Literature DB >> 29095484

Quantitative evaluation of contrast agent uptake in standard fat-suppressed dynamic contrast-enhanced MRI examinations of the breast.

Evanthia Kousi1, Joely Smith2, Araminta E Ledger1, Erica Scurr1, Steven Allen3, Robin M Wilson3, Elizabeth O'Flynn1, Romney J E Pope3, Martin O Leach1, Maria A Schmidt1.   

Abstract

PURPOSE: To propose a method to quantify T1 and contrast agent uptake in breast dynamic contrast-enhanced (DCE) examinations undertaken with standard clinical fat-suppressed MRI sequences and to demonstrate the proposed approach by comparing the enhancement characteristics of lobular and ductal carcinomas.
METHODS: A standard fat-suppressed DCE of the breast was performed at 1.5 T (Siemens Aera), followed by the acquisition of a proton density (PD)-weighted sequence, also fat suppressed. Both sequences were characterized with test objects (T1 ranging from 30 ms to 2,400 ms) and calibration curves were obtained to enable T1 calculation. The reproducibility and accuracy of the calibration curves were also investigated. Healthy volunteers and patients were scanned with Ethics Committee approval. The effect of B0 field inhomogeneity was assessed in test objects and healthy volunteers. The T1 of breast tumors was calculated at different time points (pre-, peak-, and post-contrast agent administration) for 20 patients, pre-treatment (10 lobular and 10 ductal carcinomas) and the two cancer types were compared (Wilcoxon rank-sum test).
RESULTS: The calibration curves proved to be highly reproducible (coefficient of variation under 10%). T1 measurements were affected by B0 field inhomogeneity, but frequency shifts below 50 Hz introduced only 3% change to fat-suppressed T1 measurements of breast parenchyma in volunteers. The values of T1 measured pre-, peak-, and post-contrast agent administration demonstrated that the dynamic range of the DCE sequence was correct, that is, image intensity is approximately directly proportional to 1/T1 for that range. Significant differences were identified in the width of the distributions of the post-contrast T1 values between lobular and ductal carcinomas (P < 0.05); lobular carcinomas demonstrated a wider range of post-contrast T1 values, potentially related to their infiltrative growth pattern.
CONCLUSIONS: This work has demonstrated the feasibility of fat-suppressed T1 measurements as a tool for clinical studies. The proposed quantitative approach is practical, enabled the detection of differences between lobular and invasive ductal carcinomas, and further enables the optimization of DCE protocols by tailoring the dynamic range of the sequence to the values of T1 measured.
© 2017 The Authors. Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.

Entities:  

Keywords:  DCE-MRI; breast; breast cancer; calibration; quantification

Mesh:

Substances:

Year:  2017        PMID: 29095484      PMCID: PMC5814859          DOI: 10.1002/mp.12652

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


Introduction

Dynamic Contrast‐Enhanced MRI (DCE‐MRI) is a powerful clinical tool for the detection, diagnosis, and staging of the breast cancer.1, 2 Following the administration of a gadolinium chelate contrast agent, tissue perfusion can be estimated using changes in the signal intensity over time on a series of images obtained with fast 3D T1‐weighted pulse sequences. Clinical DCE‐MRI examinations of the breast are commonly performed with fat suppression and provide time‐signal intensity curves to qualitatively assess the enhancement kinetics of the contrast agent uptake in tumors. In contrast, pharmacokinetic modeling offers a quantitative approach to investigate tumor vascularity associated with malignancy and has been shown to improve the diagnostic performance of MRI as well as the prediction of treatment response.3, 4, 5 In this setting, a separate proton density sequence is used as a reference, enabling the calculation of T1. Pharmacokinetic modeling requires rapid data acquisitions, sacrificing spatial resolution and breast coverage, and is not currently a part of the standard clinical practice. The breast MRI is recommended for the assessment of biopsy‐proven invasive lobular carcinomas (ILCs). This is due to the lower diagnostic performance of other imaging modalities in accurately defining the extent of disease,6, 7 which can be caused by the diffuse growth pattern of some ILCs. Differences in the DCE‐MRI enhancement characteristics between ILCs and invasive ductal carcinomas (IDCs) have been previously demonstrated.8, 9 In this work, we propose to combine a high‐resolution fat‐suppressed clinical DCE‐MRI sequence with the ability to perform quantitative T1 measurements by introducing a proton density‐weighted sequence as a reference. We demonstrate our method by comparing the enhancement characteristics of two groups of breast patients: ILCs and IDCs. In addition, we evaluate the accuracy and reproducibility of the obtained T1 values. Furthermore, we show that the dynamic range of our DCE‐MRI sequence is suited to the range of T1 values measured in our clinical breast examinations, enabling contrast agent uptake to be correctly depicted.

