Literature DB >> 2909676

Thermosensitive determination of patency in lumboperitoneal shunts. Technical note.

Y Ishiwata1, Y Chiba, T Yamashita, G Gondo, K Ide, T Kuwabara.   

Abstract

Surface cooling and thermistor recording over shunt tubing was used in 23 studies of cerebrospinal fluid shunt patency in 19 patients with lumboperitoneal shunts and normal-pressure hydrocephalus. Shunt patency was shown by downward reflection of the recording trace similar to that obtained for ventriculoperitoneal shunts. Obstruction was demonstrated by a flat-line recording or an upward deflection.

Entities:  

Mesh:

Year:  1989        PMID: 2909676     DOI: 10.3171/jns.1989.70.1.0143

Source DB:  PubMed          Journal:  J Neurosurg        ISSN: 0022-3085            Impact factor:   5.115


  2 in total

1.  Noninvasive thermal evaluation for shunt failure in the emergency room.

Authors:  Jordan Xu; Cassie Poole; Ronald Sahyouni; Jefferson Chen
Journal:  Surg Neurol Int       Date:  2019-12-27

2.  Multimodal Sensing Capabilities for the Detection of Shunt Failure.

Authors:  Milenka Gamero; Woo Seok Kim; Sungcheol Hong; Daniel Vorobiev; Clinton D Morgan; Sung Il Park
Journal:  Sensors (Basel)       Date:  2021-03-03       Impact factor: 3.576