Literature DB >> 29135865

Electrode Impedance Fluctuations as a Biomarker for Inner Ear Pathology After Cochlear Implantation.

June Choi1, Matthew R Payne, Luke J Campbell, Christo W Bester, Carrie Newbold, Hayden Eastwood, Stephen J O'Leary.   

Abstract

OBJECTIVES/HYPOTHESIS: Cochlear implant surgery now aims to preserve residual low frequency hearing. The current research explores whether fluctuations in the electrical impedance of cochlear implant electrodes may act as a biomarker for pathological changes that lead to the delayed loss of residual hearing. STUDY
DESIGN: Secondary analysis of a double-blinded randomized trial, where methylprednisolone was administered intravenously before cochlear implantation with a view to preserving residual hearing.
METHODS: Seventy-four patients with residual hearing after cochlear implant surgery were investigated for an impedance "spike," defined as a median rise of ≥4 kΩ across all electrodes from the baseline measurements. Spikes were related to objective and subjective hearing loss, dizziness, and tinnitus.
RESULTS: An impedance spike occurred in 14% (10/74) of enrolled patients. Three months after surgery, five patients exhibited spikes and three of these patients had a total loss of their residual hearing. 4.3% of the 69 patients without spikes lost residual hearing. At 1 year, 9 of 10 patients who exhibited spikes had lost all their residual hearing. 8.1% of the 37 patients who did not experience a spike lost their residual hearing. Seventy percent of patients exhibiting a spike also experienced vertigo. The administration of steroids at the time of surgery did not influence the occurrence of spikes.
CONCLUSION: Our results suggest that there is a relationship between a spike and the loss of residual hearing. It seems that rises in impedance can reflect pathology within the inner ear and predict the future loss of residual hearing.

Entities:  

Mesh:

Substances:

Year:  2017        PMID: 29135865     DOI: 10.1097/MAO.0000000000001589

Source DB:  PubMed          Journal:  Otol Neurotol        ISSN: 1531-7129            Impact factor:   2.311


  3 in total

1.  Access and Polarization Electrode Impedance Changes in Electric-Acoustic Stimulation Cochlear Implant Users with Delayed Loss of Acoustic Hearing.

Authors:  Viral D Tejani; Hyejin Yang; Jeong-Seo Kim; Helin Hernandez; Jacob J Oleson; Marlan R Hansen; Bruce J Gantz; Paul J Abbas; Carolyn J Brown
Journal:  J Assoc Res Otolaryngol       Date:  2021-10-22

2.  Residual Hair Cell Responses in Electric-Acoustic Stimulation Cochlear Implant Users with Complete Loss of Acoustic Hearing After Implantation.

Authors:  Viral D Tejani; Jeong-Seo Kim; Jacob J Oleson; Paul J Abbas; Carolyn J Brown; Marlan R Hansen; Bruce J Gantz
Journal:  J Assoc Res Otolaryngol       Date:  2021-02-04

3.  Exploiting Routine Clinical Measures to Inform Strategies for Better Hearing Performance in Cochlear Implant Users.

Authors:  Alan P Sanderson; Edward T F Rogers; Carl A Verschuur; Tracey A Newman
Journal:  Front Neurosci       Date:  2019-01-15       Impact factor: 4.677

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.