Methods

Imaging protocol

Subjects were scanned at 1.5 T (Aera, Siemens, Erlangen, Germany) using an eight‐channel breast receiver coil. Volunteers were scanned with an approval of the Research Ethics Committee (UK NHS HRA/NRES Committee London‐Chelsea 1406/18‐06‐1997) and written consent was obtained. The retrospective analysis of patient studies was approved by the Research Ethics Committee (“Evaluation of Breast MRI Protocols”, Service Evaluation). A standard clinical breast DCE‐MRI protocol was performed using spectrally selective (Spectral Attenuated Inversion Recovery, SPAIR) pulses for fat suppression and three‐dimensional (3D) T1‐weighted spoiled gradient‐echo sequences (TE/TR = 2/4.5 ms, flip angle = 18°, pixel size = 1.31 × 1.31 × 1 mm, parallel imaging factor 2, number of slices = 160, acquisition matrix = 290 × 320, FOV = 380 × 420 mm2). The readout gradient direction was anterior/posterior to minimize cardiac motion artifacts over the breasts. A single dose of contrast agent (Dotarem, Guerbet, France) was administered at 2–3 mL/s (MedRad,USA) depending on the size of the largest vascular access device that could be fitted to the patient. One pre‐ and eight post‐contrast transaxial 3D data sets were acquired in 56 s each, in agreement with the current national guidelines.10 A proton density (PD)‐weighted sequence was subsequently obtained with identical parameters to the DCE‐MRI sequence using a lower flip angle of 4°.

Quantitative assessment of contrast agent uptake

In quantitative DCE examinations undertaken for pharmacokinetic modeling, T1 is calculated from a combination of the two data sets obtained with differing amounts of T1 weighting (spoiled gradient echoes with high and low flip angles).11 The sequence with the high flip angle is T1 weighted while the sequence with low flip angle has minimal T1 weighting. The concentration of contrast agent in each voxel is calculated quantitatively from T1 values: where T1pre is the native T1 of the tissue and r is the relaxivity of the administered contrast agent. Although an analytical solution of the Bloch equations is not usually practical for the fat‐suppressed spoiled gradient‐echo sequences used in DCE, these sequences follow the same principles and both experimental work and numerical simulations demonstrate fat‐suppressed and non‐fat‐suppressed sequences have similar contrast characteristics: image intensity should be proportional to 1/T1 over the T1 range of interest.12 Therefore, subtracting a post‐contrast image (S1) from the pre‐contrast baseline (Spre) demonstrates the contrast agent uptake qualitatively. However, even in a perfect system with no changes in image intensity associated with imperfect excitation or receiver coil sensitivity, quantitative analysis is not possible because there will be (a) variations of image intensity in the same examination associated with differences in the proton density (equilibrium magnetization) from voxel to voxel and (b) changes to coil filling factor and changes to system gain which makes it impossible to compare directly an examination to another or a patient to another. It is possible to define an enhancement ratio (ER) as: As a ratio, ER is not affected by proton density or by coil sensitivity, but is affected by the native T1 value. Variations in ER within a lesion may not relate to different contrast agent uptake, but simply to different native T1 values; the same applies to longitudinal changes in a patient study. ER is only semiquantitative. This is the main motivation to develop a method to calculate T1 using fat‐suppressed sequences. To calculate T1, we introduce a low flip angle image which is practically proton density weighted and also fat suppressed. Using two sequences with different T1 weighting, we calculate T1 post‐contrast for each pixel. For that purpose, the behavior of each sequence is studied with a test object comprising a very wide range of T1 values, and empirical curves are used, instead of the solution of the Bloch equations.11 In a similar way to pharmacokinetic studies, we related the DCE/PD image ratio to the T1 value using an experimental measurement on test objects to provide a calibration curve. This method provided a direct measurement of the T1 value at the end of the DCE protocol; the last frame of the DCE and the PD acquisition are used jointly to calculate T1 (T1post). This approach presumes a slow clearance of the contrast agent from the patient's system between the last DCE and PD acquisitions, and therefore, no significant changes in the contrast agent concentration are expected over those few minutes. In order to calculate T1 values in previous frames, we used the same principles employed in quantitative DCE examinations (without fat suppression); we presumed that all changes of image intensity were associated with T1 (i.e., there was no change to the equilibrium magnetization; there were no other hardware changes; and the signal intensity for DCE was a known function of T1).

