Literature DB >> 29146671

Predictors and Long-Term Clinical Outcome of Longitudinal Stent Deformation: Insights From Pooled Analysis of Korean Multicenter Drug-Eluting Stent Cohort.

Tae-Min Rhee1, Kyung Woo Park2, Joo Myung Lee1, Michael S Lee1, Ki-Hyun Jeon1, Hyun-Jae Kang1, Bon-Kwon Koo1, Jay Young Rhew1, Kwang Soo Cha1, Jang-Ho Bae1, Kyoo-Rok Han1, Si-Hoon Park1, Woo-Jung Park1, Seung-Woon Rha1, Seok-Kyu Oh1, Hyuck Moon Kwon1, Ki-Bae Seung1, Taehoon Ahn1, Sang-Hyun Kim1, Hyo-Soo Kim1.   

Abstract

BACKGROUND: There are limited data on the frequency of and factors associated with quantitative coronary angiography (QCA)-defined longitudinal stent deformation (LSD) in various contemporary drug-eluting stents platforms. This study sought to evaluate the predictors of LSD and its long-term clinical implication. METHODS AND
RESULTS: A patient-level pooled analysis was performed with 7350 lesions in 5871 patients treated with platinum-chromium-based everolimus-eluting stent (Promus Element), cobalt-chromium-based everolimus-eluting stent (Promus/Xience V), or cobalt-chromium-based zotarolimus-eluting stent (Endeavor Resolute). QCA was performed to analyze differences of stent length between immediate post-deployment and final post-procedure. Independent factors associated with LSD were identified. Clinical outcomes at 3 years were compared between those with and without QCA-based LSD. The frequency of QCA-based LSD was 1.12% (82 cases). Nine of these cases were angiographically overt. Left main or ostial lesion, bifurcation treatment with provisional side branch stenting or ballooning, additional downstream intervention of a distal lesion, intravascular ultrasound use, and adjunctive post-dilatation were independently associated with QCA-based LSD. The type of stent was not associated with QCA-based LSD. Rates of target lesion failure were nominally higher in lesions with QCA-based LSD than in those without (8.97% versus 5.88%; hazard ratio, 1.415; 95% confidence interval, 0.631-3.175; P=0.399).
CONCLUSIONS: LSD is uncommon with contemporary drug-eluting stents, regardless of the type of stent platform. LSD is mainly associated with procedural factors, especially with additional downstream procedures which require the passage of devices through the stent. Careful manipulation of poststent imaging or procedural devices is required to prevent LSD. More data are needed to clarify the impact of LSD on clinical events.
© 2017 American Heart Association, Inc.

Entities:  

Keywords:  coronary angiography; drug-eluting stents; percutaneous coronary intervention; risk factors; stents

Mesh:

Substances:

Year:  2017        PMID: 29146671     DOI: 10.1161/CIRCINTERVENTIONS.117.005518

Source DB:  PubMed          Journal:  Circ Cardiovasc Interv        ISSN: 1941-7640            Impact factor:   6.546


  3 in total

1.  Needles in Our Technology Haystacks: Defining Efficacy Is Easy, Characterizing Complications Is the Challenge.

Authors:  Elazer R Edelman; Pei-Jiang Wang
Journal:  Circ Cardiovasc Interv       Date:  2017-11       Impact factor: 6.546

2.  Longitudinal stent elongation or shortening after deployment in the coronary arteries: which is dominant?

Authors:  Magdy Algowhary; Mohammed Aboel-Kassem F Abdelmegid
Journal:  Egypt Heart J       Date:  2021-05-17

Review 3.  Intravascular imaging in coronary stent restenosis: Prevention, characterization, and management.

Authors:  Amr Abouelnour; Tommaso Gori
Journal:  Front Cardiovasc Med       Date:  2022-08-09
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.