| Literature DB >> 29085227 |
Ming Li1, Lin Chen2, Li-Ming Liu3, Yong-Li Li2, Bo-An Li2, Bo Li2, Yuan-Li Mao2, Li-Fang Xia2, Tong Wang2, Ya-Nan Liu2, Zheng Li4, Tong-Sheng Guo5.
Abstract
AIM: To investigate and compare the analytical and clinical performance of TianLong automatic hypersensitive hepatitis B virus (HBV) DNA quantification system and Roche CAP/CTM system.Entities:
Keywords: Analytical performance; Clinical performance; Hepatitis B virus; Hepatitis B virus DNA quantification; Real-time quantification PCR
Mesh:
Substances:
Year: 2017 PMID: 29085227 PMCID: PMC5645617 DOI: 10.3748/wjg.v23.i37.6845
Source DB: PubMed Journal: World J Gastroenterol ISSN: 1007-9327 Impact factor: 5.742
Detail information of national standard materials
| HS-0001 | GBW (E) 090137 | 201511003 | HBV DNA serum standard material | 1.41 × 103 IU/mL |
| HS-0002 | GBW (E) 090138 | 201512004 | HBV DNA serum standard material | 5.9 × 104 IU/mL |
| HS-0003 | GBW (E) 090139 | 201511003 | HBV DNA serum standard material | 4.6 × 105 IU/mL |
Limit of detection test results
| 2.5 | 10 | 15 | 40% | 12 | 13 | 48% | 11 | 14 | 44% | 44.0% |
| 5 | 21 | 4 | 84% | 20 | 5 | 80% | 22 | 3 | 88% | 84.0% |
| 10 | 25 | 0 | 100% | 24 | 1 | 96% | 25 | 0 | 100% | 98.7% |
| 15 | 25 | 0 | 100% | 25 | 0 | 100% | 25 | 0 | 100% | 100.0% |
| 30 | 25 | 0 | 100% | 25 | 0 | 100% | 25 | 0 | 100% | 100.0% |
Limit of quantification test results
| 10 | 1 | 0.79 | 40.57% | 19 | 76.00% | 65.33% |
| 2 | 0.59 | 47.58% | 13 | 52.00% | ||
| 3 | 0.71 | 36.53% | 17 | 68.00% | ||
| 20 | 1 | 1.45 | 14.11% | 21 | 84.00% | 82.67% |
| 2 | 1.45 | 16.65% | 20 | 80.00% | ||
| 3 | 1.36 | 20.55% | 21 | 84.00% | ||
| 30 | 1 | 1.49 | 6.09% | 25 | 100.00% | 100.00% |
| 2 | 1.40 | 8.51% | 25 | 100.00% | ||
| 3 | 1.46 | 5.55% | 25 | 100.00% | ||
Accuracy test results
| GBW (E) 090137 | 1.41 × 103 IU/mL | 3.15 | 3.11 ± 0.02 | 0.04 |
| GBW (E) 090138 | 5.9 × 104 IU/mL | 4.77 | 4.80 ± 0.07 | 0.07 |
| GBW (E) 090139 | 4.6 × 105 IU/mL | 5.66 | 5.70 ± 0.05 | 0.05 |
Figure 1The correlation of measured hepatitis B virus DNA concentrations between the TL system assay and the CAP/CTM assay. HBV: Hepatitis B virus.
Figure 2The difference of measured hepatitis B virus DNA concentrations between the TL system assay and the CAP/CTM assay. HBV: Hepatitis B virus.
Precision and reproducibility test results
| Intra assay precision | 6.97 | 0.23% | 4.06 | 1.10% | 2.29 | 3.10% | ||
| Inter assay precision | 1st Day | 6.94 | 0.91% | 4.19 | 1.82% | 2.26 | 3.79% | |
| 2nd Day | 6.95 | 1.04% | 4.31 | 1.49% | 2.41 | 3.25% | ||
| 3rd Day | 7.08 | 0.64% | 4.31 | 1.67% | 2.39 | 3.72% | ||
| 4th Day | 7.10 | 0.81% | 4.32 | 1.63% | 2.42 | 3.47% | ||
| Reproducibility | 7.02 | 1.36% | 4.28 | 2.08% | 2.37 | 4.41% | ||
n = 20;
n = 15;
n = 60 (15 tests/d × 4 d).
Figure 3Linear correlation test standard curve. HBV: Hepatitis B virus.
Hepatitis B virus genotypes coverage test Ct value
| A | 25.62 | 25.89 | 25.71 | Undetected |
| B | 25.58 | 25.78 | 25.78 | Undetected |
| C | 25.94 | 25.86 | 25.99 | Undetected |
| D | 22.61 | 22.78 | 22.78 | Undetected |
| E | 22.67 | 22.70 | 22.68 | Undetected |
| F | 22.68 | 22.72 | 22.71 | Undetected |
| G | 24.50 | 24.78 | 24.76 | Undetected |
| H | 24.81 | 24.80 | 24.85 | Undetected |