Test objects

Plastic tubes filled with aqueous solutions of CuSO4 of different concentrations were used to generate the calibration curves. Standard Inversion Recovery (IR) measurements were employed to provide the reference values for this test object, comprising T1 values within the range 30 ms–2,600 ms. These solutions were scanned with the DCE and PD sequences to produce a curve representing the ratio between the image intensity obtained with DCE and PD images, now referred to as image ratio, as a function of R1 = 1/T1. A separate curve was calculated to provide calibrated image intensity values for the imaging sequences also as a function of R1. A least‐square smooth spline line fitting was performed using R statistical software (R v.3.0.2, www.r-project.org) and the % coefficient of variation was used to evaluate the reproducibility of the calibration curves. The DCE pulse sequence employed has a short TE, and therefore, we do not expect the contrast to have T2—or —weighting for the test objects used in calibration and for the breast. This measurement was repeated on two different occasions separated by 4 months to calibrate the T1 measurement and to evaluate the stability of the calibration curves. For this measurement, the power applied by the fat suppression SPAIR pulse was set to zero, and therefore, the calibration curves were not affected by B0 inhomogeneity. The effect of B0 inhomogeneity was investigated separately, with the same set of solutions. The central frequency was changed in four steps of 50 Hz in both directions (−200 Hz < ω0 < 200 Hz), and therefore, the applied fat suppression pulse partially suppressed water signals. Errors in T1 measurements were attributed to off‐resonance effects. In addition, a uniform test object (T1~110 ms) was scanned with the same sequences to verify whether the ratio between DCE and PD images was constant over the breast volume, as this could be affected by B1 inhomogeneity13 and uniformity filters.14 The percent ratio image uniformity (PRIU) was calculated over the coil volume to be occupied by the breasts according to the following equation:

Clinical examinations

The DCE examinations of 20 patients with histologically confirmed breast tumors were analyzed. Tumors comprised ten lobular carcinomas [six grade 2 ILCs, four lobular carcinomas in situ (LCIS)] and ten ductal carcinomas [two grade 1, five grade 2, and one grade 3 IDCs, two high‐grade ductal carcinomas in situ (DCIS)]. All post‐contrast 3D data sets were registered to the pre‐contrast data set prior to analysis using a rigid registration method with six degrees of freedom (3D Slicer v. 4.4.0, www.slicer.org). The largest transaxial cross section for each tumor was chosen and the tumor outline performed using in‐house software (IDL 8.4 Boulder, CO, USA) and was approved by a specialist breast Radiologist. In addition, DCE and PD data sets were obtained for two healthy volunteers by changing the central frequency in steps of 25 Hz in order to investigate further the off‐resonance effects on clinical T1 calculations. The fibroglandular tissue was segmented using the k‐means clustering algorithm over the entire breast volume (IDL 8.4 Boulder, CO, USA). The T1 relaxation time of the tumors was calculated on a pixel‐by‐pixel basis using the test object calibration curves at three time points: (a) before contrast administration (T1pre), (b) at peak‐contrast uptake, that is, the shortest T1 (T1peak), and (c) the post‐contrast (final) frame of the DCE examination (T1post). In cases of late tumor enhancement, T1peak = T1post. Gadolinium concentration [Gd] and the %ER of the tumors were also calculated pixel‐by‐pixel at peak‐ and post‐enhancement frames, using Eqs. (1) and (2), respectively. The following characteristics of lobular and ductal carcinomas were compared: peak‐enhancement frame, median, and interquartile range (IQR) of the T1 values at the pre‐, peak‐, and the post‐enhancement frames, and of [Gd] and %ER at peak‐ and post‐enhancement. The Wilcoxon rank‐sum test was used for statistical analysis with a significance level of P < 0.05 (R v.3.0.2, www.r-project.org).

