Wilfredo R Matias1,2, Fabrice E Julceus3, Cademil Abelard3, Leslie M Mayo-Smith4, Molly F Franke1, Jason B Harris4,5, Louise C Ivers1,2,6.
Abstract
BACKGROUND: Rapid diagnostic tests (RDT) for cholera are promising tools for detecting cholera in areas with limited laboratory infrastructure. However, evidence on the characteristics of the many available RDTs is scarce, and their use has been limited by suboptimal performance. We evaluated the performance characteristics of three cholera RDTs from Span Diagnostics, Artron Laboratories, and Standard Diagnostics in a regional laboratory in Haiti. METHODOLOGY/PRINCIPALEntities:
Mesh:
Substances:
Year: 2017 PMID: 29091945 PMCID: PMC5665506 DOI: 10.1371/journal.pone.0186710
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Descriptive characteristics of Crystal VC, Artron and SD Bioline RDTs for cholera.
| Test Name | Manufacturer | Intended Use (per company) | Reported Sensitivity | Reported Specificity |
|---|---|---|---|---|
| Crystal VC Dipstick | Arkray Healtchare Pvt., India | Rapid visual immunochromatographic test for qualitative and differential detection of | 88–100% | 61–87.3% |
| Artron | Artron Laboratories Inc., Canada | Rapid, convenient immunochromatographic assay for the qualitative detection of either | 99% | 99% |
| SD Bioline Cholera Ag O1/O139 RDT | Standard Diagnostics Inc., South Korea | Rapid, qualitative test for the detection of | 95.4% | 94.1% |
*At the time study was performed this test was manufactured by Span Diagnostics Ltd., India.
Fig 1Profile of study specimens.
Results of 3 RDTs for the detection of V. cholerae O1 and O139.
| Culture | ||||
|---|---|---|---|---|
| Positive | Negative | Total | ||
| Crystal VC | O1 | 189 | 29 | 218 |
| O1O139 | 93 | 36 | 129 | |
| O139 | 1 | 25 | 26 | |
| Negative | 3 | 135 | 138 | |
| Total | 286 | 225 | 511 | |
| Artron | O1 | 70 | 11 | 81 |
| O1O139 | 3 | 6 | 9 | |
| O139 | 0 | 3 | 3 | |
| Negative | 1 | 35 | 36 | |
| Total | 74 | 55 | 129 | |
| SD Bioline | O1 | 197 | 15 | 212 |
| O1O139 | 0 | 0 | 0 | |
| O139 | 0 | 0 | 0 | |
| Negative | 46 | 193 | 239 | |
| Total | 243 | 208 | 451 | |
Performance characteristics of 3 RDTs compared to bacterial culture as the gold standard for the detection of V. cholerae O1.
| Crystal VC | Artron | SD Bioline | ||||
|---|---|---|---|---|---|---|
| % | 95% CI | % | 95% CI | % | 95% CI | |
| Sensitivity | 98.6 | 96.5–99.6 | 98.6 | 92.7–100.00 | 81.1 | 75.6–85.8 |
| Specificity | 71.1 | 64.7–76.9 | 69.1 | 55.2–80.9 | 92.8 | 88.4–95.9 |
| Positive Predictive Value | 81.3 | 76.8–85.2 | 81.1 | 71.5–88.6 | 92.9 | 88.6–96.0 |
| Negative Predictive Value | 97.6 | 93.9–99.3 | 97.4 | 86.5–99.9 | 80.8 | 75.2–85.6 |
Performance characteristics of Crystal VC and Artron RDTs for the detection of V. cholerae O1 when O1 or O139 bands are considered positive.
| Crystal VC | Artron | |||
|---|---|---|---|---|
| % | 95% CI | % | 95% CI | |
| Sensitivity | 99.0 | 97.0–99.8 | 98.6 | 92.7–100.00 |
| Specificity | 60.0 | 53.3–66.5 | 63.6 | 49.6–76.2 |
| Positive Predictive Value | 75.9 | 71.2–80.1 | 78.5 | 68.8–86.3 |
| Negative Predictive Value | 97.8 | 93.8–99.5 | 97.2 | 85.5–99.9 |
Fig 2Representative images of positive Crystal VC, Artron and SD Bioline RDTs.