| Literature DB >> 29168773 |
Shintaro Kurasawa1, Shouhei Koyama2, Hiroaki Ishizawa3, Keisaku Fujimoto4, Shun Chino5.
Abstract
This paper describes and verifies a non-invasive blood glucose measurement method using a fiber Bragg grating (FBG) sensor system. The FBG sensor is installed on the radial artery, and the strain (pulse wave) that is propagated from the heartbeat is measured. The measured pulse wave signal was used as a collection of feature vectors for multivariate analysis aiming to determine the blood glucose level. The time axis of the pulse wave signal was normalized by two signal processing methods: the shortest-time-cut process and 1-s-normalization process. The measurement accuracy of the calculated blood glucose level was compared with the accuracy of these signal processing methods. It was impossible to calculate a blood glucose level exceeding 200 mg/dL in the calibration curve that was constructed by the shortest-time-cut process. In the 1-s-normalization process, the measurement accuracy of the blood glucose level was improved, and a blood glucose level exceeding 200 mg/dL could be calculated. By verifying the loading vector of each calibration curve to calculate the blood glucose level with a high measurement accuracy, we found the gradient of the peak of the pulse wave at the acceleration plethysmogram greatly affected.Entities:
Keywords: blood flow; blood glucose level; fiber Bragg grating; non-invasive measurement; partial least squares regression; pulse wave signal
Mesh:
Substances:
Year: 2017 PMID: 29168773 PMCID: PMC5751605 DOI: 10.3390/s17122702
Source DB: PubMed Journal: Sensors (Basel) ISSN: 1424-8220 Impact factor: 3.576
Figure 1Photo and schematic diagram of the fiber Bragg grating (FBG) sensor system.
Figure 2The time-series change in the blood glucose level (subject D).
Reference blood glucose data set.
| Subject (Gender) | Number of Measurements | Blood Glucose Level (mg/dL) | ||
|---|---|---|---|---|
| Maximum | Minimum | Average | ||
| Calibration Data Set | ||||
|
| 50 | 178 | 80 | 119 |
|
| 50 | 232 | 93 | 143 |
|
| 50 | 176 | 89 | 127 |
|
| 50 | 207 | 83 | 138 |
| Validation Data Set | ||||
|
| 10 | 153 | 82 | 113 |
|
| 10 | 188 | 97 | 138 |
|
| 10 | 164 | 89 | 115 |
|
| 10 | 202 | 85 | 129 |
Figure 3Measured pulse wave signal and basic acceleration plethysmogram. (a) Pulse wave signal measured with the FBG sensor; (b) Acceleration plethysmogram.
Figure 4Pulse wave signal in each signal processing method. (a) Pulse wave signals in the shortest-time-cut process. (Blood glucose level, Min: 83 mg/dL, Max: 207 mg/dL, Ave.: 136 mg/dL); (b) Pulse wave signals in the 1-s-normalization process. (Blood glucose level, Min: 83 mg/dL, Max: 207 mg/dL, Ave.: 136 mg/dL).
Figure 5Calibration curve and validation results for calculated blood glucose level (Subject: D, shortest-time-cut and 1-s-normalization processing). (a) Sub.D-calibration curve in Shortest; (b) Sub.D-validation result in Shortest; (c) Sub.D-calibration curve in 1-s; (d) Sub.D-validation result in 1-s.
Calibration curve and validation results for each subject.
| Subject | A | B | C | D | |||||
|---|---|---|---|---|---|---|---|---|---|
| Processing Method | Shortest | 1-s | Shortest | 1-s | Shortest | 1-s | Shortest | 1-s | |
|
| SEC (mg/dL) | 17 | 15 | 34 | 21 | 15 | 14 | 33 | 19 |
| r | 0.67 | 0.77 | 0.58 | 0.86 | 0.84 | 0.87 | 0.44 | 0.86 | |
| factors | 4 | 4 | 4 | 4 | 4 | 4 | 4 | 4 | |
|
| SEP (mg/dL) | 20 | 10 | 23 | 16 | 7 | 12 | 26 | 14 |
| A-zone (%) | 60 | 80 | 80 | 80 | 100 | 100 | 50 | 90 | |
| B-zone (%) | 40 | 20 | 20 | 20 | 0 | 10 | 50 | 10 | |
Figure 6Loading vector of calibration curve in subject D. (a) Loading vector of calibration curve in shortest-time-cut process; (b) Loading vector of calibration curve in the 1-s-normalization process.