Results

Test object study

Figure 1 shows the calibration curves produced using phantom data for the clinical pulse sequences on two separate occasions. Calibration curves were produced for the ratio image intensity [Fig. 1(a)] and the DCE image intensity [Fig. 1(b)]. The image intensity of the DCE sequence can be considered approximately directly proportional to R1 for R1 < 0.01 ms−1 (or T1 > 100 ms). The calibration curves proved to be reproducible: the calculated coefficient of variation for different T1 values varied from 0.1% to 9% for the ratio image and 0.5%–10% for the DCE image. Fat‐supressed T1 measurements were in agreement with IR measurements for the test objects [Fig. 1(c)]: the average absolute difference between fat‐suppressed T1 measurements and IR measurements of the test tube solutions was 7% (range 0.28%–16%); the largest difference was found for the longest T1 (2600 ms).
Figure 1

Calibration curves as measured in two separate occasions, A (black) and B (grey): ratio (a) and DCE (b) image intensity as a function of R1 (ms−1) for test objects with T1 ranging from 30 ms to 2600 ms. (c) Fat‐suppressed T1 measurements agree with the IR measurements for the test objects.

Calibration curves as measured in two separate occasions, A (black) and B (grey): ratio (a) and DCE (b) image intensity as a function of R1 (ms−1) for test objects with T1 ranging from 30 ms to 2600 ms. (c) Fat‐suppressed T1 measurements agree with the IR measurements for the test objects. The curves in Figs. 1(a) and 1(b) are affected by off‐resonance effects if field inhomogeneity causes the fat suppression pulse to suppress water. Frequency shifts under 50 Hz introduced changes up to 19% to T1 measurements in the range 100 ms < T1 < 1000 ms. Figure 2 shows DCE and ratio images of tube solutions, with T1 ranging from 30 ms to 2400 ms, as a function of frequency shift demonstrating a progressive suppression of water signal that results in distorted calibration curves. In volunteer studies, T1 values obtained for normal breast parenchyma showed small variations (< 3%) for frequency shifts below 50 Hz. The T1 of the breast parenchyma was also measured on a set of patients with unilateral disease (Appendix 1).
Figure 2

Test objects (T1 range 30–2400 ms) DCE (top row) and ratio (bottom row) images acquired with different resonance frequencies showing errors being introduced by a gradual suppression of the water signal resulting in distorted calibration curves (right). [Color figure can be viewed at wileyonlinelibrary.com]

Test objects (T1 range 30–2400 ms) DCE (top row) and ratio (bottom row) images acquired with different resonance frequencies showing errors being introduced by a gradual suppression of the water signal resulting in distorted calibration curves (right). [Color figure can be viewed at wileyonlinelibrary.com] Figure 3 shows a transaxial slice of a uniform test object acquired with DCE [Fig. 3(a)] and PD [Fig. 3(b)] sequences and the ratio between them [Fig. 3(c)]. Ratio image is uniform over the coil volume that is occupied by small or large breasts (PRIURight Coil = 93% and 87%, respectively, PRIULeft Coil = 91% and 90%, respectively), suggesting only relatively small errors associated with spatial variations in B0 and B1.
Figure 3

Transaxial slice at the center of a uniform test object acquired with DCE (a) and PD sequences (b) and the corresponding ratio image (c). The right side of the breast coil is shown. High intensity uniformity of the ratio image is demonstrated over the coil area that is occupied by the breast. Dashed lines show the area at the center of the breast coil that is occupied by a small (white dashed line) or large (black dashed line) breast.

Transaxial slice at the center of a uniform test object acquired with DCE (a) and PD sequences (b) and the corresponding ratio image (c). The right side of the breast coil is shown. High intensity uniformity of the ratio image is demonstrated over the coil area that is occupied by the breast. Dashed lines show the area at the center of the breast coil that is occupied by a small (white dashed line) or large (black dashed line) breast.

