates flexion on the femoral side.
KA2: KneeAlign2, PSI: patient-specific instrumentation, CON: conventional intramedullary instrument, HKA: hip-knee-ankle angle.
Statistically significant (p<0.05).
Outliers in Femoral Component Positions
| Characteristic | KA2 (%) | PSI (%) | CON (%) | p-value |
|---|---|---|---|---|
| Hip-knee-ankle angle (°) | 9.1 | 22.5 | 28.1 | 0.283 |
| Coronal plane | ||||
| Femoral component | 0 | 12.9 | 31.3 | <0.01 |
| Sagittal plane | ||||
| Femoral component | 9.0 | 37.5 | 0 | <0.01 |
Deviation of >3° from the target in any plane was defined as an outlier.
KA2: KneeAlign2, PSI: patient-specific instrumentation, CON: conventional intramedullary instrument.
Statistically significant (p<0.05).
Postoperative Characteristics
| Characteristic | KA2 | PSI | CON | p-value |
|---|---|---|---|---|
| Surgical duration (min) | 123.3±20.8 | 114.1±20.6 | 121.1±18.3 | 0.080 |
| Postoperative estimated blood loss (mL) | 422.5±187.5 | 402.2±189.3 | 563.6±261.9 | <0.01 |
| Postoperative ROM (°) | 122.3±9.9 | 125.2±13.7 | 123.5±9.6 | 0.298 |
| Postoperative KSS | 92.1±6.7 | 91.8±8.0 | 91.4±8.0 | 0.469 |
| Postoperative knee function score | 75.7±12.3 | 74.8±15.2 | 78.4±14.1 | 0.281 |
Values are presented as mean±standard deviation.
KA2: KneeAlign2, PSI: patient-specific instrumentation, CON: conventional intramedullary instrument, ROM: range of motion, KSS: Knee Society score.
Statistically significant (p<0.05).