DETERMINATION OF NATIONAL DIAGNOSTIC REFERENCE LEVELS IN COMPUTED TOMOGRAPHY EXAMINATIONS OF IRAN BY A NEW QUALITY CONTROL-BASED DOSE SURVEY METHOD.

Mehdi Sohrabi1, Masoumeh Parsi1, Fereidoun Mianji2.   

Abstract

National diagnostic reference levels (NDRLs) of Iran were determined for the four most common CT examinations including head, sinus, chest and abdomen/pelvis. A new 'quality control (QC)-based dose survey method', as developed by us, was applied to 157 CT scanners in Iran (2014-15) with different slice classes, models and geographic spread across the country. The NDRLs for head, sinus, chest and abdomen/pelvis examinations are 58, 29, 12 and 14 mGy for CTDIVol and 750, 300, 300 and 650 mGy.cm for DLP, respectively. The 'QC-based dose survey method' was further proven that it is a simple, accurate and practical method for a time and cost-effective NDRLs determination. One effective approach for optimization of the CT examination protocols at the national level is the provision of an adequate standardized training of the radiologists, technicians and medical physicists on the patient radiation protection principles and implementation of the DRL concept in clinical practices.

Entities:  

Mesh:

Year:  2018        PMID: 29136248     DOI: 10.1093/rpd/ncx252

Source DB:  PubMed          Journal:  Radiat Prot Dosimetry        ISSN: 0144-8420            Impact factor:   0.972


  3 in total

1.  Statistical analysis for obtaining optimum number of CT scanners in patient dose surveys for determining national diagnostic reference levels.

Authors:  Mehdi Sohrabi; Masoumeh Parsi; Sanaz Hariri Tabrizi
Journal:  Eur Radiol       Date:  2018-06-28       Impact factor: 5.315

2.  Determination of Diagnostic Reference Level (DRL) in Common Computed Tomography Examinations with the Modified Quality Control-Based Dose Survey Method in Four University Centers: A Comparison of Methods.

Authors:  Jalal Tabesh; Maziyar Mahdavi; Gholamhasan Haddadi; Rezvan Ravanfar Haghighi; Reza Jalli
Journal:  J Biomed Phys Eng       Date:  2021-08-01

3.  Assessment and comparison of radiation dose and image quality in multi-detector CT scanners in non-contrast head and neck examinations.

Authors:  Daryoush Khoramian; Soroush Sistani; Razzagh Abedi Firouzjah
Journal:  Pol J Radiol       Date:  2019-01-23
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.