Clinical study

Figure 4 shows the pre‐, peak‐, and the post‐contrast enhancement frames of two breast examinations, followed by the image with low flip angle (PD) and the corresponding R1 measurements using the test object calibration curves. The % ER was also calculated pixel‐by‐pixel using Eq. (2) and compared with [Gd] post‐enhancement [Eq. (1)]. The graphs show that although both cases have the same ER range, the range of values for the gadolinium concentration is different. Axes have been scaled equally to highlight this observation.
Figure 4

Pre‐, peak‐, and post‐contrast enhancement images followed by the corresponding PD images and R1 maps of a grade II invasive lobular carcinoma (Examination 1) and grade II invasive ductal carcinoma (Examination 2); the mean native T1 ± standard deviation for examinations 1 and 2 is 1311 ± 298 ms and 750 ms ± 139 ms, respectively. Graphs relate the % Enhancement ratio and [Gd] for these examinations. Only in the first example, the [Gd] rises monotonically with the Enhancement Ratio, in the second example, a more complex relationship is mediated by variations in native T1. R1 maps for examinations 1 and 2 have been scaled independently. Within individual examinations, R1 maps have been scaled equally. [Color figure can be viewed at wileyonlinelibrary.com]

Pre‐, peak‐, and post‐contrast enhancement images followed by the corresponding PD images and R1 maps of a grade II invasive lobular carcinoma (Examination 1) and grade II invasive ductal carcinoma (Examination 2); the mean native T1 ± standard deviation for examinations 1 and 2 is 1311 ± 298 ms and 750 ms ± 139 ms, respectively. Graphs relate the % Enhancement ratio and [Gd] for these examinations. Only in the first example, the [Gd] rises monotonically with the Enhancement Ratio, in the second example, a more complex relationship is mediated by variations in native T1. R1 maps for examinations 1 and 2 have been scaled independently. Within individual examinations, R1 maps have been scaled equally. [Color figure can be viewed at wileyonlinelibrary.com] Median T1, [Gd], and %ER values were calculated for every patient; mean ± standard deviation is shown in Table 1. There were no significant differences between the two cancer groups (P > 0.05, P values in Table 1). Peak‐ enhancement occurred in the final frame for four (three ILC and one LCIS) of ten lobular carcinomas suggesting slower uptake of the contrast agent for these tumors. Although peak‐ enhancement occurred earlier for nine of ten ductal carcinomas, this difference was not significant (P = 0.8, Table 1). There were no statistically significant differences in IQR for T1pre, [Gd]peak, [Gd]post, %ERpeak, and %ERpost between the two cancer groups (P > 0.05, P values in Table 2), but the T1 IQR was significantly higher for lobular carcinomas in peak‐ and post‐enhancement frames (P < 0.05, P values in Table 2).
Table 1

Mean ± standard deviation of median T1, [Gd], and %ER across the different time points and median peak‐enhancement frame for the lobular and ductal cancer groups

Tumor groupT1pre (ms)T1peak (ms)T1post (ms)[Gd]peak (×10−4 mmol/ml)[Gd]post (×10−4 mmol/ml)%ERpeak %ERpost Peak‐enhancement frame
Lobular carcinomas (n = 10)1275 ± 623587 ± 367591 ± 3633.9 ± 2.23.9 ± 2.3116 ± 33114 ± 335 (4–8)
Ductal carcinomas (n = 10)1087 ± 471477 ± 193494 ± 2013.9 ± 1.93.7 ± 1.7118 ± 32111 ± 276 (3–8)
P‐value0.9110.81110.8
Table 2

Mean ± standard deviation of IQR T1, [Gd], and %ER for the lobular and ductal cancer groups across the different time points

Tumor groupIQR T1pre (ms)IQR T1peak (ms)IQR T1post (ms)IQR [Gd]peak (×10−4 mmol/ml)IQR [Gd]post (×10−4 mmol/ml)IQR %ERpeak IQR %ERpost
Lobular carcinomas (n = 10)388 ± 251206 ± 110200 ± 1182.1 ± 1.32.4 ± 2.462 ± 2957 ± 25
Ductal carcinomas (n = 10)251 ± 110108 ± 39113 ± 481.5 ± 0.71.5 ± 0.749 ± 1248 ± 10
P‐value0.14 0.02 0.04 0.580.770.430.68

Bold values indicate significant differences.

Mean ± standard deviation of median T1, [Gd], and %ER across the different time points and median peak‐enhancement frame for the lobular and ductal cancer groups Mean ± standard deviation of IQR T1, [Gd], and %ER for the lobular and ductal cancer groups across the different time points Bold values indicate significant differences. Global T1 histograms (100 ms bin size) for the pre‐, peak‐, and post‐contrast enhancement frames are shown in Fig. 5. The number of voxels in the first bin (T1 ≤ 100 ms) falling to the nonlinear part of the calibration curves (Fig. 1) was 0, 18, and 16 for T1pre, T1peak, and T1post, that is, no more than 0.4% of the total number of tumor voxels (4029). Therefore, the obtained T1 values demonstrate that the dynamic range of our DCE sequence suits the range of the T1 values measured in clinical examinations before and after contrast administration; the range of the T1 values for our cohort falls within the range, for which the image intensity is approximately directly proportional to 1/T1 (or R1) for our DCE sequence (Fig. 1). The distribution of the native T1 values (T1pre) is similar between lobular and ductal carcinomas [Fig. 5(a)], whereas the distributions of the T1 values after the contrast administration (T1peak and T1post) suggest greater enhancement variability for the lobular carcinomas [Figs. 5(b) and 5(c)].
Figure 5

Lobular carcinoma (LC) and ductal carcinoma (DC) T1 distributions with the corresponding distribution lines for the pre‐, peak‐, and post‐contrast phases. The bimodal distribution of the post‐enhancement T1 values observed for LC simply shows the uptake variation in the analyzed cases.

Lobular carcinoma (LC) and ductal carcinoma (DC) T1 distributions with the corresponding distribution lines for the pre‐, peak‐, and post‐contrast phases. The bimodal distribution of the post‐enhancement T1 values observed for LC simply shows the uptake variation in the analyzed cases.

Discussion

The value of quantitative measurements in breast has already been demonstrated in the context of pharmacokinetic modeling. Using a small number of patients, we demonstrated a method to assess T1 in clinical fat‐suppressed DCE‐MRI examinations, thus quantifying contrast agent uptake. The proposed quantitative approach broadens the scope of the clinical DCE examination, is practical and achievable on any clinical MRI system, as it simply requires a calibration with test objects. At this stage, the T1 calculations are performed off‐line, but they could be easily integrated as a postprocessing step. The use of quantitative methods enables direct comparisons of examinations in a longitudinal patient study or examinations from different patients; quantitative parameters such as T1 (pre‐ and post‐contrast) and contrast agent concentration can be measured separately. The semiquantitative enhancement ratio, in contrast, is affected by both T1 and contrast‐agent concentration and is therefore more difficult to interpret signal changes relate to the baseline image intensity in a T1‐weighted acquisition. In addition to proposing and demonstrating a novel approach to quantitative breast MRI, we also demonstrated that the dynamic range of our DCE pulse sequence is suitable for our clinical workload; the image intensity is approximately directly proportional to 1/T1 over the range of T1 values we measured in breast lesions. Considering that the national guidelines for the breast screening program require that the contrast characteristics of the DCE sequence are evaluated,10 it is essential to provide methods to do so. To the best of our knowledge, this article is the first to demonstrate that the DCE contrast characteristics are correct for the actual range of T1 values found in clinical practice within our patient population, taking into account specific constraints such as the rate of injection, the contrast agent dose and type, for example, which may vary in different populations. Inaccurate T1 measurements can be caused by the spatial variation in flip angle as a result of B1 field inhomogeneity13 and inaccurate RF transmitter power calibration. The use of uniformity filters in clinical examinations is also a factor that could potentially introduce errors, decreasing the level of confidence in quantification studies. These filters have shown to alter noise distribution resulting in SNR changes.14 Although B1 inhomogeneity is more pronounced at higher fields (≥ 3 T), all the aforementioned sources of error affect both standard T1 calculations in pharmacokinetic modeling and fat‐suppressed T1 measurements. B0 variations are an additional issue, specific to breast DCE with fat suppression; good B0 homogeneity is required to avoid suppression of water signals. In this study, these main factors to affect accuracy of the T1 values obtained with the proposed method were investigated thoroughly. Calibration curves proved to be highly reproducible and a good agreement between IR and fat‐suppressed T1 measurements was demonstrated; the discrepancy found for very long T1 values is probably due to a lower SNR. Spatial variation of the B1 field was investigated with a uniform test object. B1 inhomogeneity and the use of uniformity filters were found not to affect significantly the T1 calculation over the breast volume for our 1.5 T system in a conductive test object. B0 inhomogeneity was simulated altering the central frequency. Small frequency variations introduce calibration errors if the water signal is suppressed, but unintentional water suppression is relatively rare.15 Recent developments in shimming are encouraging16 and will in general lead to improved performance in commercial systems. A separate issue is that DCE images may have fat and water out of phase; in case of fat suppression failure, no quantitative measurements are possible for voxels containing both fat and water. Nevertheless, the measurements of T1 on breast lesions are likely to be less affected by fat suppression failure than the measurements on breast parenchyma, as breast tumors are not expected to have a significant fat content. Comparing lobular carcinomas and invasive ductal carcinomas, we found later enhancement for lobular carcinomas in our cohort, in accordance with previous studies,8, 9 but these differences were not statistically significant. Also, similar peak‐enhancement was found for the two cancer groups, in agreement with Mann et al.8 T1 relaxation time is tissue specific and having a quantitative method to measure it allowed the analysis to go further and interpret the distribution of the T1 values for the lobular and ductal carcinomas. Significant differences in the IQR for T1peak and T1post between the two patient groups were detected, potentially reflecting their distinct growth and invasion patterns. Lobular cancers may grow in a loosely cohesive manner invading surrounding tissue, whereas ductal cancers usually follow a self‐contained solid growth pattern.17 Ductal carcinomas are therefore more likely to present similar characteristics within a patient population. Figure 4 aims to demonstrate that contrast agent uptake is not necessarily proportional to ER, which is only semiquantitative. In only one of the cases presented, contrast agent uptake rises approximately monotonically with increasing ER, and this could be attributed to differences in the baseline T1 values between the two tumors. However, the effects of noise and their dependence on T1 cannot be excluded. The supplementary figure also demonstrates the relationship between ER and [Gd] post‐enhancement for lesions with different native T1 values. We acknowledge the limitations of this study on a small number of subjects; however, our scope was to demonstrate the potential of the proposed approach. We detected significant differences in the range of T1 values post‐contrast between lobular and ductal carcinomas, but no significant differences between the range of contrast agent concentration values or enhancement ratio values. These findings merit further investigation, as T1 values could be proposed as independent biomarkers and be directly related to other tumor characteristics in larger cohorts. Although the sequences employed complied with the DCE‐MRI national guidelines for temporal resolution, we cannot exclude that some variations observed might be system and protocol dependent. Qualitative assessment of the enhancement curves may be reader dependent leading to inconsistent interpretation of uptake in tumors.3 In quantitative studies, many variations can also arise from different MR systems and DCE sequence parameters.18, 19, 20, 21 Ledger et al. highlighted the effect of flip angle and k‐space sampling on fat suppression efficiency, dynamic range, and therefore the relationship between signal intensity and 1/T1 for the range of the expected T1 values.12 Therefore, the proposed method for quantitative T1 measurements in fat‐suppressed DCE may also need to be validated for other sequence parameters and in other systems. Further work is currently in progress. In conclusion, fat‐suppressed T1 measurements are viable in breast DCE, resulting in quantitative measurements of contrast agent uptake. The proposed quantitative approach enables the optimization of DCE protocols by tailoring the dynamic range of the sequence to the values of T1 measured for each population. T1 measurements from clinical fat‐suppressed DCE demonstrated the variations in the T1 range between ductal and lobular cancer within a relatively small number of patients. This work has demonstrated the feasibility of fat‐suppressed T1 measurements as a tool for clinical studies. Fig. S1. %Enhancement Ratio versus [Gd] and native T1 distribution for different lesions. Click here for additional data file.
Table A1

Median T1 values of the contralateral normal‐appearing breast tissue

SubjectMedian T1 (ms)
11036
21130
31319
41069
5751
6635
7757
8507
91501
10629
11881
12952
13689
141410
  22 in total

1.  Magnetic resonance mammography of invasive lobular versus ductal carcinoma: systematic comparison of 811 patients reveals high diagnostic accuracy irrespective of typing.

Authors:  Matthias Dietzel; Pascal A Baltzer; Tibor Vag; Tobias Gröschel; Mieczyslaw Gajda; Oumar Camara; Werner A Kaiser
Journal:  J Comput Assist Tomogr       Date:  2010-07       Impact factor: 1.826

2.  Mammographic detection and staging of invasive lobular carcinoma.

Authors:  Jeroen Veltman; C Boetes; L van Die; P Bult; J G Blickman; J O Barentsz
Journal:  Clin Imaging       Date:  2006 Mar-Apr       Impact factor: 1.605

Review 3.  Dynamic contrast-enhanced breast MR imaging.

Authors:  Marianne Moon; Daniel Cornfeld; Jeffrey Weinreb
Journal:  Magn Reson Imaging Clin N Am       Date:  2009-05       Impact factor: 2.266

4.  On shimming approaches in 3T breast MRI.

Authors:  Ileana Hancu; Ambey Govenkar; Robert E Lenkinski; Seung-Kyun Lee
Journal:  Magn Reson Med       Date:  2012-05-03       Impact factor: 4.668

5.  Application of a mixed imaging sequence for MR imaging characterization of human breast disease.

Authors:  T E Merchant; G R Thelissen; P W de Graaf; C W Nieuwenhuizen; H C Kievit; W Den Otter
Journal:  Acta Radiol       Date:  1993-07       Impact factor: 1.990

6.  Comparison of enhancement characteristics between invasive lobular carcinoma and invasive ductal carcinoma.

Authors:  Ritse M Mann; Jeroen Veltman; Henkjan Huisman; Carla Boetes
Journal:  J Magn Reson Imaging       Date:  2011-08       Impact factor: 4.813

7.  Physiologic changes in breast magnetic resonance imaging during the menstrual cycle: perfusion imaging, signal enhancement, and influence of the T1 relaxation time of breast tissue.

Authors:  Jean-Paul Delille; Priscilla J Slanetz; Eren D Yeh; Daniel B Kopans; Leoncio Garrido
Journal:  Breast J       Date:  2005 Jul-Aug       Impact factor: 2.431

8.  Kinetic curves of malignant lesions are not consistent across MRI systems: need for improved standardization of breast dynamic contrast-enhanced MRI acquisition.

Authors:  Sanaz A Jansen; Akiko Shimauchi; Lindsay Zak; Xiaobing Fan; Abbie M Wood; Gregory S Karczmar; Gillian M Newstead
Journal:  AJR Am J Roentgenol       Date:  2009-09       Impact factor: 3.959

9.  Measurement of pharmacokinetic parameters in histologically graded invasive breast tumours using dynamic contrast-enhanced MRI.

Authors:  A Radjenovic; B J Dall; J P Ridgway; M A Smith
Journal:  Br J Radiol       Date:  2007-12-10       Impact factor: 3.039

10.  Quality assurance in MRI breast screening: comparing signal-to-noise ratio in dynamic contrast-enhanced imaging protocols.

Authors:  Evanthia Kousi; Marco Borri; Jamie Dean; Rafal Panek; Erica Scurr; Martin O Leach; Maria A Schmidt
Journal:  Phys Med Biol       Date:  2015-11-25       Impact factor: 3.609

View more
  2 in total

1.  Combination of magnetic resonance imaging and targeted contrast agent for the diagnosis of myocardial infarction.

Authors:  Jiangjun Qin; Shuchang Zhou; Zhiwei Li; Yinan Chen; Qun Qin; Tao Ai
Journal:  Exp Ther Med       Date:  2018-08-13       Impact factor: 2.447

2.  Systematic Review of Magnetic Resonance Lymphangiography From a Technical Perspective.

Authors:  Michael Mills; Malou van Zanten; Marco Borri; Peter S Mortimer; Kristiana Gordon; Pia Ostergaard; Franklyn A Howe
Journal:  J Magn Reson Imaging       Date:  2021-02-24       Impact factor: 4.813

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.